Wan-Soo Yoon1, Hyoung-Woo Lho1, Dong-Sup Chung1.
Abstract
OBJECTIVE: Though the operating microscope (OM) has been the standard optical system in neurosurgery, a new technology called three-dimensional (3D) exoscope has emerged as an alternative. Herein, two types of 3D exoscopes for brain tumor surgery are presented. In addition, the advantages and limitations compared with the OM are discussed.Entities:
Keywords: 3D exoscope; Brain neoplasms; Microscopy; Surveys and questionnaires
Year: 2021 PMID: 33626859 PMCID: PMC7969046 DOI: 10.3340/jkns.2020.0199
Source DB: PubMed Journal: J Korean Neurosurg Soc ISSN: 1225-8245
Clinical characteristics of patients performed the neurosurgical resection under exoscopes
| No | Age | Sex | Type of exoscope | Histopathology | Location | Size (cm) | Volume (mL) | Extent of resection | Surgical complication |
|---|---|---|---|---|---|---|---|---|---|
| 1 | 68 | M | VOMS-100 | Metastasis | Cbll | 3.2×3.6×2.5 | 15.6 | GTR | None |
| 2 | 78 | F | VOMS-100 | GBM | Parietal | 5.6×5.1×3.9 | 43.1 | GTR | None |
| 3 | 66 | F | VOMS-100 | Metastasis | Frontal | 4.3×3.3×4.0 | 27.9 | GTR | None |
| 4 | 67 | M | VOMS-100 | Meningioma | Frontal convexity | 4.8×3.8×5.6 | 48.1 | GTR | None |
| 5 | 66 | F | VOMS-100 | Meningioma | Parasagittal | 9×4.5×8.2 | 131 | GTR | None |
| 6 | 49 | F | VITOM 3D | Metastasis | Medial prefrontal | 4×3.4×3.8 | 19.3 | STR | None |
| 7 | 57 | M | VITOM 3D | Anaplastic astrocytoma | Frontal | 6.2×5.0×4.6 | 66.1 | STR | None |
| 8 | 39 | F | VITOM 3D | Anaplastic xanthoastrocytoma | Frontal | 8×5.6×4.8 | 65.3 | GTR | None |
| 9 | 62 | M | VITOM 3D | GBM | Medial prefrontal | 3.6×3.3×3.8 | 16.1 | GTR | None |
| 10 | 26 | F | VITOM 3D | Meningioma | Frontal convexity | 3.8×3×2.1 | 20.2 | GTR | None |
| 11 | 39 | M | VITOM 3D | Hemangioblastoma | Cbll | 4.3×3×2.6 | 11.8 | GTR | Remote EDH |
M : male, Cbll : cerebellum, GTR : grossly total removal, F : female, GBM : glioblastoma, STR : subtotal removal, EDH : epidural hematoma
Fig. 1.Operating room during exoscopic surgery using VITOM 3D. The 3D monitor is displayed in front of surgeons. Surgeons can see the surgical field with the naked eye due to the long working distance of the exoscope and can watch the magnified image on the monitor while wearing 3D glasses.
Comparison of technical specifications between exoscopes and microscope
| VOMS-100 | VITOM 3D | Kinevo 900 | |
|---|---|---|---|
| Working distance (cm) | 21/22/23/31/35 | 20–50 | 20–62.5 |
| Optical zoom | 1–2× | 1–4× | 1–6× |
| Magnification | 7–80× | 8–30× | 0.9–41.3× |
| Depth of field (focal length) (cm) | 5/9.5/15/25 | 3.5–10 | 17/26 |
| Fluorescence detection | No | No | Yes |
| Integration with navigation system | No | No | Yes |
| Light source | LED | LED | Xenon |
| Auto focusing | No | No | Yes |
| Displayed video image | Full-HD | 4K | Full-HD/4K |
| Auto-balance | No | No | Yes |
| Controller type | Foot panel | Mounted jog dial | Foot panel |
| Movement by controller | Possible | Impossible | Possible |
| Weight (kg) | 95 | N/D | 395 |
| Cost[ | Approximately 50 million | Approximately 150 million | Approximately 500 million |
Source of Kinevo 900 (https://www.zeiss.com/meditec/int/product-portfolio/surgical-microscopes/kinevo-900.html).
LED : light-emitting diode, HD : high definition, N/D : not described, KRW : Korea won
Questionnaire about the experience of exoscopes compared with microscope
| Question | VOMS-100 | VITOM 3D | KINEVO 900 | |
|---|---|---|---|---|
| Image quality on the displayed monitor | ||||
| 1. How was the quality of the overall image? | ( )/5 | ( )/5 | ( )/5 | |
| 2. How was the quality of the image when hemorrhage was visible in the surgical field? | ( )/5 | ( )/5 | ( )/5 | |
| 3. How was the quality of the image when the surgical field was magnified? | ( )/5 | ( )/5 | ( )/5 | |
| 4. How was the overall brightness of the image? | ( )/5 | ( )/5 | ( )/5 | |
| 5. How was the brightness of the image when the deep surgical field was shown? | ( )/5 | ( )/5 | ( )/5 | |
| Handling of the equipment | ||||
| 6. How easy was it to move instruments in the surgical field? | ( )/5 | ( )/5 | ( )/5 | |
| 7. How easy was it to focus on the surgical field? | ( )/5 | ( )/5 | ( )/5 | |
| 8. How convenient was it to prepare and install? | ( )/5 | ( )/5 | ( )/5 | |
| 9. How much space did the equipment occupy in the operating theater? | ( )/5 | ( )/5 | ( )/5 | |
| 10. How was the overall convenience during surgery? | ( )/5 | ( )/5 | ( )/5 | |
| Ergonomics | ||||
| 11. How much eye fatigue did you experience during surgery? | ( )/5 | ( )/5 | ( )/5 | |
| 12. How much body discomfort did you experience? | ( )/5 | ( )/5 | ( )/5 | |
| Educational usefulness | ||||
| 13. How appropriate was it for educational purposes? | ( )/5 | ( )/5 | ( )/5 | |
| 3D glasses | ||||
| 14. Which was more comfortable, the eyepiece lenses or 3D glasses? | (1) E. lens | (2) 3D glasses | (3) N/C | |
| 15. Did you experience dizziness while wearing the 3D glasses? | (1) Yes | (2) No | (3) N/C | |
| Expectation | ||||
| 16. Do you think the exoscope can currently be used as a substitute for the OM in brain tumor surgery? | (1) Yes | (2) No | (3) N/C | |
| 17. Do you expect the exoscope will replace the OM in the future? | (1) Yes | (2) No | (3) N/C | |
3D : 3-dimensional, E. : eyepiece, N/C : no choice, OM : operating microscope
Fig. 2.Results of questionnaire regarding the experience of using exoscopes (VOMS-100 or VITOM 3D) compared with the OM (Kinevo 900). A : Questions regarding image quality on the display monitor. VITOM 3D was equivalent to Kinevo 900 except for quality of the magnified image (Q. 3). However, VOMS-100 was inferior to Kinevo 900 in all aspects (Q. 1–5). B : Questions regarding handling of equipment. Exoscopes were inferior in focusing on the surgical field (Q. 7). However, exoscopes were superior in accessibility of instruments in the surgical field (Q. 6) and amount of space occupied in the operating theater (Q. 9). C : Questions regarding ergonomics and educational usefulness. Difference was not observed between exoscope and OM. D : Questions regarding 3D glasses and expectations. Discomfort in wearing the 3D glasses was not reported, and respondents had high expectations regarding substitution of exoscopes for the OM. *p<0.05. Q : question, N/C : no choice, OM : operating microscope.