Literature DB >> 33676070

Adverse events associated with Bonebridge and Osia bone conduction implant devices.

Hannah R Crowder1, Daniel E Bestourous2, Brian K Reilly3.   

Abstract

PURPOSE: Active transcutaneous Bone Conduction Implants (BCIs) are relatively new to the market and may offer improved outcomes while reducing skin-related complications associated with previous models. The purpose of this study is to examine medical device reports (MDRs) submitted to the Food and Drug Administration's (FDA) Manufacturer and User Device Facility Experience (MAUDE) database to identify adverse events with the active, transcutaneous BCIs, Bonebridge and Osia.
METHODS: A search of the FDA MAUDE database was conducted using product code "PFO" (for Active Implantable Bone Conduction Hearing System), brand names "Bonebridge" and "Osia." Data was collected on device malfunction, patient injury, inciting events, and subsequent interventions between July 1, 2018 and November 1, 2020.
RESULTS: The search query yielded 83 reports that met inclusion criteria, 56 regarding Bonebridge and 27 regarding Osia. A total of 91 adverse events were reported, including 45 device malfunctions and 46 patient injuries. Of all adverse events reported for Bonebridge, 15 (26.3%, 15/57) documented patient injuries, while the majority (73.7%, 42/57) documented device malfunctions. Of all adverse events reported for Osia, 3 (8.8%, 3/34) were reported concerning malfunctions, while 31 (91.2%, 31/34) were reported for patient injuries. The most commonly reported adverse events included lack of conduction or hearing (n = 26, 28.6%), infection (n = 14, 15.4%), and intermittent or reduced conduction or hearing (n = 12, 13.2%). From the MAUDE database reported adverse events and the total number of Osia implants given to us from Cochlear over this 28 month period, we estimate patient injuries to occur in roughly 2.1% of patients.
CONCLUSION: There are limitations to the database which make systemic analysis challenging. This study suggests that patients with transcutaneous, active BCIs may be experiencing fewer soft tissue injuries, but similar device malfunctions as those with previous models.
Copyright © 2021 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bone conduction implant; Bonebridge; Hearing aid; Osia; Pediatric otology

Year:  2021        PMID: 33676070     DOI: 10.1016/j.amjoto.2021.102968

Source DB:  PubMed          Journal:  Am J Otolaryngol        ISSN: 0196-0709            Impact factor:   1.808


  3 in total

1.  Clinical Experience of Using Active Transcutaneous Bone Conduction Implants (Bonebridge) in Children Under 5 Years Old.

Authors:  Seong Hoon Bae; Youngrak Jung; Jinsei Jung; Jae Young Choi
Journal:  Clin Exp Otorhinolaryngol       Date:  2022-03-04       Impact factor: 3.340

2.  Long-term data of the new transcutaneous partially implantable bone conduction hearing system Osia®.

Authors:  Ann-Kathrin Rauch; Thomas Wesarg; Antje Aschendorff; Iva Speck; Susan Arndt
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-11-18       Impact factor: 3.236

3.  Pediatric morphometric study to guide the optimized implantation of the Osia® 2 implant system.

Authors:  Balint Posta; Adam Perenyi; Linda Szabo; Roland Nagy; Gabor Katona; Zsuzsanna Csakanyi; Laszlo Rovo; Zsofia Bere
Journal:  Eur Arch Otorhinolaryngol       Date:  2022-03-16       Impact factor: 3.236

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.