Martin Kolen1, Wilson A Smith1, Fokko M Mulder1. 1. Materials for Energy Conversion and Storage (MECS), Department of Chemical Engineering, Delft University of Technology, Van der Maasweg 9, 2629 HZ Delft, The Netherlands.
Abstract
Direct electrolytic N2 reduction to ammonia (NH3) is a renewable alternative to the Haber-Bosch process. The activity and selectivity of electrocatalysts are evaluated by measuring the amount of NH3 in the electrolyte. Quantitative 1H nuclear magnetic resonance (qNMR) detection reduces the bench time to analyze samples of NH3 (present in the assay as NH4 +) compared to conventional spectrophotometric methods. However, many groups do not have access to an NMR spectrometer with sufficiently high sensitivity. We report that by adding 1 mM paramagnetic Gd3+ ions to the NMR sample, the required analysis time can be reduced by an order of magnitude such that fast NH4 + detection becomes accessible with a standard NMR spectrometer. Accurate, internally calibrated quantification is possible over a wide pH range.
Direct electrolytic N2 reduction to ammonia (NH3) is a renewable alternative to the Haber-Bosch process. The activity and selectivity of electrocatalysts are evaluated by measuring the amount of NH3 in the electrolyte. Quantitative 1H nuclear magnetic resonance (qNMR) detection reduces the bench time to analyze samples of NH3 (present in the assay as NH4 +) compared to conventional spectrophotometric methods. However, many groups do not have access to an NMR spectrometer with sufficiently high sensitivity. We report that by adding 1 mM paramagnetic Gd3+ ions to the NMR sample, the required analysis time can be reduced by an order of magnitude such that fast NH4 + detection becomes accessible with a standard NMR spectrometer. Accurate, internally calibrated quantification is possible over a wide pH range.
Ammonia (NH3) is one of the largest chemical commodities
responsible for about 1.5% of global energy use and associated CO2 emissions from the Haber–Bosch process. Its primary
use is as a feedstock for nitrogen-based fertilizers. Electrochemical,
fossil-fuel-free methods to produce ammonia are gaining significant
interest for the reduction of CO2 emissions, as well as
enable ammonia as a carbon-free energy carrier and storage material.[1−3] Direct electrolytic N2 reduction is a renewable alternative
to synthesize NH3. To suppress the undesirable hydrogen
evolution reaction, a selective electrocatalyst is needed.[4]The activity and selectivity of electrocatalysts
are quantified
by measuring the accumulated NH3 in the electrolyte with
appropriate detection methods. Recently, a number of papers were published
concerning the difficulty of obtaining reproducible results in nitrogen
reduction research.[5,6] Such difficulty is related to
experimental procedures and significant amounts of NH3 from
dust, ambient air, 15N2, and desorption from
cell surfaces. In addition, NOx contamination or nitrogen-containing
catalyst precursors can be reduced to NH3 during electrolysis,
which can be falsely attributed to N2 reduction.[7,8] Reliable testing and analysis procedures including control experiments
with an isotope labeled 15N2 are necessary to
avoid false positives.[5] Conventional spectrophotometric
NH3 detection methods such as the indophenol blue method
cannot distinguish between isotopologues of NH3 and require
considerable bench time.[9,10]1H NMR spectroscopy
is a fast, accessible, and isotopically selective alternative to spectrophotometric
methods for NH3 detection, but as we will show below, the
sensitivity on a standard 400 MHz spectrometer is insufficient.[11] Here, we present a powerful liquid-state NMR
method with sufficient sensitivity on relatively easily accessible
NMR spectrometers, i.e., requiring limited field strength and normal
sensitivity probes.Several alternatives to spectrophotometric
NH3 detection
methods have been proposed recently. Ion chromatography can be used
for ammonia detection but isotopologues cannot be distinguished and
an overlap of NH4+ with other cations poses
a threat to the accuracy of the method.[12] Yu et al. proposed ultrahigh-performance liquid chromatography–mass
spectroscopy (UPLC-MS) to measure derivatized solutions of NH3.[13] The method is very sensitive
and capable of distinguishing isotopologues of NH3 but
requires careful control over the pH. Quantitative 1H NMR
has been widely adopted to quantify 15NH4+ from control experiments with 15N2.
