Evaluating drug sensitivity is improved by directly quantifying death kinetics, rather than correlates of viability, such as metabolic activity. This is challenging, requiring time-lapse microscopy and genetically encoded labels to distinguish live and dead cells. Here, we describe fluorescence-based and lysis-dependent inference of cell death kinetics (FLICK). This method requires only a standard fluorescence plate reader, retaining the high-throughput nature and broad accessibility of common viability assays. However, FLICK specifically quantifies death, including an accurate inference of death kinetics. For complete details on the use and execution of this protocol, please refer to Richards et al. (2020).
Evaluating drug sensitivity is improved by directly quantifying death kinetics, rather than correlates of viability, such as metabolic activity. This is challenging, requiring time-lapse microscopy and genetically encoded labels to distinguish live and dead cells. Here, we describe fluorescence-based and lysis-dependent inference of cell death kinetics (FLICK). This method requires only a standard fluorescence plate reader, retaining the high-throughput nature and broad accessibility of common viability assays. However, FLICK specifically quantifies death, including an accurate inference of death kinetics. For complete details on the use and execution of this protocol, please refer to Richards et al. (2020).
Timing: 1 dayIn two 96-well, optical bottom, black-walled plates, add to the first column of both plates 40,000 cells in 180 μL of cell culture media (16 wells total). This first well will be used to generate a 2-fold cell dilution across the plate. The final concentration for this first well will be 20,000 cells/well (i.e., 20,000 cells per 90 μL).This protocol has been written for use with a Tecan Spark multimode plate reader, with fluorescence readings taken from the bottom of the plate. Optical bottom plates are not necessary if the plate reader used acquires fluorescence emission from the top of the wells. However, even in this case, optical bottom plates may still be helpful for visualizing cell lysis (see Before you begin, Step 8). Additionally, this protocol has been optimized for U2-OS cells. Testing an altered range of cell densities may be optimal for other types of cells.CRITICAL: Media containing phenol red has an insignificant effect on the SYTOX fluorescence signal or signal linearity. Other cell line-specific media reagents should be checked for auto-fluorescence at 504/523 nm excitation and emission wavelengths.Add 90 μL of media to all remaining wells in both plates (176 wells total).Serially dilute cells 1:2 from the first column of the plates into the adjacent columns by removing 90 μL and adding it to the adjacent column to the right. Continue the dilution until reaching the 11th column. The last column should not contain any cells on either plate and is used for determining background fluorescence signal.Discard 90 μL from the 11th column wells.At this point all wells should have 90 μL of media. The first column should have 20,000 cells; the 11th column should have approximately 20 cells; the last column should contain only media (Figure 1A).
Figure 1
Optimizing SYTOX dyes for measuring cell death in a fluorescence plate reader
(A) Plating design for establishing a linear range and acquisition settings for SYTOX fluorescence. SYTOX dilution represented using green shades. Cell dilution represented using blue shades.
(B) Visual inspection of cell permeabilization by Triton. Phase and SYTOX fluorescence images shown for cells exposed to 0.1% Triton-X. Images collected using a 10× objective. Scale bar applies to all images, which are at the same magnification. Cell boundary for the highlighted cell was added for illustrative purposes and is not part of the analysis. Inset image of the highlighted cell is a cropped 2× version of the same image.
(C) Optimal permeabilization times and degree of signal stability after cell lysis was determined by measuring SYTOX fluorescence at varied times after cell permeabilization. Data are mean fluorescence in the well ±SD, with data from 4 individual replicate wells overlayed.
(D) Establishing acquisition settings for SYTOX fluorescence. Linearity of SYTOX fluorescence evaluated at varied fluorescence gain settings (90–170). SYTOX signal at a gain of 140 highlighted in black. Data are mean ±SD of replicates following a 2 h Triton permeabilization with 5 μM SYTOX green, using the plating design shown in (A).
Incubate plates at 37°C and 5% CO2 for >4 h until the cells are adhered. Incubation time may be longer or shorter depending on cell line of interest.Prepare SYTOX reagent at 10× final concentration.Prepare 10× SYTOX solutions in media at the following concentrations: 50, 40, 30, and 20 μM.Add 10, 8, 6, or 4 μL of 5 mM SYTOX green to 1 mL of culturing media to create the 50, 40, 30, and 20 μM 10× stock solutions, respectively.SYTOX is used in this protocol to label dead cells. SYTOX fluorescence increases upon binding to DNA, and fluorescence depends on loss of plasma membrane integrity. This general mechanism is beneficial, as SYTOX can be used to evaluate the degree of cell death independently from the mechanism of cell death. The FLICK method we describe for inferring death kinetics can be used with any dye that labels dead cells independent of mechanism. Other dyes that label dead cells in a death mechanism-specific manner (e.g., apoptotic specific) may be valuable for quantifying the kinetics of activation for specific death mechanisms.Prepare 10× Triton-X permeabilization buffer.Make a 10× Triton-X solution in cell culture grade PBS buffer by adding 50 μL Triton-X to 5 mL PBS (1.0% solution). Mix the solution by inverting the tube. Avoid shaking.To help dissolve Triton-X, the solution can be put in a 37°C water bath for 5 min. The final concentration of Triton-X when added to cells can be between 0.05%–0.15%. The 10× solution should adjusted accordingly if using a final concentration other than 0.1%.Add 10× SYTOX solutions (Figure 1A).Each 10× SYTOX solution will be added to 4 rows of cells creating quadruplicate measurements for each dilution (Figure 1A).To the first 4 rows on plate 1, add 10 μL of 50 μM 10× SYTOX solution from step 5 for a final concentration of 5 μM.To the last 4 rows on plate 1, add 10 μL of 40 μM 10× SYTOX solution from step 5 for a final concentration of 4 μM.To the first 4 rows on plate 2, add 10 μL of 30 μM 10× SYTOX solution from step 5 for a final concentration of 3 μM.To the last 4 rows on plate 2, add 10 μL of 20 μM 10× SYTOX solution from step 5 for a final concentration of 2 μM.These data will be used to optimize the concentration of SYTOX and the fluorescence acquisition settings on the plate reader. The technical replicates are not intended to capture other sources of biological or experimental variation. See Optimizing Experimental Design, step 14 for details on optimizing plating designs.Permeabilize cells (Figures 1B and 1C).Add 10 μL of 1% Triton-X solution prepared in step 6 to all wells for a final concentration of 0.1%.Place plates in incubator at 37°C and 5% CO2 for approximately 1.5–2 h to allow for cell permeabilization.CRITICAL: The timing for permeabilization will vary depending on cell line and Triton-X concentration. It may be necessary to incubate longer to ensure complete cell lysis. Check for cell permeabilization under a microscope (Figure 1B). Following lysis, cell morphology should be significantly altered. No discernable intact cells should remain in the well, with only cell debris remaining (see Figure 1B, 2 h). A key observation is the change in nuclear morphology and lack of clear nuclear boundary (see Figure 1B, change between 1 and 2 h). Even following dramatic changes to nuclear morphology, total well fluorescence is largely unaffected (Figure 1C). If fluorescence and/or phase microscopy is not available, permeabilization time can be optimized by reading the plate over time, to determine when the signal hits a stable plateau.Set up the plate reader for quantification of SYTOX fluorescence.The excitation and emission wavelengths should be set to 504/523 nm with a bandwidth of 5 nm, if possible.Flash number should be set at 5.The settings above are specified for the Tecan Spark. Settings will vary based on the plate reader. Some models may allow control of fluorescence integration time rather than flash number. Integration time should be kept short to minimize photobleaching.CRITICAL: Once cells are lysed, the SYTOX signal is stable (Figure 1C). Avoid shaking or agitating the plate at this step, which can result in dislodging/aggregation of the cells, leading to inaccurate readings in some plate readers (Grootjans et al., 2016).Collect SYTOX fluorescence using range of gain settings (Figure 1D).For each of the four SYTOX concentrations, collect fluorescence data using a spectrum of gain values. Gain settings should include levels that produce low signals at all cell numbers, through gain settings that produce saturated fluorescence signals in wells containing high cell numbers. Saturated signals are reported as “invalid” (INV) or “NaN” in most plate readers, and refer to signals that are above the maximum detectable limit (65,000 fluorescence units for the Tecan Spark). Refer to the plate reader manual to determine the maximum fluorescence signal that can be detected.Data analysis to determine optimal acquisition settings.Remove the background fluorescence from each measurement by calculating the mean signal from wells that do not contain cells; subtract this mean value from the signal for all wells.For each SYTOX concentration and gain condition, compare the background subtracted fluorescence signal to the cell number (Figure 1D).Determine the correlation coefficient (R2) for each condition using linear regression.In MATLAB, the built-in function “corrcoef” can be used to compute the Pearson correlation coefficient (R).In Excel, correlation coefficient can be calculated using the function “CORREL.”Select the SYTOX concentration and gain setting that gives the best combination of linearity and dynamic range.The maximum signal (wells that contain 20,000 cells) should be less than 80%–90% of the detection limit (65,000 relative fluorescence units on the Tecan Spark).In most titration experiments a range of gain values produce large dynamic range and R2 > 0.99 (Figure 1D, gain values between 120–140 are all acceptable). See Table 1 for example settings used for other cell lines.
