| Literature DB >> 33684722 |
Yan Jin1, Yang Li1, Muhammad Erfan Uddin1, Alex Sparreboom1, Shuiying Hu2.
Abstract
A simple, rapid, and sensitive LC-MS/MS method for determining concentrations of the anticancer alkaloid vincristine in micro volumes of mouse plasma was developed and validated in positive ion mode. Separation of vincristine and the internal standard [2H3]-vincristine was achieved on an Accucore aQ column with a gradient mobile phase delivered at a flow rate of 0.4 mL/min and a run time of 2.2 min. Calibration curves were linear (r2 > 0.99, n = 8) up to 250 ng/mL, with a lower limit of quantitation of 2.5 ng/mL. The matrix effect and extraction recovery for vincristine were ranging 108-110% and 88.4-107%, respectively. The intra-day and inter-day precision of quality controls tested at 3 different concentrations were always less than 15%, and accuracy ranged from 91.7 to 107%. The method was successfully applied to evaluate the pharmacokinetic profile of vincristine in wild-type and CYP3A-deficient mice in support of a project to provide mechanistic insight into drug-drug interactions and to identify sources of inter-individual pharmacokinetic variability associated with vincristine-induced peripheral neuropathy.Entities:
Keywords: Mouse plasma; Pharmacokinetics; UHPLC-MS/MS; Vincristine
Mesh:
Substances:
Year: 2021 PMID: 33684722 PMCID: PMC7987214 DOI: 10.1016/j.jchromb.2021.122591
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.318