The acidified form of ammonia, ammonium (14NH4+), and its isotopologue 15NH4+ have an unmistakable fingerprint in the 1H NMR
spectrum.[11]Quantitative 1H NMR is based on the proportional relationship
between the signal integral I and the number of protons N responsible for that particular signalwhere KS is a
proportionality factor that depends on the physicochemical properties
of the sample. To achieve accurate quantification, changes in KS have to be accounted for by a suitable quantification
method.[14] Pulse length-based concentration
determination (PULCON) uses the principle of reciprocity to correlate
the absolute intensity of two spectra measured in different solutions.[15] Nielander et al. successfully applied the PULCON
method to NH4+ quantification.[11]Since fluctuations of KS affect all
resonances in the spectrum equally, the ratio of two peaks is independent
of KS and can therefore be used for quantification.
Typically, an internal standard of known concentration is added as
a reference. The concentration of NH4+ can be
quantified by either relative or absolute quantification. For relative
quantification, a calibration curve is generated by measuring standard
solutions of NH4+ and an internal standard.[14] The prerequisites for accurate ammonia quantification
with relative quantification were recently described..[16] Absolute quantification allows the calculation
of the NH4+ concentration directly from the
integral of the peaks of NH4+ and the internal
standard without requiring a calibration curve according towhere I, N, and C are the integral area,
number of nuclei,
and concentration of NH4+ and standard, respectively.
Absolute quantification requires that the total time spent to acquire
one scan, the interscan delay tscan, is
at least 5 times the longest longitudinal relaxation time T1 in the sample.[14] Despite the advantages of absolute quantification (calibration-free
and robust) so far, no absolute quantification method has been proposed
for ammonia detection. Hodgetts et al. reported that a d1, T1, and proton exchange-induced
loss of coherence affects the NH4+ peak, rendering
absolute quantification not suitable for NH4+ detection.[16] By adding a suitable paramagnetic
salt, accurate absolute quantification becomes possible because the T1 values of both the internal standard and NH4+ are reduced so that there is insufficient time
for the proton exchange to induce a loss of coherence as we will see
below.The lower limit of quantification (LOQ) of a detection
method is
the lowest concentration of NH3 that can be measured within
an acceptable time and with an acceptable accuracy. The LOQ depends
on the sensitivity, which is calculated from the signal-to-noise ratio
(SNR) at a certain interscan delay tscanTo reduce the minimum LOQ by a factor of 2, the analysis time
has
to be quadrupled.[17] Nielander et al. could
detect 1 μM NH4+ in ethanol within 1 h
(tscan = 2 s) with a 900 MHz NMR and cryo-probe.[11] The sensitivity difference between a 900 MHz
NMR and a standard laboratory NMR (400 MHz) is substantial. The type
of probe and the field strength difference lead to 1 order of magnitude
lower sensitivity, which leads to 2 orders of magnitude longer analysis
time to achieve the same LOQ on a 400 MHz NMR without cryo-probe.[18] To compensate for the lower sensitivity, longer
experiments (typically several hours) are necessary to accumulate
enough NH4+ in the electrolyte to reach the
detection limit, which is unfavorable and which in addition increases
the risk of false negatives due to deactivation and of false positives
due to contamination. Higher 1H NMR sensitivity is needed
to enable laboratories with more standard NMR spectrometers to quantify
NH3 efficiently.The type of the pulse sequence influences
the NH4+ sensitivity strongly.[11,16] The signal from the
hydrogen atoms of the solvent has to be suppressed to avoid baseline
distortions and low receiver gain. Nielander et al. showed that pulse
sequences that utilize pulsed field gradients in combination with
selective excitation pulses are very effective at suppressing water