Table 1
Example SYTOX calibration settings (all using 5μM SYTOX green)
Cell
Permeabilization time (h)
Gain
A549
2
95
BT-20
3
130
HeLA
2
95
MCF10A
2
130
MCF7
2
95
MDA-MB-231
4
120
MDA-MB-468
3
130
PC9
3
120
Primary Human Fibroblast
3
140
T47D
4
130
U2-OS
2
130
Example SYTOX calibration settings (all using 5μM SYTOX green)Optimizing SYTOX dyes for measuring cell death in a fluorescence plate reader(A) Plating design for establishing a linear range and acquisition settings for SYTOX fluorescence. SYTOX dilution represented using green shades. Cell dilution represented using blue shades.(B) Visual inspection of cell permeabilization by Triton. Phase and SYTOX fluorescence images shown for cells exposed to 0.1% Triton-X. Images collected using a 10× objective. Scale bar applies to all images, which are at the same magnification. Cell boundary for the highlighted cell was added for illustrative purposes and is not part of the analysis. Inset image of the highlighted cell is a cropped 2× version of the same image.(C) Optimal permeabilization times and degree of signal stability after cell lysis was determined by measuring SYTOX fluorescence at varied times after cell permeabilization. Data are mean fluorescence in the well ±SD, with data from 4 individual replicate wells overlayed.(D) Establishing acquisition settings for SYTOX fluorescence. Linearity of SYTOX fluorescence evaluated at varied fluorescence gain settings (90–170). SYTOX signal at a gain of 140 highlighted in black. Data are mean ±SD of replicates following a 2 h Triton permeabilization with 5 μM SYTOX green, using the plating design shown in (A).
Optimize experimental design
Timing: 1 dayOption 1 – Set up a T0 control plate to determine live and dead cell numbers at assay start.When designing an experiment to calculate lethal fraction over time, a T0 control plate should be used to estimate the initial lethal fraction, the proportion of cells that are alive versus dead, at the start of the assay. This plate is seeded at the same time as experimental plates with the same number of cells, but will be lysed at the start of the experiment. An entire 96 well plate should be used for this calculation (see Quantification and statistical analysis step 2).Option 2 – Estimate the live and dead cells at assay start from previous experiments.Fluorescence readings from a previous experiment can be used as an approximation to determine the initial cell counts if the same number of cells were seeded in both experiments.CRITICAL: Do not lyse a subset of wells on an experimental plate as the T0 control. The lysis buffer can permeate to other wells across the plate over time, causing lysis.Consider the optimal plating layout (Figures 2A and 2B).
Figure 2
Optimal plating designs for analysis of drug sensitivity using common pharmacometrics
(A and B) Schematics of 96-well plating layouts containing drugs at varied doses (graded colors), vehicle controls (gray), and untreated wells (white). (A) Edge wells are often omitted due to different growth characteristics. Vehicle controls arrayed systematically across the plate can assist with identifying/normalizing plating biases if needed. This standard design has no technical replicates and biological replicates located on separate plates. (B) Partial use of edge wells, with high doses of drug on one edge, and vehicle controls across top edge and two columns of the plate. Using this design, normalization to vehicle controls should be performed using the plate mean across all vehicle wells to minimize edge effects.
(C) Variable influence of plating design, dependent on choice of downstream analysis metrics. Data shown for relative viability (RV or RV edge; defined as live cells in treated wells compared to control) or Fractional Viability (FV or FV edge; defined as live cells in each well compared to total of live and dead cells in the same well). Data are for DMSO treated wells using a plating design, similar to (A). For RV, DMSO treated wells were normalized relative to the plate average for all DMSO treated wells. For RV edge, DMSO treated wells were normalized relative to the average of all untreated wells, which were located on the plate edge. FV refers to the fraction of live cells in DMSO treated wells. FV edge refers to the fraction of live cells in edge wells. Use of edge wells in normalization increases variability of RV but not FV metrics. Data are shown for 150 individual wells with median values highlighted with a black line.
Evaporation and slow growth often occur on the outer edges of the plate. We have observed that drug-induced lethal fraction is not generally altered by growth rate. Additionally, because lethal fraction is internally controlled (i.e., does not require comparison between two wells), the outer edges can be used if downstream analyses focus exclusively on lethal fraction. However, one should expect lower total cell counts in these wells. Relative viability, which requires comparison between separate wells on a plate should not be calculated using the edge wells (Figure 2C).Other sources of variability in this assay could include inconsistencies in cell plating between plates or on separate days. Additionally, variations in Triton permeabilization efficiency may exist for plates that are lysed at different times or from different permeabilization buffers. The plating design should avoid confounding theses sources of variation with treatment groups.Optimal plating designs for analysis of drug sensitivity using common pharmacometrics(A and B) Schematics of 96-well plating layouts containing drugs at varied doses (graded colors), vehicle controls (gray), and untreated wells (white). (A) Edge wells are often omitted due to different growth characteristics. Vehicle controls arrayed systematically across the plate can assist with identifying/normalizing plating biases if needed. This standard design has no technical replicates and biological replicates located on separate plates. (B) Partial use of edge wells, with high doses of drug on one edge, and vehicle controls across top edge and two columns of the plate. Using this design, normalization to vehicle controls should be performed using the plate mean across all vehicle wells to minimize edge effects.(C) Variable influence of plating design, dependent on choice of downstream analysis metrics. Data shown for relative viability (RV or RV edge; defined as live cells in treated wells compared to control) or Fractional Viability (FV or FV edge; defined as live cells in each well compared to total of live and dead cells in the same well). Data are for DMSO treated wells using a plating design, similar to (A). For RV, DMSO treated wells were normalized relative to the plate average for all DMSO treated wells. For RV edge, DMSO treated wells were normalized relative to the average of all untreated wells, which were located on the plate edge. FV refers to the fraction of live cells in DMSO treated wells. FV edge refers to the fraction of live cells in edge wells. Use of edge wells in normalization increases variability of RV but not FV metrics. Data are shown for 150 individual wells with median values highlighted with a black line.The T0 control plate can be used to correct for the growth variability if necessary. This should only be done if users have validated that a systematic plating location bias reproducibly occurs on a given plate reader.Delivering compound and SYTOX reagent.When possible, 10× compound stocks should be made in media containing 10× SYTOX. This allows the operator to dispense both reagents into the well in a single pipette stroke. The cell plating volume should be adjusted so a final volume of 100 μL in the well is achieved, after adding SYTOX and compound.100 μL final volume is used in order to limit the amount of reagents needed, and because U2-OS cells grow at expected rates when plated at these densities in this volume. Plating volumes can be adjusted if needed, based on the growth characteristics of the cells in the study.CRITICAL: At high concentrations of drug, the percentage of DMSO may be above the tolerance of cells. It is important to have vehicle controls on the plate to determine the contribution of death from DMSO.
