Left ventricular mass on positron emission tomography: Validation against cardiovascular magnetic resonance.

Maan Malahfji1, Ahmed Ibrahim Ahmed1, Yushui Han1, Aaron K Jung1, Talal Alnabelsi1, Faisal Nabi1, Dipan J Shah1, Ahmed Aljizeeri2,3, Alexis Poitrasson-Rivière4, Edward Ficaro4, John J Mahmarian1, Mouaz H Al-Mallah5.   

Abstract

BACKGROUND: Left ventricular hypertrophy (LVH) is an important clinical finding that is independently associated with mortality and cardiovascular events. We aimed to assess the interstudy variability of LV mass quantitation between PET and CMR.
METHODS: Patients who underwent both PET and CMR within 1 year were identified from prospective institutional registries. LV mass on PET was compared against LV mass on CMR using several statistical measures of agreement.
RESULTS: A total of 105 patients (mean age 60 ± 14 years, 67.6% male) were included. The median (interquartile range, IQR) duration between CMR and PET was 47 (11-154) days. The median (IQR) LV mass values were 168.0 g (126.0-202.0) on CMR and 174.0 g (150.0-212.0) with PET (absolute mean difference 29.42 ± 25.3). There was a good correlation (Spearman ρ = 0.81, P < 0.001; Intraclass Correlation Coefficient 0.78, 95% CI 0.70-0.85, P < 0.001) with moderate limits of agreement (95% limits of agreement - 63.78 to 83.7.) Results were consistent, albeit with moderate correlation, in subgroups of patients with LVH, in patients with myocardial infarction, in patients with LV ejection fraction < 50%, and those with limited image quality. LV mass on PET tended to be underestimated at high values compared to CMR.
CONCLUSION: We demonstrate good correlation and reproducibility of LV mass quantitation by PET against the reference standard of CMR across a wide range of normal and diseased hearts with a tendency of PET to underestimate mass at higher mass values.
© 2021. American Society of Nuclear Cardiology.

Entities:  

Keywords:  CMR; PET; nuclear cardiac imaging

Mesh:

Year:  2021        PMID: 33629247     DOI: 10.1007/s12350-021-02537-6

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   3.872


  2 in total

1.  Left ventricular function and perfusion from gated SPECT perfusion images: an integrated method.

Authors:  T L Faber; C D Cooke; R D Folks; J P Vansant; K J Nichols; E G DePuey; R I Pettigrew; E V Garcia
Journal:  J Nucl Med       Date:  1999-04       Impact factor: 10.057

2.  Comparison of 64-slice CT with gated SPECT for evaluation of left ventricular function.

Authors:  Tiziano Schepis; Oliver Gaemperli; Pascal Koepfli; Ines Valenta; Klaus Strobel; Andrea Brunner; Sebastian Leschka; Lotus Desbiolles; Lars Husmann; Hatem Alkadhi; Philipp A Kaufmann
Journal:  J Nucl Med       Date:  2006-08       Impact factor: 10.057

  2 in total
  2 in total

1.  Added value to stress myocardial perfusion imaging studies with measurement of left ventricular mass.

Authors:  Anahita N Tavoosi; Yoshito Kadoya; Terrence D Ruddy
Journal:  J Nucl Cardiol       Date:  2021-10-19       Impact factor: 3.872

2.  Diagnosis of left ventricular hypertrophy using non-ECG-gated 15O-water PET.

Authors:  Jens Sörensen; Jonny Nordström; Tomasz Baron; Stellan Mörner; Sven-Olof Granstam; Mark Lubberink; Lars Tolbod; Jeffrey van den Berg; Frank A Flachskampf; Tanja Kero; Peter Magnusson; Hendrik J Harms
Journal:  J Nucl Cardiol       Date:  2021-07-20       Impact factor: 3.872

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.