Laura Stephens1, Wendy Hintz-Prunty1, Hans-Inge Bengtsson2, James A Proudfoot3, Sandip Pravin Patel4,5, H Elizabeth Broome1,4. 1. Department of Pathology, University of California, San Diego, San Diego, CA. 2. CellaVision AB, Ideon Science Park, Lund, Sweden. 3. Department of Clinical and Translational Research Institute, University of California, San Diego, San Diego, CA. 4. Moores Cancer Center, University of California, San Diego, San Diego, CA. 5. Department of Medicine, University of California, San Diego, San Diego, CA.
Abstract
BACKGROUND: To optimize precision of nucleated blood cell counting, the clinical laboratory scientist should post the automated differential rather than the manual differential if the former is within the 95% CI of the latter, as determined by the "Rumke statistic." The objective of this study was to determine the potential impact of real-time, computer-assisted use of Rumke statistics for more judicious use of the automated vs digitally imaged, manual differential. METHODS: Complete blood counts with automated differentials produced by a XE5000™ hematology analyzer (Sysmex) were compared with both the DM96 (CellaVision™ AB) preclassification differentials and the posted reclassified manual differentials, using the Rumke 95% CIs as calculated using the Clopper-Pearson method. RESULTS: A total of 44.7% of manual differentials had no statistical or clinical justification over the automated differential. In addition, 31.1% of manual differentials had statistical discrepancies between the instrument absolute neutrophil count (IANC) and manual absolute neutrophil count (ANC). Nineteen of these IANC/manual ANC discrepant samples had ANCs below 1500/μL, a decision level for cancer treatment. Holding the IANC when it is less than 2000/μL until after manual smear review would have prevented the posting of any IANC vs manual ANC discrepant results at the 1500/μL ANC decision threshold. CONCLUSIONS: A real-time operator alert concerning the statistical identity of imaging device differentials vs automated differentials could have reduced manual differentials by nearly 45%. Not posting unnecessary manual differentials for the cases with IANC/manual ANC discrepancies would have likely reduced clinical error/confusion.
BACKGROUND: To optimize precision of nucleated blood cell counting, the clinical laboratory scientist should post the automated differential rather than the manual differential if the former is within the 95% CI of the latter, as determined by the "Rumke statistic." The objective of this study was to determine the potential impact of real-time, computer-assisted use of Rumke statistics for more judicious use of the automated vs digitally imaged, manual differential. METHODS: Complete blood counts with automated differentials produced by a XE5000™ hematology analyzer (Sysmex) were compared with both the DM96 (CellaVision™ AB) preclassification differentials and the posted reclassified manual differentials, using the Rumke 95% CIs as calculated using the Clopper-Pearson method. RESULTS: A total of 44.7% of manual differentials had no statistical or clinical justification over the automated differential. In addition, 31.1% of manual differentials had statistical discrepancies between the instrument absolute neutrophil count (IANC) and manual absolute neutrophil count (ANC). Nineteen of these IANC/manual ANC discrepant samples had ANCs below 1500/μL, a decision level for cancer treatment. Holding the IANC when it is less than 2000/μL until after manual smear review would have prevented the posting of any IANC vs manual ANC discrepant results at the 1500/μL ANC decision threshold. CONCLUSIONS: A real-time operator alert concerning the statistical identity of imaging device differentials vs automated differentials could have reduced manual differentials by nearly 45%. Not posting unnecessary manual differentials for the cases with IANC/manual ANC discrepancies would have likely reduced clinical error/confusion.