without removing the NH4+ signal.[11] These pulse sequences use pulsed field gradients
to dephase the water resonance and selective pulses to ensure that
during acquisition water is completely out of phase while NH4+ is in phase.[19]The T1 of a molecule influences the
sensitivity, because for a given interscan delay tscan, T1 determines the percentage
of spins that can relax back to equilibrium in between scans. A smaller
percentage relaxation leads to less acquired signal per scan according
towhere M and M0 are the magnetization in the z-axis following tscan and at full relaxation,
respectively. The interscan delay tscan is composed of the recycle delay d1 and
the acquisition time. It is noteworthy that for some pulse sequences,
the percentage relaxation only depends on the recycle delay, not on
the acquisition time, as will be discussed in more detail below. Reducing
the interscan delay, for example, by fast sampling is a well-known
strategy to improve the 1H NMR sensitivity.[20] Another strategy to lower the interscan delay
is to shorten the T1 of the analyte, which
has the advantage that the same percentage relaxation can be achieved
at a lower interscan delay.[14]T1 is determined by the fluctuating magnetic interactions
due to nearby magnetic moment fluctuations and due to positional changes
of surrounding nuclei and moments. Interactions with unpaired electrons
of paramagnetic substances are 1000 times larger than typical interactions
between nuclear magnetic moments. Therefore, a small amount of a paramagnetic
substance is sufficient to lower T1 drastically.[21] This concept is applied in contrast agents for
medical magnetic resonance imaging (MRI). The so-called paramagnetic
relaxation enhancement (PRE) is also a common strategy to overcome
sensitivity barriers for small organic molecules and proteins, because
as T1 decreases, more scans can be acquired
in the same amount of time.[18,22,23]
Results and Discussion
The Gd3+ ion is widely
used for PRE in medical MRI due
to its large magnetic moment from seven unpaired electrons.[24] We investigated the influence of paramagnetic
Gd3+ ions on the sensitivity for aqueous ammonia detection
to enable 1H NMR as a routine analysis tool for NH4+ quantification, with the use of an internal standard
(absolute quantification). In agreement with Nielander et al., we
found that pulse sequences that are suppressing the water resonance
by dephasing it during acquisition are well suited for NH4+ detection.[11] With excitation
sculpting (ES), the water resonance is suppressed effectively and
a flat baseline is obtained around the NH4+ triplet.
However, at 40 μM NH4+, the SNR is only
13.6 for a 12.8 min measurement on a 400 MHz NMR with room-temperature
probe (see Figure a). With this sensitivity, it takes 4 h until the accumulated ammonia
in the electrolyte produced by a catalyst with intermediate activity
becomes quantifiable by NMR (calculation in the Supporting Information, SI). Therefore, we sought to improve the sensitivity
by adding 1 mM paramagnetic Gd3+ to the NMR tube. Maleic
acid (MA) was added as an internal standard to quantify the amount
of NH4+ with absolute quantification. The singlet
of maleic acid at ca. 6.21 ppm is sufficiently separated from the
NH4+ triplet at ca. 6.9 ppm. The T1 values of both NH4+ and MA decrease
drastically after the addition of Gd3+. T1 decreases from 2.16 to 0.14 s and from 2.05 to 0.13
s for NH4+ and MA, respectively. This 15.4-fold
reduction of the T1 of NH4+ enables a reduction of the interscan delay by the same factor,
which, according to eq , leads to a potential 3.9-fold sensitivity increase (). The linewidth of NH4+ increases only slightly
with the addition of Gd3+ from
3.6 to 4.2 Hz.
Figure 1
1H NMR sensitivity gain from 0 mM Gd3+ (black,
green, and gray) to 1 mM Gd3+ (red, blue, orange) in the
NMR tube measured with different acquisition parameters and pulse
sequences. (a) 40 μM NH4+ measured with
identical total acquisition time (12.8 min) and a recycle delay d1 of 10 and 0.75 s for 0 mM Gd3+ and
1 mM Gd3+, respectively. Pulse sequence: Excitation sculpting.