Key resources table
Materials and equipment
This protocol uses a Tecan Spark Multimode plate reader. Essentially any other plate reader should also be suitable. The minimal requirement is only the ability to quantify a single channel of fluorescence associated with labeling of dead cells. Labeling of dead cells in this protocol is performed using the SYTOX green dye. SYTOX dyes in other colors are also suitable, assuming these can be read in the chosen plate reader. SYTOX dyes are used in this protocol as their fluorescence is increased when bound to DNA in dead cells, and these dyes are not toxic to live cells, enabling kinetic analysis. Other dyes with these features should also be usable with appropriate titration.
Step-by-step method details
Plating cells
Timing: 20 minFollowing experimental design, cells should be grown and collected to be plated into 96 well, optical bottom, black-walled plates. The number of plates needed is dictated by the number of conditions to be tested; however, an additional T0 control plate must also be plated. Care should be taken in accurately counting and seeding cells. This will reduce noise in the fluorescence measurements and provide consistent data between replicates. Two technical replicates are performed on the same plate to account for plating bias. Additionally, biological replicates should also be performed on separate plates or experiments performed on subsequent days.Cell line of interest should be cultured according to established protocols.This protocol is suitable for cells with any growth morphologies or characteristics. Very small cells or cells that grow with uneven density (clumpy or in colonies) will generate noisier data, particularly at low cell densities. This may compromise the low end sensitivity of the assay.Count cells using a hemocytometer or automatic cell counter.Dilute cells to the appropriate concentration in cell culture media that allows for the proper seeding density. Cells should be seeded at a starting density at which untreated cells do not become confluent during the assay, taking into account the doubling time, well size, and length of the experiment. Following counting, cells should be resuspended at a concentration that allows for plating at 90 μL/well.Resuspend cells by gently pipetting up and down, then add 90 μL of cell suspension to plates.See step 14 in Before you begin and Figure 2 for optimal experimental designs, including number of technical and/or biological replicates per experiment.Place in incubator at 37°C and 5% CO2 to adhere overnight (4–18 h).Pause point: The assay can begin immediately after the cells have adhered to the plate. Most commonly, drugs are added the following morning after plating, but the assay could also start anytime within the following 2–3 days while the wells are not confluent and the cells are growing normally. Whatever decision is made should be consistent across all plates in a given experiment.
Adding SYTOX and compound
Timing: 1 hSYTOX green solutions with and without compound are prepared and added to experimental wells. A 1% Triton-X solution is prepared and added to the T0 control plate to determine initial total cell fluorescence.Prepare a 1% Triton-X permeabilization buffer in sterile PBS.Approximately 1 mL is required to permeabilize an entire 96-well plate. Extra should be made to account for dead volume, particularly if dispensed from a reservoir using a multi-channel pipette.Prepare a 10× SYTOX solution (see Before you begin step 11) by diluting 5 mM SYTOX in enough volume of culture media for your experimental design. The total volume should be enough for all dilutions, experimental wells, and T0 control plate.Dilute the compounds of interest to 10× the final desired concentration in 10× SYTOX solution. This should also be done for a vehicle control. If doing dose-response experiments, it is recommended to create 10× drug dilutions in round-bottom plates by serial dilution.Round-bottom plates are recommended to reduce the dead volume during pipettingAdd 10 μL of the SYTOX+compound (or vehicle) solution to appropriate wells. Add 10 μL of 10× SYTOX only solution to the T0 control plate. The final volume of the well should now be 100 μL with 1× SYTOX and 1× compound (or vehicle).Add 10 μL of 1% Triton-X permeabilization buffer to all wells of the T0 plate and place in the incubator at 37°C and 5% CO2 for >2 h (or the length of time optimized in Before you begin, step 8).
Fluorescence measurements
Timing: 1–3 daysThis step quantifies the raw fluorescence of dead cells at chosen time points. After the final measurement is taken, cells are lysed using Triton-X buffer to quantify the total cell fluorescence at the end of the assay.Set the fluorescence plate reader to the appropriate excitation, emission, and gain settings for the chosen SYTOX concentration.Take an initial reading for the experimental plates. Once the measurement is finished, keep the plate stored in the incubator at 37°C and 5% CO2 until the next time point.Once the T0 control plate is fully permeabilized (step 10), measure the fluorescence. This measurement of Total Starting Fluorescence will be used later in Quantification and statistical analysis, step 2. Discard the plate when finished.Continue taking measurements at selected time points until the end of the experiment.Common assay lengths are generally 48–72 h after drug addition, but these depend on the growth rate of cells being tested.Time points should be ∼ 4 h apart, at a roughly constant interval, and selected to capture multiple (2–3) time points within the increasing phase of the response and the plateau phase of the response. Selected time points should be the same across biological replicates.Both the frequency and the interval of measurements will affect the accuracy of the kinetic fitting (Quantification and statistical analysis, step 3). In particular, the inference of death onset time (DO) depends on capturing changes in cell death within both the increasing phase and plateau phase of the response. See Troubleshooting for more details. Plates should remain in the incubator between time point measurements unless the plate reader used is equipped with a temperature and CO2 controlled environmental chamber. Temperature fluctuations caused by frequent removal from the incubator may compromise cell health and/or growth rate.Prior to reading the final time point, prepare fresh 1% Triton-X permeabilization buffer. Prepare enough volume to lyse all experimental wells.After the final measurement, add 10 μL of 1% permeabilization buffer to each well. Lyse the plate for > 2 h at 37°C and 5% CO2Once all wells have been fully lysed, take a final fluorescence measurement. This final measurement is proportional to the total cell count at the end of the assay.CRITICAL: Before taking the total cell count measurement, check the plate for complete lysis under a microscope. See Before you begin, step 8, and Figures 1B and 1C for more details.
Expected outcomes
The FLICK method is designed to provide a specific measurement of cell death, rather than an indirect surrogate of cell viability. Other assays that are specific to cell death generally use time-lapse microscopy to track live and dead cells over time. In the FLICK method, this is achieved by direct measurement of the dead cell population over time, and a computational inference of the live cell population. Additionally, because FLICK does not require the use of any genetically encoded labels, this method can be used more flexibly, to quantify responses in diverse settings, including for primary cells.The outlined procedure will generate temporally resolved fluorescence data that is proportional to the number of dead cells contained within a well (Figure 3A). Additionally, this protocol generates an experimentally determined measurement that is proportional to the total number of cells (e.g., live and dead) at the start and end of the assay. Total cell fluorescence at intermediate time points (i.e., any time point between the beginning and end of the assay) is computationally inferred using the experimentally observed cell fluorescence at the start and end of the assay as constraints (see Quantification and statistical analysis, Figure 3B). For all wells, fluorescence following endpoint permeabilization should be equal to- or higher than the T0 fluorescence (Figures 4A and 4B, see Troubleshooting section for additional details). For vehicle-treated wells, fluorescence measurements after endpoint permeabilization should be substantially higher than measurements taken at the end point of the assay prior to permeabilization, reflecting an expected low percentage of dead cells in the population (Figures 4B and 4C). Outcomes for drug-treated wells will vary depending on the concentration and mechanism of action (death, growth arrest, or a combination of death and arrest) (Schwartz et al., 2020). Death-inducing drugs should cause an increase in the fluorescence signal over time and may plateau at later time points. Alternatively, drugs that primarily induce growth arrest may not cause an increase in SYTOX fluorescence, but a noticeable decrease in the total cell fluorescence after permeabilization should be observed.
