| Literature DB >> 33680550 |
Jintaek Im1, Cheol Song1.
Abstract
We present a SMART injector with two parallel common-path optical coherence tomography fibers to enable angle measurements and injection depth corrections for oblique subretinal injection. The two optical fibers are attached to opposite sides of a 33 G needle with known offsets and designed to pass through a 23 G trocar that has an inner diameter of 0.65 mm. By attaching a SMART system to a rotational stage, the measured angles are calibrated for minimal error from reference angles. A commercial eye model was used to evaluate the control performance, and injection experiments were performed on a phantom made of agarose gel and a porcine eye.Entities:
Year: 2021 PMID: 33680550 PMCID: PMC7901332 DOI: 10.1364/BOE.410492
Source DB: PubMed Journal: Biomed Opt Express ISSN: 2156-7085 Impact factor: 3.732