(b) 40 μM NH4+ measured with identical
total acquisition time (10.7 min) and recycle delay (0.5 s). Pulse
sequence: Excitation sculpting. (c) 40 μM NH4+ measured with the same acquisition parameters as (b) but
with the double pulsed field gradient spin echo (DPFGSE) pulse sequence.
All SNR values are averages from a triplet measurement. (d) Effect
of 1 mM Gd3+ on T1 of NH4+ and maleic acid. Field strength: 400 MHz.
1H NMR sensitivity gain from 0 mM Gd3+ (black,
green, and gray) to 1 mM Gd3+ (red, blue, orange) in the
NMR tube measured with different acquisition parameters and pulse
sequences. (a) 40 μM NH4+ measured with
identical total acquisition time (12.8 min) and a recycle delay d1 of 10 and 0.75 s for 0 mM Gd3+ and
1 mM Gd3+, respectively. Pulse sequence: Excitation sculpting.
(b) 40 μM NH4+ measured with identical
total acquisition time (10.7 min) and recycle delay (0.5 s). Pulse
sequence: Excitation sculpting. (c) 40 μM NH4+ measured with the same acquisition parameters as (b) but
with the double pulsed field gradient spin echo (DPFGSE) pulse sequence.
All SNR values are averages from a triplet measurement. (d) Effect
of 1 mM Gd3+ on T1 of NH4+ and maleic acid. Field strength: 400 MHz.To show that the measured sensitivity gain matches
the sensitivity
gain predicted from T1 measurements, we
measured a sample of 40 μM NH4+ with and
without 1 mM Gd3+ using different acquisition parameters
(Figure a–c).
In Figure a, the total
analysis time is identical in both measurements and d1 is set to 5T1 so that NH4+ has the same percentage relaxation in both cases.
With 1 mM Gd3+, the sensitivity is significantly (factor
2.4) higher but not as much as expected from the T1 decrease (factor 3.9). We will later show that a sensitivity
gain close to the predicted value can be measured directly by removing
an additional 90° pulse that is by default included in the ES
pulse sequence. With the default version of ES, the sensitivity gain
is lower than expected from the T1 decrease
because the additional 90° pulse removes the contribution of
the acquisition time to the percentage relaxation. Consequently, the
acquisition time only adds time to the total analysis time without
improving signal strength and the percentage relaxation depends only
on d1. Since the acquisition time makes
up a larger fraction of the interscan delay at low interscan delays,
the decrease of sensitivity is more pronounced with 1 mM Gd3+ where the acquisition time makes up 0.75 s of the total 1.5 s interscan
delay. Without Gd3+, only 2 s out of 12 s interscan delay
is the acquisition time that leads to a smaller sensitivity loss.
In other words, the interscan delay could be factors 1.2 and 2 smaller
for 0 mM Gd3+ and 1 mM Gd3+, respectively. Therefore,
the sensitivity gain with 1 mM Gd3+ would increase by a
factor 1.3 () from 2.4 to 3.1 if the sensitivity loss
would have been equal in both cases. Taking into account ≈15%
sensitivity loss due to line broadening, the sensitivity gain is 3.6,
which is close to the predicted value.The experiment shown
in Figure a is not
sufficient to prove a sensitivity gain because
it only shows that with a higher recycle delay, less scans can be
acquired in the same amount of time. Less scans will always lead to
lower SNR. To prove a sensitivity increase, it is necessary to show
that a larger recycle delay is necessary with 0 mM Gd3+ but not with 1 mM Gd3+. This is shown in Figure b, where both 0 mM Gd3+ and 1 mM Gd3+ were measured with low recycle delay (0.5
s) and identical total acquisition time. The sensitivity without Gd3+ is 3.9 times lower, indicating that a large fraction of
the signal is lost due to low percentage relaxation. The percentage
relaxation at a recycle delay of 0.5 s is 20.7 and 97.2% for T1 of 2.16 and 0.14 s, respectively. Therefore,
4.7 times more signal can be expected with 1 mM Gd3+ in
the same amount of time. We assume that 15% of that signal increase
is lost due to line broadening with Gd3+, which results
in sensitivity improvement by a factor of 4.08. This value agrees
well with the experimentally observed value of 3.9.To study
if adding Gd3+ also improves the sensitivity
with other pulse sequences, we measured the sensitivity gain with
the double pulsed field gradient spin echo (DPFGSE) pulse sequence
(Figure c) using the
same acquisition parameters as in Figure b. The sensitivity gain with DPFGSE (2.1)
is lower than with ES (3.9). The reason for this is that with DPFGSE,
the percentage relaxation has to be calculated using the full interscan
delay including acquisition time, not just the d1 as for ES. Using the same methodology as in (b), we calculate
the percentage relaxation with and without Gd3+ and arrive
at an expected sensitivity gain of 1.8, which agrees well with the
experimentally measured value. The sensitivity gain is lower than
in (b) because with 1 mM Gd3+, the chosen tscan of 1.25 s is almost 9 times longer than the T1 of NH4+, which means
that tscan is much longer than the necessary
5T1, and as a result, sensitivity is lost.
