Occupational radiation exposure of orthopedists at different locations during pedicle screw insertion: An anthropomorphic phantom study.

Hsien-Wen Chiang1, Re-Wen Wu2, Tsung-Cheng Yin2, Jung-Hui Li1, Yue-Sheng Wang1, Li-Han Lin1.   

Abstract

BACKGROUND: The purpose of this work was to evaluate orthopedic surgeons' exposure to occupational radiation doses from scattering using a mobile flat panel C-arm X-ray machine at different standing positions during an intraoperative pedicle screw implantation.
OBJECTIVE: Evaluate the radiation dose received by medical staff, by applying flat X-ray machine in surgical room during an intraoperative pedicle screw implantation.
METHODS: A mobile flat-panel C-arm X-ray machine at a dedicated orthopedic operating room was used to image an anthropomorphic female phantom which was set in a prone position on the operating table. The X-ray was projected horizontally, and 1 minute continuous fluoroscopy was used for lumbar spine and thoracolumbar spine during pedicle screw implantation. Scattering radiation doses to orthopedic surgeons were measured at different standing positions and body heights (50, 100, 150 cm above the ground) with and without limited collimations.
RESULTS: The dose area product (DAP) in this experiment is normalized as 343 μGy⋅m2. In the four areas, the lowest scattered radiation measured by DF is 11.2 vs. 0.7 μSv, outside and inside the lead suit, respectively, with or without restricted field, 150 cm above the ground, and the lowest scattered radiation dose inside the lead suit. It is 1.3 vs. 0.5 μSv. Comparing the highest dose of the TF at with the lowest dose of the DF, the average result is 73.7 vs. 11.1 μSv, P< 0.05.
CONCLUSIONS: Using a mobile flat-panel C-arm X-ray machine during a pedicle screw implantation, the minimum scattering radiation to surgeons was found to be at the terminal DF area based on the analysis of the scattering doses orthopedic surgeons were exposed to.

Entities:  

Keywords:  Fluoroscopy; effective dose; pedicle screw insertion; scattered radiation

Year:  2021        PMID: 33682760     DOI: 10.3233/THC-218021

Source DB:  PubMed          Journal:  Technol Health Care        ISSN: 0928-7329            Impact factor:   1.285


  1 in total

1.  Image Quality and Dose Comparison of 3 Mobile Intraoperative Three-Dimensional Imaging Systems in Spine Surgery.

Authors:  Norah Foster; Christopher Shaffrey; Avery Buchholz; Raymond Turner; Lexie Zidanyue Yang; Donna Niedzwiecki; Allen Goode
Journal:  World Neurosurg       Date:  2022-01-01       Impact factor: 2.210

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.