Figure 3
Overview of the FLICK assay
(A) Schematic of steps for data collection using FLICK.
(B) Schematic of steps for analysis of data generated using FLICK.
Figure 4
Kinetic analysis of drug response using the FLICK assay
Analysis of response to SGI-1027 at varied doses. Due to varied rates of death onset, SGI-1027 results in cell death that appears to peak at intermediate doses.
(A) Quantification of dead cell fluorescence over time.
(B) Quantification of live cell fluorescence at assay endpoint. The live cell fluorescence corresponds to the difference between the SYTOX fluorescence pre-Triton permeabilization (PRE) and post-Triton (POST) permeabilization.
(C) Population size at any intermediate time point is estimated based on the experimentally observed population size at the beginning and end of the assay.
(D) Lethal fraction (LF) kinetics for SGI-1027. Unlike dead cell fluorescence, LF reveals dose-dependent increase in drug efficacy.
(E and F) Kinetic parameters from the LED analysis. (E) Death onset time (DO) at each tested dose of SGI-1027. Camptothecin shown for comparison. (F) Death rate (DR) at each tested dose of SGI-1027. Camptothecin shown for comparison. For (A) and (D)–(F), data are mean ± SD of 4 biological replicates. For (F) and (G), dose ranges tested were half-log dilutions starting at 31.6 μM. If death was not significantly above baseline values at the assay endpoint, rates were not calculated (n.a., not applicable).
Overview of the FLICK assay(A) Schematic of steps for data collection using FLICK.(B) Schematic of steps for analysis of data generated using FLICK.Kinetic analysis of drug response using the FLICK assayAnalysis of response to SGI-1027 at varied doses. Due to varied rates of death onset, SGI-1027 results in cell death that appears to peak at intermediate doses.(A) Quantification of dead cell fluorescence over time.(B) Quantification of live cell fluorescence at assay endpoint. The live cell fluorescence corresponds to the difference between the SYTOX fluorescence pre-Triton permeabilization (PRE) and post-Triton (POST) permeabilization.(C) Population size at any intermediate time point is estimated based on the experimentally observed population size at the beginning and end of the assay.(D) Lethal fraction (LF) kinetics for SGI-1027. Unlike dead cell fluorescence, LF reveals dose-dependent increase in drug efficacy.(E and F) Kinetic parameters from the LED analysis. (E) Death onset time (DO) at each tested dose of SGI-1027. Camptothecin shown for comparison. (F) Death rate (DR) at each tested dose of SGI-1027. Camptothecin shown for comparison. For (A) and (D)–(F), data are mean ± SD of 4 biological replicates. For (F) and (G), dose ranges tested were half-log dilutions starting at 31.6 μM. If death was not significantly above baseline values at the assay endpoint, rates were not calculated (n.a., not applicable).Drug-induced cell death can be observed directly from the SYTOX fluorescent signal, which is proportional to the number of dead cells; however, this signal alone should not be used to evaluate drug response, due to the lack of insight about population size (Forcina et al., 2017). The peculiar concentration dependence of SGI-1027 provides an example of unreliable insights derived from counting dead cells alone, when population size is not also considered. SGI-1027 is a DNA methyl-transferase inhibitor which leads to both growth arrest and cell death, but with varied timing and varied intensity, depending on the dose used. At low concentrations (1 μM), SGI-1027 causes low levels of cell death (Figure 4A). At these low concentrations, however, the total cell count at the end of the experiment is reduced compared to the vehicle control, revealing a modest anti-proliferative effect at this concentration (Figures 4B and 4C). At higher concentrations (3.16 and 31.6 μM), all cells are killed by the end of the assay; however, the data do not follow an expected dose-dependent response. Instead, SGI-1027 at 3.16 μM kills a larger total number of cells than at a higher dose of 31.6 μM (Figure 4A). This discrepancy is caused by the fact that exposure to 31.6 μM SGI-1027 results in a very fast onset time of death (DO) and rate of death (DR), essentially killing all plated cells before any cell division can occur. Computing the drug-induced lethal fraction – which considers both the number of dead cells and the population size – reveals a more reliable insight about the drug response, including the dose-dependent variation in death onset time (Figures 4D–4F). Additionally, lethal fraction or its inverse, fractional viability, also reveals a more typical dose-dependent increase in cell death (Figure 4D, 5A, and 5B).
Figure 5
Pharmacological analyses of drug response
Comparison of common pharmaco-metrics.
(A) Definitions for fractional viability (FV), relative viability (RV), and normalized growth rate inhibition (GR). FV is the fraction of a cell population that is alive. RV is the size of the drug-treated cell population, compared to the vehicle control population. GR is the relative population growth rate of drug-treated cells, compared to control.
(B–D) Sample data for response to SGI-1027 (top) or Camptothecin (bottom) in U2-OS cells. (B) FV. (C) RV (D) GR. For each panel in (B)–(D), the IC50, or equivalent, is shown. IC50 refers to the dose associated with 50% reduction from the starting value. Curve fitting generates the EC50 parameter, which refers to the dose of the observed half-maximal response. Data are mean ± SD of biological replicates.
Pharmacological analyses of drug responseComparison of common pharmaco-metrics.(A) Definitions for fractional viability (FV), relative viability (RV), and normalized growth rate inhibition (GR). FV is the fraction of a cell population that is alive. RV is the size of the drug-treated cell population, compared to the vehicle control population. GR is the relative population growth rate of drug-treated cells, compared to control.(B–D) Sample data for response to SGI-1027 (top) or Camptothecin (bottom) in U2-OS cells. (B) FV. (C) RV (D) GR. For each panel in (B)–(D), the IC50, or equivalent, is shown. IC50 refers to the dose associated with 50% reduction from the starting value. Curve fitting generates the EC50 parameter, which refers to the dose of the observed half-maximal response. Data are mean ± SD of biological replicates.Lethal fraction or fractional viability produce reliable insights regarding the degree of drug-induced cell killing. However, these measures do not capture drug-induced changes in the population size. Alternatively, relative viability, the most commonly used measurement for evaluating drug sensitivity, ignores the degree of cell killing, instead reporting only changes in population size (Figures 5A and C). Importantly, because this protocol facilitates quantification of both live and dead cells, these methods can be used to quantify any commonly used drug response metric, including relative viability or the normalized growth rate inhibition value (GR) (Figures 5A and 5D). Drugs vary in the degree to which they activate cell death versus inhibit growth (Schwartz et al., 2020). For drugs like SGI-1027, that primarily activate death without strongly inhibiting cell proliferation, FV, RV, and GR metrics will produce very similar insights (Figures 4B–4D). Alternatively, other drugs like camptothecin activate cell death at high doses, but also strongly inhibit cell proliferation, even at low doses. For these drugs, FV, RV, and GR metrics tend to produce different insights (Figures 4B–4D). Each of these metrics can be appropriate, depending on the focus of a study, but each metric produces unique and incomplete insights. Additionally, using data generated from this protocol, the complementary insights derived from cell death-focused or population size-focused pharmacometrics can be integrated by calculating drug GRADE (Schwartz et al., 2020). GRADE analysis reveals the degree of drug-induced cell death versus drug-induced proliferative arrest induced by a given drug. Other common assays generate signals proportional to metabolic activity (CellTiter-Glo, MTT, Alomar Blue). Unlike FLICK, these other assays do not provide specific information on the degree of drug-induced cell death, and can only be used to score relative viability.
Quantification and statistical analysis
Quantify the fluorescence associated with dead and live cells over time
Determine the SYTOX fluorescence of dead cells (FDt) at each time point for each condition (Figure 4A)SYTOX fluorescence is proportional to the number of dead cells and should be measured immediately after the assay start, followed by measurements at any additional timepoints (t) of interest (Table 2).
