Development of a Portable Respiratory Gas Analyzer for Measuring Indirect Resting Energy Expenditure (REE).

Hyo-Chang Seo1, Daehyeon Shin1, Chae Hun Leem2, Segyeong Joo1.   

Abstract

OBJECTIVE: A rapidly growing home healthcare market has resulted in the development of many portable or wearable products. Most of these products measure, estimate, or calculate physiologic signals or parameters, such as step counts, blood pressure, or electrocardiogram. One of the most important applications in home healthcare is monitoring one's metabolic state since the change of metabolic state could reveal minor or major changes in one's health condition. A simple and noninvasive way to measure metabolism is through breath monitoring. With breath monitoring by breath gas analysis, two important indicators like the respiratory quotient (RQ) and resting energy exposure (REE) can be calculated. Therefore, we developed a portable respiratory gas analyzer for breath monitoring to monitor metabolic state, and the performance of the developed device was tested in a clinical trial. Approach. The subjects consisted of 40 healthy men and women. Subjects begin to measure exhalation gas using Vmax 29 for 15 minutes. After that, subjects begin to measure exhalation gas via the developed respiratory gas analyzer. Finally, the recorded data on the volume of oxygen (VO2), volume of carbon dioxide (VCO2), RQ, and REE were used to validate correlations between Vmax 29 and the developed respiratory gas analyzer.
RESULTS: The results showed that the root-mean-square errors (RMSE) values of VCO2, VO2, RQ, and REE are 0.0315, 0.0417, 0.504, and 0.127. Bland-Altman plots showed that most of the VCO2, VO2, RQ, and REE values are within 95% of the significance level.
CONCLUSIONS: We have successfully developed and tested a portable respiratory gas analyzer for home healthcare. However, there are limitations of the clinical trial; the number of subjects is small in size, and the age and race of subjects are confined. The developed portable respiratory gas analyzer is a cost-efficient method for measuring metabolic state and a new application of home healthcare.
Copyright © 2021 Hyo-Chang Seo et al.

Entities:  

Mesh:

Substances:

Year:  2021        PMID: 33680417      PMCID: PMC7904359          DOI: 10.1155/2021/8870749

Source DB:  PubMed          Journal:  J Healthc Eng        ISSN: 2040-2295            Impact factor:   2.682


  15 in total

1.  Monitoring energy metabolism with indirect calorimetry: instruments, interpretation, and clinical application.

Authors:  Kalman E Holdy
Journal:  Nutr Clin Pract       Date:  2004-10       Impact factor: 3.080

2.  The Effects of Exercises on the Relationship between Pulse Transit Time and Arterial Blood Pressure.

Authors:  Y Wong; Y Zhang
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2005

Review 3.  Indirect calorimetry: technical aspects.

Authors:  L E Matarese
Journal:  J Am Diet Assoc       Date:  1997-10

4.  Indirect calorimetry in humans: a postcalorimetric evaluation procedure for correction of metabolic monitor variability.

Authors:  Peter Schadewaldt; Bettina Nowotny; Klaus Strassburger; Jörg Kotzka; Michael Roden
Journal:  Am J Clin Nutr       Date:  2013-02-27       Impact factor: 7.045

5.  The measurement of energy expenditure.

Authors:  Richard D Branson; Jay A Johannigman
Journal:  Nutr Clin Pract       Date:  2004-12       Impact factor: 3.080

6.  Pulse transit time as an indicator of arterial blood pressure.

Authors:  L A Geddes; M H Voelz; C F Babbs; J D Bourland; W A Tacker
Journal:  Psychophysiology       Date:  1981-01       Impact factor: 4.016

Review 7.  Indirect calorimetry: applications in practice.

Authors:  Jennifer A Wooley
Journal:  Respir Care Clin N Am       Date:  2006-12

8.  Instrumentation for monitoring gas exchange and metabolic rate in critically ill patients.

Authors:  D R Westenskow; C A Cutler; W D Wallace
Journal:  Crit Care Med       Date:  1984-03       Impact factor: 7.598

9.  Clinical use of the respiratory quotient obtained from indirect calorimetry.

Authors:  Stephen A McClave; Cynthia C Lowen; Melissa J Kleber; J Wesley McConnell; Laura Y Jung; Linda J Goldsmith
Journal:  JPEN J Parenter Enteral Nutr       Date:  2003 Jan-Feb       Impact factor: 4.016

10.  Community-Based ECG Monitoring System for Patients with Cardiovascular Diseases.

Authors:  Bor-Shyh Lin; Alice M Wong; Kevin C Tseng
Journal:  J Med Syst       Date:  2016-01-22       Impact factor: 4.460

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.