The previous examples demonstrate that after addition of 1 mM Gd3+, a significant sensitivity gain is observed with different
acquisition parameters and pulse sequences, and this sensitivity gain
agrees well with the expected values predicted from T1 measurements.We measured the accuracy of NH4+ quantification
with 1 mM Gd3+ by calculating the NH4+ concentration from the intensities of MA and NH4+ using eq and
comparing it to the gravimetrically measured concentration (Figure a,b). The method
has very good linearity (R2 = 0.999) and
an acceptable relative error (≤10%) in the NH4+ concentration range of 30–388 μM with the ES
pulse sequence. The relative error is randomly distributed around
the abscissa, which suggests that it is caused by integration errors.
Higher accuracy (relative error ≤ 5.3%) was obtained with an
isotope labeled 15NH4+ (see Figure S2). 15NH4+ can be quantified with higher accuracy because it appears in the
NMR spectrum as a doublet, which has inherently higher SNR than the 14NH4+ triplet.
Figure 2
(a) Linearity and (b)
accuracy of NH4+ quantification
with 1 mM Gd3+. Error bars around each point represent
the standard deviation for each triplicate measurement. Pulse sequence:
Excitation sculpting, recycle delay: 0.5 s, acquisition time: 0.75
s, total analysis time: 10.7 min, field strength: 400 MHz.
(a) Linearity and (b)
accuracy of NH4+ quantification
with 1 mM Gd3+. Error bars around each point represent
the standard deviation for each triplicate measurement. Pulse sequence:
Excitation sculpting, recycle delay: 0.5 s, acquisition time: 0.75
s, total analysis time: 10.7 min, field strength: 400 MHz.The pH of the catholyte, which is used for detection, can
vary
over time due to acidic or alkaline species produced in the electrochemical
reaction or because of migration of ions induced by the electric field.[25,26] N2 reduction experiments are especially prone to pH changes
because the electrolyte volume is minimized to maximize the signal
for ammonia detection. Both UPLC-MS and the indophenol method are
sensitive to pH changes because the pH influences the reaction that
is carried out prior to analysis.[9,13] Therefore,
additional dilution steps can be necessary to measure accurately with
these methods.To investigate if the accuracy of our 1H NMR method
depends on the pH, we acidified a sample of 388 μM NH3 with different concentrations of H2SO4 (Figure a). Based on the
previous finding that the T1 values of
NH4+ and MA are very close to each other, we
chose a recycle delay of 0.5 s (3T1) for
this experiment. For acid concentrations above 37 mM H2SO4, the relative error continuously increases when more
acid is added. To investigate if the growing relative error might
be caused by changing T1 values, we measured
the T1 values of NH4+ and MA at 370 mM H2SO4 (Figure b). The gap between the T1 values of NH4+ and MA is slightly
larger at 370 mM H2SO4 than at 37 mM H2SO4 which might explain the larger error. After increasing
the recycle delay from 0.5 to 1.5 s to compensate for the increased T1 gap, the relative error decreases to <2%
between 37 and 222 mM H2SO4. This suggests that
the detection method is accurate over a wide pH range if a higher d1 is chosen to compensate for T1 changes.