Table 2
Fluorescence of dead cells (FDt)
Camptothecin dose (μM)
Replicate #
SYTOX fluorescence (relative fluor. units, RFU)
FD0
FD16
FD20
FD24
FD40
FD44
FD48
0
1
234
236
251
247
325
332
354
0.01
1
239
273
264
269
341
349
405
0.0316
1
253
252
258
267
335
346
351
0.1
1
224
253
259
261
387
445
388
0.316
1
238
265
277
299
407
471
501
1
1
242
371
593
856
1,058
1,222
1,364
3.16
1
257
647
1,324
1,765
2,630
3,169
3,630
10
1
239
502
1,142
1,862
3,201
3,457
3,820
0
2
226
236
253
258
347
385
439
0.01
2
230
279
246
257
344
388
429
0.0316
2
246
254
251
266
347
361
389
0.1
2
221
232
245
253
304
340
358
0.316
2
267
268
282
282
392
550
515
1
2
250
352
567
776
1,128
1,393
1,674
3.16
2
238
600
1,255
1,688
2,659
3,253
3,819
10
2
262
604
1,317
2,154
3,308
3,836
4,131
Fluorescence of dead cells (FDt)Model the population growth over time (Figures 4B and 4C).After the last time point has been taken, determine the SYTOX fluorescence of total cells (FT) at the end of the experiment by measuring the fluorescence following permeabilization with Triton (Figure 4B and Table 3, FT48).
Table 3
Total cell fluorescence at assay endpoint (FT48)
Camptothecin dose (μM)
Rep. #
SYTOX fluorescence of total cells (RFU)
FTM0
FT16
FT20
FT24
FT40
FT44
FT48
0
1
–
–
–
–
–
–
14,747
0.01
1
–
–
–
–
–
–
14,713
0.0316
1
–
–
–
–
–
–
12,829
0.1
1
–
–
–
–
–
–
8,507
0.316
1
–
–
–
–
–
–
6,411
1
1
–
–
–
–
–
–
6,828
3.16
1
–
–
–
–
–
–
4,376
10
1
–
–
–
–
–
–
4,809
0
2
–
–
–
–
–
–
17,605
0.01
2
–
–
–
–
–
–
15,027
0.0316
2
–
–
–
–
–
–
11,393
0.1
2
–
–
–
–
–
–
7,935
0.316
2
–
–
–
–
–
–
6,011
1
2
–
–
–
–
–
–
6,648
3.16
2
–
–
–
–
–
–
4,462
10
2
–
–
–
–
–
–
4,636
Fluorescence values following Triton permeabilization are proportional to the total number of cells (live + dead).Total cell fluorescence at assay endpoint (FT48)Determine the SYTOX fluorescence following Triton permeabilization for the T0 Control plate (FT0).An average starting value (FTM0) should be determined by calculating the 50% trimmed mean from the fluorescence of permeabilized cells (Table 4). The trimmed mean refers to the mean of the data, following removal of the highest and lowest outliers.
Table 4
Fluorescence of total population at assay start (FT0)
SYTOX fluorescence of total population at T0 (RFU)
FT0
3,197
3,470
3,475
3,685
3,700
3,883
3,927
3,988
…
…
…
…
…
…
…
…
3,994
4,316
4,320
4,519
5,021
5,814
6,359
6,477
50% trimmed mean (FTM0)
4,081
In MATLAB, use the built-in function “trimmean” (i.e., trimmean(x,50)).In Excel or similar, sort the T0 data and remove the top and bottom 25%. Compute the mean of the remaining 50%.Fluorescence of total population at assay start (FT0)Computing an average starting fluorescence value is done here for robustness, as well-to-well variations are likely to fluctuate from plate to plate. Computing this average from a trimmed portion of the data may reduce the influence of single outlier wells that may exist for technical reasons (incomplete lysis, error in plating, etc.). A similar method is used in other studies (Hafner et al., 2016).For each condition, fit the starting (FTM0) and final (FT48) total cell fluorescence values to an exponential curve:t = time pointk = growth rate per houra = initial starting populationIn MATLAB, this can be performed using the built-in function “fit” and the fit type “exp1.”In PRISM, this can be performed using the nonlinear regression curve fitting package, with the exponential growth equation.In excel, this can be performed by generating a signal vs. time scatter plot, and adding an exponential trendlineEstimate the total fluorescence at each intermediate time point (FTt) by solving the above equation at all desired time points (t), once the a and k parameters have been solved from the curve fitting (Figure 4C and Table 5).
Table 5
Fluorescence of the total population (FTt)
Camptothecin dose (μM)
Rep. #
SYTOX fluorescence of the total population (RFU)
FTM0
FT16
FT20
FT24
FT40
FT44
FT48
0
1
4,081
6,262
6,970
7,758
11,905
13,250
14,747
0.01
1
4,081
6,258
6,963
7,749
11,882
13,222
14,713
0.0316
1
4,081
5,978
6,577
7,236
10,600
11,661
12,829
0.1
1
4,081
5,213
5,542
5,892
7,527
8,002
8,507
0.316
1
4,081
4,744
4,926
5,115
5,946
6,174
6,411
1
1
4,081
4,845
5,057
5,279
6,267
6,541
6,828
3.16
1
4,081
4,177
4,201
4,226
4,325
4,351
4,376
10
1
4,081
4,310
4,370
4,430
4,679
4,744
4,809
0
2
4,081
6,643
7,504
8,476
13,798
15,586
17,605
0.01
2
4,081
6,302
7,025
7,831
12,093
13,480
15,027
0.0316
2
4,081
5,746
6,259
6,819
9,601
10,459
11,393
0.1
2
4,081
5,094
5,384
5,690
7,103
7,507
7,935
0.316
2
4,081
4,643
4,795
4,953
5,635
5,820
6,011
1
2
4,081
4,802
5,001
5,209
6,129
6,383
6,648
3.16
2
4,081
4,204
4,236
4,267
4,396
4,429
4,462
10
2
4,081
4,258
4,304
4,350
4,538
4,587
4,636
This estimate is based on the assumption of exponential population growth at uniform rate. Population growth over time in the absence of drug is likely to be approximately exponential. In the presence of drug, the population growth dynamics may not be exponential and may not be uniform over time. Our simulations demonstrate that the accuracy of lethal fraction kinetics do not depend on the validity of this assumption, as lethal fraction kinetics computed using this method were similar under many different assumed growth models. Additionally, lethal fraction kinetics computed using this method were similar to those experimentally calculated from time-lapse microscopy. This robustness depends on the length of the assay, as in short-term assays (48–72 h; effectively 2–3 population doublings), many different growth models produce a similar inference of live cell numbers or death kinetics. For longer assays, other growth models may provide more accurate estimates, including growth models with non-uniform rates over time (Richards, et al. 2020). See comment in Limitations below.Fluorescence of the total population (FTt)Subtract the measured fluorescence of dead cells (Table 2) from the total cell fluorescence generated in Table 5. The remaining fluorescence is proportional to the size of the live population at each time point (Table 6).
