Literature DB >> 33632579

High Rates of Spacer Fracture in the Setting of Extended Trochanteric Osteotomy With a Specific Thin-Core Articulating Antibiotic Hip Spacer.

Alex J Lancaster1, Victor R Carlson1, Christopher E Pelt1, Lucas A Anderson1, Christopher L Peters1, Jeremy M Gililland1.   

Abstract

BACKGROUND: Two-stage revision remains the standard of care for prosthetic joint infection after total hip arthroplasty. However, there are substantial complications associated with articulating antibiotic hip spacers. Handmade and molded spacers have been shown to have higher rates of spacer fracture than antibiotic-coated prostheses (ACPs). The aim of this study is to review outcomes with an implant that is often categorized as an ACP spacer, the Zimmer-Biomet StageOne Select Femoral Spacer (ZBSO).
METHODS: A retrospective review was performed of 63 patients who underwent placement of a ZBSO. Patients were compared based on whether or not an extended trochanteric osteotomy (ETO) was performed using Fisher's exact and t-tests.
RESULTS: Five patients were excluded due to lack of follow-up or death shortly after stage 1 surgery, leaving 58 patients. Spacer fracture was noted in 5 of 58 patients (8.6%). Sixteen patients underwent ETO and 25.0% suffered a spacer fracture compared to 2.3% without ETO (odds ratio 13.7, P = .0248). There was no association between patient demographics or ETO length and spacer fracture. Two patients had periprosthetic fractures (3.4%) and 4 had dislocations (6.9%). Forty-nine patients (84.4%) went on to second-stage revision; of those 26.5% failed to clear the infection and required an average of 2.2 additional surgeries.
CONCLUSION: The ZBSO spacer has overall complication rates similar to previously reported spacer series. Although the ZBSO looks like an ACP spacer, in the setting of ETO, it behaves like a molded or handmade spacer with a high rate of spacer fracture (25%) due to the small diameter of the core. This implant should be used with caution in combination with an ETO.
Copyright © 2021 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  antibiotic-coated prostheses; articulating hip spacer; extended trochanteric osteotomy; hip arthoplasty; prosthetic joint infection

Year:  2021        PMID: 33632579     DOI: 10.1016/j.arth.2021.01.086

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

Review 1.  Mechanical complications of hip spacers: a systematic review of the literature.

Authors:  Andrea Sambri; Michele Fiore; Claudia Rondinella; Lorenzo Morante; Azzurra Paolucci; Claudio Giannini; Calogero Alfonso; Massimiliano De Paolis
Journal:  Arch Orthop Trauma Surg       Date:  2022-04-12       Impact factor: 2.928

2.  Extended Trochanteric Osteotomy with Intermediate Resection Arthroplasty Is Safe for Use in Two-Stage Revision Total Hip Arthroplasty for Infection.

Authors:  Sebastian Hardt; Vincent Justus Leopold; Thilo Khakzad; Matthias Pumberger; Carsten Perka; Christian Hipfl
Journal:  J Clin Med       Date:  2021-12-22       Impact factor: 4.241

3.  Extended trochanteric osteotomy in revision hip arthroplasty: a case series study and systematic literature review.

Authors:  Khalid Hamad; Sujith Konan
Journal:  Arthroplasty       Date:  2022-04-03
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.