Figure 3
Influence of H2SO4 concentration
in the NMR
tube on the accuracy of NH4+ quantification
with d1 of 0.5 s (red) and 1.5 s (black).
Error bars around each point represent the standard deviation for
each triplicate measurement. (a) T1 of
NH4+ and maleic acid at two different H2SO4 concentrations. (b) Acquisition parameters:
at = 0.75 s, nt = 512, Pulse sequence: Excitation sculpting, field
strength: 400 MHz.
Influence of H2SO4 concentration
in the NMR
tube on the accuracy of NH4+ quantification
with d1 of 0.5 s (red) and 1.5 s (black).
Error bars around each point represent the standard deviation for
each triplicate measurement. (a) T1 of
NH4+ and maleic acid at two different H2SO4 concentrations. (b) Acquisition parameters:
at = 0.75 s, nt = 512, Pulse sequence: Excitation sculpting, field
strength: 400 MHz.In Figure , even
with a high recycle delay of 1.5 s, an unusually high error remains
at the highest and lowest acid concentrations. The error at the lowest
acid concentrations is in agreement with the results by Hodgetts et
al. and is caused by deprotonation of MA below 20 mM H2SO4.[16] Spectra acquired at
the highest acid concentration had phasing issues, which had to be
corrected by postprocessing the spectrum using the autophasing algorithm
in the software package MestReNova. We suspect that the phasing issues
are caused by tuning and matching, which become more difficult at
high salt concentrations.[14] To achieve
maximum accuracy, the acid concentration should not exceed 222 mM.As discussed previously, with the default settings of the ES pulse
sequence, the acquisition time does not contribute to the percentage
relaxation so that sensitivity is lost. To determine the maximum sensitivity
for NH4+ detection with 1 mM Gd3+, we deactivated the additional 90° pulse at the beginning of
the pulse sequence so that both acquisition time and recycle delay
contribute to the percentage relaxation. This leads to a significant
increase in sensitivity (see Figure ). The sensitivity can be further increased by reducing
the interscan delay from 5T1 to 3T1 which is feasible in this case because the T1 values of NH4+ and MA
are very close to each other. The SNR of a 40 μM NH4+ sample measured for 14.6 min (interscan delay 3T1) is 47.4. This corresponds
to a 1.4-fold sensitivity increase compared with the activated 90°
pulse. The relative error is similar to an interscan delay of 5T1 and 3T1 (<6%),
indicating that the interscan delay can be reduced without sacrificing
accuracy. As discussed above, at high acid concentrations, a higher
recycle than 3T1 might be necessary to
compensate for T1 changes.
Figure 4
Effect on sensitivity
of removing the additional 90° pulse
from excitation sculpting pulse sequence and reducing the interscan
delay from 0.72 s (5T1) to 0.43 s (3T1) (green). Comparison with literature sensitivity
in water. A “standardized sensitivity” was calculated
to compare sensitivities measured on different spectrometers (see
main text). Error bars around each point represent the standard deviation
for each triplicate measurement. NH4+: 40 μM,
Gd3+: 1 mM, field strength: 400 MHz.