Table 6
Fluorescence of live cells (FLt)
Camptothecin dose (μM)
Rep. #
SYTOX fluorescence of live cells (RFU)
FL0
FL16
FL20
FL24
FL40
FL44
FL48
0
1
3,847
6,026
6,719
7,511
11,580
12,918
14,393
0.01
1
3,842
5,985
6,699
7,480
11,541
12,873
14,308
0.0316
1
3,828
5,726
6,319
6,969
10,265
11,315
12,478
0.1
1
3,857
4,960
5,283
5,631
7,140
7,557
8,119
0.316
1
3,843
4,479
4,649
4,816
5,539
5,703
5,910
1
1
3,839
4,474
4,464
4,423
5,209
5,319
5,464
3.16
1
3,824
3,530
2,877
2,461
1,695
1,182
746
10
1
3,842
3,808
3,228
2,568
1,478
1,287
989
0
2
3,855
6,407
7,251
8,218
13,451
15,201
17,166
0.01
2
3,851
6,023
6,779
7,574
11,749
13,092
14,598
0.0316
2
3,835
5,492
6,008
6,553
9,254
10,098
11,004
0.1
2
3,860
4,862
5,139
5,437
6,799
7,167
7,577
0.316
2
3,814
4,375
4,513
4,671
5,243
5,270
5,496
1
2
3,831
4,450
4,434
4,433
5,001
4,990
4,974
3.16
2
3,843
3,604
2,981
2,579
1,737
1,176
643
10
2
3,819
3,654
2,987
2,196
1,230
751
505
Fluorescence of live cells (FLt)FT = fluorescence of total cells (Table 5)FD = fluorescence of dead cells (Table 2)t = a given timepointx = a given drug concentrationThis simple subtraction (fluorescence of the total population – fluorescence of the dead population) should be proportional to the number of live cells. The accuracy of this inference depends on the linearity of the SYTOX fluorescence signal with cell number (Before you begin, steps 10 and 11, Figure 1D).
Calculate lethal fraction (LF) kinetics
Calculate lethal fraction kinetics (Figure 4D).Calculate lethal fraction (LF) for each condition/timepoint (Table 7):
Table 7
Lethal fraction (LFt)
Camptothecin dose (μM)
Rep. #
Lethal fraction
LF0
LF16
LF20
LF24
LF40
LF44
LF48
0
1
0.0573
0.0377
0.0360
0.0318
0.0273
0.0251
0.0240
0.01
1
0.0586
0.0436
0.0379
0.0347
0.0287
0.0264
0.0275
0.0316
1
0.0620
0.0422
0.0392
0.0369
0.0316
0.0297
0.0274
0.1
1
0.0549
0.0485
0.0467
0.0443
0.0514
0.0556
0.0456
0.316
1
0.0583
0.0559
0.0562
0.0585
0.0684
0.0763
0.0781
1
1
0.0593
0.0766
0.1173
0.1622
0.1688
0.1868
0.1998
3.16
1
0.0630
0.1549
0.3151
0.4177
0.6080
0.7284
0.8295
10
1
0.0586
0.1165
0.2613
0.4203
0.6841
0.7288
0.7943
0
2
0.0554
0.0355
0.0337
0.0304
0.0251
0.0247
0.0249
0.01
2
0.0564
0.0443
0.0350
0.0328
0.0284
0.0288
0.0285
0.0316
2
0.0603
0.0442
0.0401
0.0390
0.0361
0.0345
0.0341
0.1
2
0.0542
0.0455
0.0455
0.0445
0.0428
0.0453
0.0451
0.316
2
0.0654
0.0577
0.0588
0.0569
0.0696
0.0945
0.0857
1
2
0.0613
0.0733
0.1134
0.1490
0.1841
0.2182
0.2518
3.16
2
0.0583
0.1427
0.2963
0.3956
0.6049
0.7345
0.8559
10
2
0.0642
0.1418
0.3060
0.4952
0.7289
0.8363
0.8911
Lethal fraction (LFt)FD = dead cell fluorescence (from Table 2)FL = live cell fluorescence (from Table 6)t = a given time pointx = a given drug concentrationFor each dose of drug, determine response kinetics using a Lag Exponential Death (LED) equation (Forcina et al., 2017) (Table 8):
Table 8
Lag exponential death (LED) fitted parameters
Camptothecin dose (μM)
Rep. #
Lethal fraction
LFi
LFp
DO
DR
LFmax
LF0
…
LF48
0
1
0.0573
…
0.0240
0.056
0.022
0.953
0.054
0.057∗
0
2
0.0554
…
0.0249
0.01
1
0.0586
…
0.0275
0.057
0.028
11.374
0.133
0.059∗
0.01
2
0.0564
…
0.0285
0.0316
1
0.0620
…
0.0274
0.061
0.030
1.841
0.060
0.062∗
0.0316
2
0.0603
…
0.0341
0.1
1
0.0549
…
0.0456
0.048
0.046
34.480
0.014
0.056∗
0.1
2
0.0542
…
0.0451
0.316
1
0.0583
…
0.0781
0.059
0.084
39.558
1.176
0.095∗
0.316
2
0.0654
…
0.0857
1
1
0.0593
…
0.1998
0.060
0.219
14.586
0.081
0.252
1
2
0.0613
…
0.2518
3.16
1
0.0630
…
0.8295
0.061
1.000
13.239
0.041
0.856
3.16
2
0.0583
…
0.8559
10
1
0.0586
…
0.7943
0.061
0.957
14.547
0.056
0.891
10
2
0.0642
…
0.8911
Lag exponential death (LED) fitted parametersLFi = initial lethal fraction prior to death onset, unconstrainedLFp = lethal fraction plateau, constrained by LFp < 1DO = onset time of death, constrained by 0 < DO < final timepointDR = maximum rate of death, constrained by DR < 2The four kinetic parameters of the LED model should be determined using a least squares regression based curve fitting.In MATLAB, this can be performed using the built-in function “fit.” See associated GitHub repository to access a custom built function.In PRISM, LED parameters can be solved using the “plateau followed by one-phase association” function in the nonlinear regression, exponential analysis package, with manually constrained death onset time (X0 in PRISM).the LED equation is useful for determining the kinetics of cell death, namely through the DO and DR parameters. These features are not observable from end point data, and differ across drugs and drug classes (Inde et al., 2020). SGI-1027 has a dose-dependent decrease in DO whereas Camptothecin has more stable DO across doses (Figure 4E). Both SGI-1027 and Camptothecin result in near complete cell killing at high doses, but the death occurs at very different rates (Figure 4F).For drugs/doses that do not induce significant levels of cell death (see Table 8, ∗), kinetic terms are unreliable. These data are instead fit to a linear model with slope = 0 (Table 9):
Table 9
Lag exponential death (LED) fitted parameters, linear model (flat fit) included
Camptothecin dose (μM)
Rep. #
Lethal fraction
LFi
LFp
DO
DR
LFmax
LF0
…
LF48
0
1
0.0573
…
0.0240
0.034
0
48
0
0.057
0
2
0.0554
…
0.0249
0.01
1
0.0586
…
0.0275
0.037
0
48
0
0.059
0.01
2
0.0564
…
0.0285
0.0316
1
0.0620
…
0.0274
0.040
0
48
0
0.062
0.0316
2
0.0603
…
0.0341
0.1
1
0.0549
…
0.0456
0.048
0
48
0
0.056
0.1
2
0.0542
…
0.0451
0.316
1
0.0583
…
0.0781
0.067
0
48
0
0.095
0.316
2
0.0654
…
0.0857
1
1
0.0593
…
0.1998
0.060
0.219
14.586
0.081
0.252
1
2
0.0613
…
0.2518
3.16
1
0.0630
…
0.8295
0.061
1.000
13.239
0.041
0.856
3.16
2
0.0583
…
0.8559
10
1
0.0586
…
0.7943
0.061
0.957
14.547
0.056
0.891
10
2
0.0642
…
0.8911
Lag exponential death (LED) fitted parameters, linear model (flat fit) includedLFi = initial lethal fraction prior to death onset, unconstrainedLack of drug-induced cell death can be determined based on the error among replicates (i.e., no significant change from LFi over the time course). Alternatively, non-killing drugs can be defined as those for which LFmax values are not 2× greater than LFi.In MATLAB, curve fitting can be performed using the built-in function “fit,” defining the fittype as “poly1.”In PRISM or Excel, data can be fit using linear regression.
