Oblique injection depth correction by a two parallel OCT sensor guided handheld SMART injector.

Jintaek Im1, Cheol Song1.   

Abstract

We present a SMART injector with two parallel common-path optical coherence tomography fibers to enable angle measurements and injection depth corrections for oblique subretinal injection. The two optical fibers are attached to opposite sides of a 33 G needle with known offsets and designed to pass through a 23 G trocar that has an inner diameter of 0.65 mm. By attaching a SMART system to a rotational stage, the measured angles are calibrated for minimal error from reference angles. A commercial eye model was used to evaluate the control performance, and injection experiments were performed on a phantom made of agarose gel and a porcine eye.
© 2021 Optical Society of America under the terms of the OSA Open Access Publishing Agreement.

Entities:  

Year:  2021        PMID: 33680550      PMCID: PMC7901332          DOI: 10.1364/BOE.410492

Source DB:  PubMed          Journal:  Biomed Opt Express        ISSN: 2156-7085            Impact factor:   3.732