Effect on sensitivity
of removing the additional 90° pulse
from excitation sculpting pulse sequence and reducing the interscan
delay from 0.72 s (5T1) to 0.43 s (3T1) (green). Comparison with literature sensitivity
in water. A “standardized sensitivity” was calculated
to compare sensitivities measured on different spectrometers (see
main text). Error bars around each point represent the standard deviation
for each triplicate measurement. NH4+: 40 μM,
Gd3+: 1 mM, field strength: 400 MHz.We remeasured the sensitivity gain after addition of 1 mM Gd3+ to obtain a direct measurement of the sensitivity gain without
the interference of the additional 90° pulse. Sensitivity increases
of 3.9- and 3.6-fold are measured with 1 mM Gd3+ for interscan
delays of 5T1 and 3T1, respectively. These values are consistent with the predicted
sensitivity gain from the T1 decrease
(3.9). Taking into account the corrected sensitivity gain that we
calculated from Figure a (3.1), we estimate that the sensitivity can be increased by a factor
of 3.5 ± 0.4 with 1 mM Gd3+, which corresponds to
an order of magnitude less analysis time or several hours less ammonia
accumulation to reach the detection limit. This sensitivity improvement
makes fast 1H NMR NH4+ quantification
accessible with a standard NMR spectrometer and reduces the cost of
essential control experiments with expensive (≈500 euros/L) 15N2.It is difficult to compare the sensitivities
of two different NMR
detection methods if these methods were applied using different spectrometers.
The sensitivity can vary an order of magnitude because of different
field strength, probe hardware, NMR tubes, postprocessing methods,
etc.[18] We attempt to compare our sensitivity
with the sensitivity measured by Hodgetts et al. by calculating a
standardized sensitivity that takes into account the influence of
field strength and type of probe (cryo- or room-temperature probe)
on sensitivity (Figure ). The calculation of the standardized sensitivity can be found in
the SI. As expected, with 1 mM Gd3+, the standardized sensitivity is significantly higher than the value
reported by Hodgetts et al. without Gd3+.
Conclusions
In summary, the 1H NMR analysis time required to quantify
NH4+ in aqueous samples can be reduced by an
order of magnitude by adding 1 mM paramagnetic Gd3+. This
improvement makes 1H NMR NH4+ quantification
more accessible and reduces the cost of control experiments with 15N2, which enables faster, more reliable N2 reduction research. A large reduction of the T1 of NH4+ and MA without significant
line broadening causes the sensitivity increase. The method has very
good linearity (R2 = 0.999) and is accurate
over a wide pH range if the interscan delay is increased to compensate
for small T1 changes.
Materials and
Methods
Materials
14NH4Cl (99.995%), 15NH4Cl (≥98 atom %, 15N ≥
99 % CP), maleic acid (≥99%), and H2SO4 (≥97.5%) were obtained from Sigma-Aldrich. Gadolinium(III)
nitrate hexahydrate (99.9%) was obtained from Fisher Scientific. DMSO-d6 (99.9% D, 0.03% V/V Tetramethylsilan) was
obtained from Cambridge Isotope Laboratories. Ultrapure water was
produced with a Milli-Q Advantage A10 water purification system (resistivity:
18.2 Ω at 25°C).
Sample Preparation
Ammonia standard
solutions (40–500
μM) were prepared fresh daily by adding a suitable amount of
NH4Cl to ultrapure water and performing serial dilutions
to the required standard concentrations. In a typical experiment,
525 μL of NH4+ standard solution was mixed
with 50 μL of 0.5 M H2SO4, 50 μL
of DMSO-d6, 25 μL of 12.5 mM maleic
acid, and 25 μL of 27 mM Gd3+ solution inside a 1.5
mL Eppendorf tube. This solution (600 μL) was transferred into
a 5 mm thin-wall NMR tube (Wilmad). All NH4+ concentrations are reported as concentration in the NMR tube unless
otherwise noted. The NMR tube was closed with Norell Sample Vault
NMR tube caps (Sigma-Aldrich). The tube was cleaned with ultrapure
water and ethanol using an NMR tube cleaner. After cleaning, the NMR
tube was dried at 60°C for 1 h and stored in a dust-free environment.
1H NMR Data Acquisition and Processing
1H NMR spectra were acquired on a 400 MHz pulsed Fourier transform
NMR spectrometer equipped with an autosampler. An autotunable, temperature-regulated
Agilent OneNMR room-temperature probe was used for all measurements.
The temperature was set to 25 °C, and the receiver gain was optimized
automatically. To avoid baseline distortions and low receiver gain,
the water resonance has to be suppressed by a suitable pulse sequence.