Calculate pharmacological dose-responses using fractional viability (FV), relative viability (RV), or normalized growth rate adjusted values (GR)
Calculate fractional viability (FV, Figures 5A and 5B)FV is a common measure of drug response for assays that count both live and dead cells (flow cytometry, quantitative microscopy, etc.). Unlike RV, FV does not report changes in population size, but rather the proportion of the population that is alive or dead. FV is the inverse of LF (FV = 1 − LF).Calculate fractional viability (FV) for each condition/timepoint (Table 10):
Table 10
Fractional viability (FVt)
Camptothecin dose (μM)
Rep. #
Fractional viability
FV0
FV16
FV20
FV24
FV40
FV44
FV48
0
1
0.9427
0.9623
0.9640
0.9682
0.9727
0.9749
0.9760
0.01
1
0.9414
0.9564
0.9621
0.9653
0.9713
0.9736
0.9725
0.0316
1
0.9380
0.9578
0.9608
0.9631
0.9684
0.9703
0.9726
0.1
1
0.9451
0.9515
0.9533
0.9557
0.9486
0.9444
0.9544
0.316
1
0.9417
0.9441
0.9438
0.9415
0.9316
0.9237
0.9219
1
1
0.9407
0.9234
0.8827
0.8378
0.8312
0.8132
0.8002
3.16
1
0.9370
0.8451
0.6849
0.5823
0.3920
0.2716
0.1705
10
1
0.9414
0.8835
0.7387
0.5797
0.3159
0.2712
0.2057
0
2
0.9446
0.9645
0.9663
0.9696
0.9749
0.9753
0.9751
0.01
2
0.9436
0.9557
0.9650
0.9672
0.9716
0.9712
0.9715
0.0316
2
0.9397
0.9558
0.9599
0.9610
0.9639
0.9655
0.9659
0.1
2
0.9458
0.9545
0.9545
0.9555
0.9572
0.9547
0.9549
0.316
2
0.9346
0.9423
0.9412
0.9431
0.9304
0.9055
0.9143
1
2
0.9387
0.9267
0.8866
0.8510
0.8159
0.7818
0.7482
3.16
2
0.9417
0.8573
0.7037
0.6044
0.3951
0.2655
0.1441
10
2
0.9358
0.8582
0.6940
0.5048
0.2711
0.1637
0.1089
Fractional viability (FVt)LF = lethal fraction (from Table 6)t = a given time pointx = a given drug concentrationDetermine pharmacological parameters for FV values by curve fitting, using the same 4-parameter logistic regression as for RV, above in Quantification and statistical analysis step 4b.Calculate relative viability (RV, Figures 5A and 5C).RV is a common measure of drug response used for most assays that rely on measures of metabolic activity (CellTiter-Glo, MTT, Alamar Blue, etc.). This measure reports the drug-induced change in population size, compared to untreated or vehicle-treated populations.Calculate relative viability (RV) for each condition/timepoint (Table 11).
Table 11
Relative viability (RVt)
Camptothecin dose (μM)
Rep. #
Relative viability
RV0
RV16
RV20
RV24
RV40
RV44
RV48
0
1
0.9990
0.9694
0.9619
0.9550
0.9252
0.9188
0.9121
0.01
1
0.9977
0.9626
0.9591
0.9511
0.9221
0.9156
0.9067
0.0316
1
0.9940
0.9211
0.9046
0.8861
0.8202
0.8048
0.7908
0.1
1
1.0016
0.7979
0.7564
0.7160
0.5705
0.5375
0.5145
0.316
1
0.9979
0.7205
0.6656
0.6124
0.4426
0.4057
0.3745
1
1
0.9969
0.7196
0.6391
0.5624
0.4162
0.3783
0.3463
3.16
1
0.9930
0.5678
0.4119
0.3129
0.1355
0.0840
0.0473
10
1
0.9977
0.6126
0.4621
0.3265
0.1181
0.0915
0.0627
0
2
1.0010
1.0306
1.0381
1.0450
1.0748
1.0812
1.0879
0.01
2
1.0000
0.9688
0.9705
0.9631
0.9387
0.9312
0.9251
0.0316
2
0.9958
0.8834
0.8602
0.8332
0.7394
0.7182
0.6974
0.1
2
1.0023
0.7820
0.7357
0.6914
0.5432
0.5098
0.4802
0.316
2
0.9904
0.7038
0.6462
0.5939
0.4189
0.3748
0.3483
1
2
0.9948
0.7158
0.6348
0.5636
0.3996
0.3549
0.3152
3.16
2
0.9979
0.5797
0.4267
0.3280
0.1388
0.0836
0.0407
10
2
0.9917
0.5878
0.4276
0.2792
0.0983
0.0534
0.0320
Replicates of the control condition are averaged before calculating RV:FL = modeled live cell fluorescence (from Table 6)t = a given time pointx = a given drug concentrationRelative viability (RVt)Model RV values using a 4-parameter logistic regression (at any timepoint but most commonly using data from the assay endpoint):x = dose in log10 scaleEmax = maximum effect (e.g., bottom of curve)RV0 = relative viability in absence of drug (e.g., top of curve)EC50 = dose in log scale corresponding to the mid-point of the sigmoidal curve (inflection point)h = Hill coefficientThe four parameters of the logistic regression model should be determined using a least squares regression based curve fitting.In MATLAB, this can be performed using the built-in function “fit,” defining the fit type using the equation listed above. See associated GitHub repository to access a custom built function.In PRISM, this can be performed using the “log(inhibitor) vs. response – variable slope” analysis function.This equation requires a log-transformation of the dose values and cannot accept a dose of “0.” These doses can be omitted from the fit or set to a value 10× lower than the lowest dose.Calculate the normalized growth rate inhibition value (GR, Figures 5A and 5D).GR reports the drug response in terms of the growth rate of drug-treated cells, compared to the untreated growth rate. This approach normalizes for differences in the apparent drug response that may be due to differences in assay length or differences in growth rates between cells of interest. GR calculations are described in detail in Hafner et al. 2016.Calculate the average fluorescence of live cells from the replicates of untreated control wells at the assay start (FL0,, see Table 5) (Table 12).
Table 12
Untreated control at assay start (FL0,)
Camptothecin dose (μM)
Replicate #
Live cell fluorescence
FL0,ctrl
FLM0,ctrl
0
1
3,847
3,851
0
2
3,855
Untreated control at assay start (FL0,)Calculate the average fluorescence of live cells from the replicates of the untreated control at the timepoint of interest (FLt, see Table 6). An example calculating the mean fluorescence of live cells at 48 h is shown in Table 13.
