Chaitanya Tellapragada1, Christian G Giske2,3.
Abstract
The study was undertaken to evaluate the performance of Unyvero Hospitalized Pneumonia (HPN) panel application, a multiplex PCR-based method for the detection of bacterial pathogens from lower respiratory tract (LRT) samples, obtained from COVID-19 patients with suspected secondary hospital-acquired pneumonia. Residual LRT samples obtained from critically ill COVID-19 patients with predetermined microbiological culture results were tested using the Unyvero HPN Application. Performance evaluation of the HPN Application was carried out using the standard-of-care (SoC) microbiological culture findings as the reference method. Eighty-three LRT samples were used in the evaluation. The HPN Application had a full concordance with SoC findings in 59/83 (71%) samples. The new method detected additional bacterial species in 21 (25%) and failed at detecting a bacterial species present in lower respiratory culture in 3 (3.6%) samples. Overall the sensitivity, specificity, positive, and negative predictive values of the HPN Application were 95.1% (95%CI 96.5-98.3%), 98.3% (95% CI 97.5-98.9%), 71.6% (95% CI 61.0-80.3%), and 99.8% (95% CI 99.3-99.9%), respectively. In conclusion, the HPN Application demonstrated higher diagnostic yield in comparison with the culture and generated results within 5 h.Entities:
Keywords: Bacterial pneumonia; COVID-19; Hospital-Acquired pneumonia; Multiplex-PCR; Rapid-diagnostic tests
Mesh:
Year: 2021 PMID: 33661410 PMCID: PMC7930892 DOI: 10.1007/s10096-021-04194-6
Source DB: PubMed Journal: Eur J Clin Microbiol Infect Dis ISSN: 0934-9723 Impact factor: 3.267