Good water suppression was obtained with pulse sequences that use
pulsed field gradients to dephase the water magnetization and selective
pulses to flip the NH4+ magnetization back into
phase during acquisition. Two pulse sequences that were preinstalled
in the software of our NMR system (vNMRj) were used in this work:
Excitation Sculpting (vNMRj: “waterES”) and double pulsed
field gradient spin echo (vNMRj: “selexcit”). The waterES
pulse sequence has the following structure:waterES: G1-P90-G1-d1-P90-G2-S180-P180-G2-G3-S180-P180-G3-aqwhere G1–G3 are the z-gradients of different
strengths, P90 and P180 are hard pulses, and S180 is a selective 180°
pulse. During the acquisition time, only the water resonance is out
of phase, whereas the rest of the spectrum is in phase, leading to
the desired suppression of the water resonance. The block “G1-P90-G1”
dephases residual magnetization prior to the next scan and can be
deactivated to increase sensitivity, as described in the main text.
The z-gradient G1 had a duration of 1.6 ms and a
strength of 1.07 G cm–1. The z-gradients
G2 and G3 had a duration of 1 ms and a strength of 1.7 G cm–1. The 180° selective pulses had the shape “Wsupp”
with a width of 2.5 ms and a power of 13 dB. The selexcit pulse sequence
has the following structure:selexcit: P90-G1-S180-G1-G2-S180-G2-aqwhere G1 and G2 are the z-gradients of different
strengths, P90 and P180 are hard pulses, and S180 is a selective 180°
pulse. During the acquisition time, only the region defined by the
selective 180° pulse is in phase, whereas the rest of the spectrum
is out of phase. The z-gradients G1 and G2 had strengths
of 0.85 and 1.28 G cm–1, respectively, and a duration
of 1 ms. The selective 180° pulse was defined as a “q3”
pulse shape with a width of 5 ms and a power of 0 dB. The position
and width of the selective pulse in the frequency domain were set
to 6.63 ppm and 540 Hz, respectively, so that the pulse is positioned
between the resonances of NH4+ and maleic acid.
The pulse shapes q3 and “Wsupp” that were used to create
the shaped pulses in waterES and selexcit are standard pulse shapes
available in the software package vNMRj. Equivalent pulse shapes should
be available in other software packages.The data were processed
in the software package MestReNova (version:
12.0.1-20560) using the automated tools provided in this software.
Unless otherwise noted, an apodization of 4 Hz was applied followed
by phasing and baseline correction. The peaks of NH4+ (t, ≈6.9 ppm, 4H) and MA (s, ≈6.21 ppm, 2H) were integrated using the line
fitting tool. Using the line fitting tool instead of directly integrating
the peaks leads to an approximately 2-fold decrease of the relative
error. The three integrals of the NH4+ peaks
were added together to calculate the total NH4+ integral. From the ratio of the integral of NH4+ and MA, the concentration of NH4+ was calculated
with absolute quantification according to eq . The linewidth of NH4+ is calculated by averaging the full width at half-maximum (FWHM)
of the three NH4+ peaks. The signal-to-noise
ratio (SNR) was calculated using the “SNR calculation”
tool in MestReNova with the noise region defined from 11 to 13 ppm.
The SNR values were calculated by averaging three measurements of
the average SNR of the three peaks of the NH4+ triplet. The relative error was calculated according towhere ccalcd and cgrav are the concentrations of NH4+ calculated from absolute quantification and from the
weight and purity of the NH4Cl that was added to prepare
the standards, respectively.The T1 values of NH4+ and MA were measured using
the ES pulse sequence with default
setting. Spectra were acquired at six different recycle delays, and
the function y(x) = a*(1 – exp(−bx)) was fitted to the
integrated peak intensities of NH4+ and MA as
a function of d1 using the software OriginPro
2015. Subsequently, the parameter b from the fitting
function was inversed to calculate T1.
An example of the T1 determination using
this method can be found in the SI.