Table 13
Untreated control at timepoint of interest (FLt,)
Camptothecin dose (μM)
Replicate #
Live cell fluorescence
FL48,ctrl
FLM48,ctrl
0
1
14,393
15,780
0
2
17,166
Untreated control at timepoint of interest (FLt,)Determine the fluorescence contributed by the proportion of the population that was still alive at each timepoint in each of the treated conditions at the assay start (FL0,d) and any timepoint of interest (FLt,d) (see Table 6).Calculate the GR value for each condition/timepoint (Table 14):
Table 14
GR values (GRt)
Camptothecin dose (μM)
Rep. #
GR index
GR0
GR16
GR20
GR24
GR40
GR44
GR48
0
1
1
0.9148
0.9139
0.9145
0.9118
0.9124
0.9126
0.01
1
1
0.8992
0.9103
0.9093
0.9095
0.9102
0.9083
0.0316
1
1
0.7911
0.7922
0.7888
0.7862
0.7863
0.7873
0.1
1
1
0.4392
0.4424
0.4439
0.4364
0.4333
0.4417
0.316
1
1
0.2482
0.2482
0.2450
0.2399
0.2353
0.2356
1
1
1
0.2479
0.1920
0.1473
0.1966
0.1907
0.1894
3.16
1
1
−0.1093
−0.2818
−0.3481
−0.3802
−0.4667
−0.5521
10
1
1
−0.0126
−0.1835
−0.3237
−0.4298
−0.4432
−0.4867
0
2
1
1.0861
1.0864
1.0852
1.0854
1.0842
1.0835
0.01
2
1
0.9103
0.9315
0.9283
0.9270
0.9251
0.9249
0.0316
2
1
0.6817
0.6866
0.6821
0.6788
0.6790
0.6787
0.1
2
1
0.3964
0.3953
0.3947
0.3950
0.3927
0.3930
0.316
2
1
0.2198
0.2166
0.2174
0.2059
0.1890
0.1967
1
2
1
0.2420
0.1856
0.1521
0.1697
0.1520
0.1369
3.16
2
1
−0.0886
−0.2561
−0.3210
−0.3731
−0.4694
−0.5846
10
2
1
−0.0618
−0.2488
−0.4157
−0.4863
−0.5813
−0.6300
GR values (GRt)FL = modeled live cell fluorescence (from Table 6)FLM = mean fluorescence of live cells (from Tables 12 and 13)t = a given timepointx = a given drug concentrationDetermine pharmacological parameters for GR values by curve fitting, using the same 4-parameter logistic regression as for RV and FV, above in Quantification and statistical analysis step 4b.
Limitations
The FLICK assay provides a single unified platform for evaluating drug response kinetics, as well as determining a variety of pharmacological measures, such as relative viability, fractional viability, and GR values. These values can be obtained at any time point in the assay window, or at the assay end point. Additionally, because FLICK requires only one fluorescence channel to mark dead cells, this assay can be optimized to be multiplexed with markers of other cell fates or responses. This protocol has been optimized for adherent cell lines, but should be easily adapted to suspension cultures.The FLICK assay has a few notable limitations. The use of a plate reader requires the user to manually move plates between the incubator and the reader. The lack of automation may limit throughput, although commercially available automated microscopes also generally accommodate only a small number of plates. More critically, FLICK requires that the user select time points for analysis. As discussed in this protocol (Fluorescence measurements, step 14) and in the Troubleshooting section, this may require some optimization depending on the kinetics of cell death observed for a given drug. An additional limitation is that the output of this assay is not cell number, but fluorescence values for a population of cells (i.e., a well of a plate). This value, however, is linearly proportional to the number of cells, so the number of cells can be estimated from fluorescence values by generating a standard curve, similar to the cell dilution used when calibrating SYTOX (see Before you begin, step 1).Some drugs emit fluorescence in the same wavelength as SYTOX and cannot be quantified using this assay. In these situations, use of SYTOX variants that emit fluorescence in other wavelengths may be helpful. This assay is also unable to measure the response to drugs that interfere with SYTOX fluorescence or the ability of SYTOX to bind DNA. Lastly, the demarcation of dead cells by SYTOX requires both plasma membrane rupture and disruption of the nuclear envelope. These membranes require active maintenance, so SYTOX will mark dead cells regardless of the mechanism of cell death caused by the drug. These events, however, do not occur with the same efficiency for all types of cell death, which may affect the ability of SYTOX to accurately quantify dead cells.Our kinetic inference of live and dead cells over time makes one critical assumption regarding the nature of growth for drug-treated populations. For any population growth that is observed over time, our approach assumes that growth was exponential at a constant rate over time. While approximately exponential growth is generally observed for healthy cells, the same is not necessarily true for cells experiencing drug-induced stress. By simulating responses using a spectrum of different growth models, we previously found that inferred response kinetics were stable regardless of which growth model was chosen (e.g., linear, sigmoidal, and non-uniform over time) (Richards et al., 2020). This overall robustness regardless of growth model depends on selection of a relatively short assay time (48–72 h; generally 2–3 population doublings). At longer time points, solutions derived from different growth assumptions will begin to diverge more substantially. In these situations, selection of a population growth model that more accurately mimics what is observed for drug-induced growth may be helpful.Even at short assay times, our simulations did identify some situations that differ substantially from the uniform exponential growth rate assumption; however, these required very peculiar growth kinetics that were not ever observed experimentally. Furthermore, the kinetic data that result from an exponential growth assumption correlate well with data derived from single cell microscopy for a panel of diverse drugs. Nonetheless, in some situations the assumption of a constant exponential growth rate may produce incorrect inferences. Kinetic inference of total cells can become inaccurate for drugs that kill at very fast rates and with early onset times. Drugs with these features compromise our calculation because the cells have decayed too much by the assay endpoint to produce a SYTOX signal. These situations could be remedied by ending the assay at an earlier time point.
Troubleshooting
Problem
Final pre-Triton measurement is higher than the post-Triton measurement (steps 14–16).
Potential solution
In principle, using Triton to lyse remaining viable cells should always lead to stable or increased SYTOX signal. If lethal fraction is ∼ 1, post-Triton measurements may be slightly lower or higher, just due to measurement noise. A post-Triton SYTOX signal that is substantially decreased from the pre-Triton measurement can result from agitating the plate and dislodging the dead cells which may settle in a portion of the well in which fluorescence readings are not captured. This issue can be remedied by allowing the dislodged cells to settle before reading the plate, or by repeating the experiment being careful not to shake or agitate the plate.Nonlinear relationship between cell number and SYTOX fluorescence (Before you begin, step 11).Assuming all signals are in the linear range (i.e., no readings of “INF,” “NaN,” “error,” etc.), non-linearity can be caused by incomplete permeabilization. Visually inspect the plate to see if Triton permeabilization has caused rupture of all cells. Alternatively, increase the Triton permeabilization time.Lethal fraction decreases over time (step 14).Lethal fraction measurements generally increase over time until reaching a stable plateau. In rare cases, LF% increases rapidly before slowly decreasing to a lower plateau. This can occur either due to the emergence of a drug resistant clone or due to degradation of the SYTOX signal over time. These two possibilities can be distinguished by directly measuring live cell kinetics over time using the STACK assay (Forcina et al, 2017).Inaccurate inference of death kinetics when comparing to other assays (Quantification and statistical analysis, step 3).In some situations, users may want to use kinetic data derived from this approach to parameterize dose/time selection for other experiments, for instance, finding time points before or after the onset of cell death. These inferred times can be inaccurate if the population growth varies significantly from the exponential growth model used. This is particularly an issue if growth in drug-treated conditions has three characteristics: 1) very non-uniform growth over time, 2) abrupt transition from growing to non-growing, and 3) transitions from growing to non-growing are unrelated to the death onset time. In the absence of these three features, inferred kinetic rates from this method should be accurate even if the population growth is non-exponential. In rare situations where these issues occur, potential solutions are to: shorten the assay length to minimize differences between varied growth assumptions, compute death kinetics using a different growth model, or use the STACK assay to directly observe death kinetics (Forcina, et al. 2017).Unreliable death onset time (Quantification and statistical analysis, step 3).Selection of times to measure SYTOX fluorescence affects the accuracy of the death onset calculation. The inferred death onset time will vary substantially between experiments if it is not appropriately constrained by death observed at the selected time points. A solution to this is to optimize the selection of time points or to add more time points. The inference of death onset time will be improved if death levels are sampled at ∼3 consecutive time points, spaced 2–4 h apart, in the exponentially increasing phase of death, and in the plateau phase of death.
Resource availability
Lead contact
Further information and requests for resources and reagents should be directed to and will be fulfilled by the Lead Contact, Michael Lee (michael.lee@umassmed.edu).
Materials availability
This protocol uses only materials and reagents that are commercially available.
Data and code availability
Sample datasets and code used for analyzing data using this protocol are available in a GitHub repositories (MJLee-Lab/fitGrowth; MJLee-Lab/fitLED; MJLee-Lab/fit_via).