A novel simple technique to enlarge the aortic annulus by two valve sizes.

Bo Yang1.   

Abstract

Entities:  

Year:  2020        PMID: 33623928      PMCID: PMC7899136          DOI: 10.1016/j.xjtc.2020.10.038

Source DB:  PubMed          Journal:  JTCVS Tech        ISSN: 2666-2507


× No keyword cloud information.
Y incision at aortomitral curtain (L) and rectangular patch to enlarge aortic root (R). Y incision at the aortomitral curtain and rectangular patch enlarged the aortic root by 2 sizes without violating the mitral valve. See Commentaries on pages 17 and 19. The current techniques of aortic root enlargement used by adult cardiac surgeons are the Nicks and Manouguian procedures. The Nicks procedure generally increases aortic annulus by one valve size. The Manouguian requires incising the mitral valve (MV) anterior leaflet and left atrium (LA), with risk of mitral regurgitation. Reported herein is a new surgical technique to enlarge the aortic annulus by 2 valve sizes without violating the LA or MV.

Case Report

A 62-year-old female patient with body height of 5′2′′ (158 cm) and surface area of 1.66 m2, status post combined pancreas—kidney transplant and second kidney transplant with significant renal dysfunction—presented with acute myocardial infarction, severe aortic stenosis, and a left ventricular ejection fraction of 35%. She underwent urgent 4-vessel coronary artery bypass, aortic annular enlargement, and aortic valve replacement (AVR). The annulus size was measured at 19 mm after excision of the leaflets and annular debridement of calcium. After an aortic root enlargement procedure, a 23-mm Magna Ease (Edwards Lifesciences; Irvine, Calif) aortic bioprosthesis was implanted. Postoperative aortic valve mean gradient was 7 mm Hg, and aortic root diameter increased from 23 to 30 mm with normal coronary blood flow and no mitral regurgitation. This patient was discharged without complications, and provided informed consent for publication of the information included in this case report.

Details of Aortic Root Enlargement (Video 1)

A partial transverse aortotomy was made 1.5 cm above the sinotubular junction and stopped above the left-non commissure post. The stenotic aortic valve was excised and the annulus was debrided of calcium. The left-non commissure post was incised from the aortotomy into the aortomitral curtain close to the left atrial dome. The incision was extended in a “Y” fashion undermining the left and noncoronary annulus to their respective nadir (Figure 1, A). This maneuver widely opens the aortic root (Figure 1, B). A rectangular-shaped Hemashield Dacron patch (Boston Scientific Corp, Natick, Mass) was trimmed in width slightly greater than the distance between the 2 cusp nadirs. This patch was sewn to the mitral annulus from left to right fibrous trigone with running 4-0 PROLENE suture (Ethicon, Somerville, NJ). The suture line was transitioned to the undermined aortic annulus at the nadir of both left and noncoronary sinuses, sutured along the longitudinal length of the patch up to the level of the transverse aortotomy incision, and secured (Figures 1, C, and 2). The upsized valve sizer was placed in the enlarged root touching 3 nadirs of aortic annulus and the position of the sizer on the patch was marked to guide the placement of valve sutures (Figure 3, A). The non-pledgetted 2-0 ETHIBOND sutures (Ethicon) were placed along the native aortic annulus in a non-everting fashion and from outside in on the patch (Figure 3, B). The bioprosthesis was placed with one strut facing the left-right commissure post, one strut close to right-non commissure post, and one facing the patch (Figure 3, C). The sutures at nadirs of noncoronary and left coronary sinuses, which were the lowest point of aortic annulus, were tied first to seat the valve well and prevent paravalvular leak. A portion of the patch lay beneath the prosthetic valve thereby enlarging the root and left ventricular outflow tract (Figure 3, D). The aortotomy was then closed from the left to the right incorporating the patch (Figure 3, E).
Figure 1

A, A “Y” incision through the left-non commissure post extended below the aortic annulus to the area between the nadirs of left and noncoronary cusps and mitral valve annulus for the aortic root enlargement. B, The aortic root was wide open after the Y incision (white arrow). Yellow arrow shows the left coronary ostium. C, In the left panel, a rectangular patch was sewn to the mitral annulus (white arrow) at the bottom and aortic annulus (white stars) on both sides, then to the incision of the transverse aortotomy. Yellow arrow shows left coronary ostium. In the right panel, an illustration of the suture lines of the anastomosis of the Dacron patch to the mitral and aortic annulus.

Figure 2

Y incision at aortomitral curtain (left) and rectangular patch to enlarge aortic root (right).

Figure 3

A, The valve position was marked on the Dacron patch to guide the placement of valve sutures with the 2-size larger sizer touching 3 nadirs of the aortic annulus of 3 aortic cusps. B, 2-0 ETHIBOND sutures were placed along the aortic annulus in a non-everting fashion and from outside in on the patch (left panel, white arrow: mitral annulus; white stars: aortic annulus sewn to the patch). The transitional stitches were placed with one arm of the suture through the aortic annulus (non-everting: from ventricle to aorta), and the other arm through the patch (one bite: inside-outside-inside) (right panel: illustration). C, The upsized bioprosthetic valve was seated right above the aortic annulus and the Dacron patch with one strut facing the left-right commissure post to position the coronary ostia above the nadir of left and right coronary cusps of the bioprosthesis. D, The left ventricular outflow tract was wide open underneath the upsized bioprosthesis (view from inside of the valve). E, The partial transverse aortotomy was closed incorporating the Dacron patch.

A, A “Y” incision through the left-non commissure post extended below the aortic annulus to the area between the nadirs of left and noncoronary cusps and mitral valve annulus for the aortic root enlargement. B, The aortic root was wide open after the Y incision (white arrow). Yellow arrow shows the left coronary ostium. C, In the left panel, a rectangular patch was sewn to the mitral annulus (white arrow) at the bottom and aortic annulus (white stars) on both sides, then to the incision of the transverse aortotomy. Yellow arrow shows left coronary ostium. In the right panel, an illustration of the suture lines of the anastomosis of the Dacron patch to the mitral and aortic annulus. Y incision at aortomitral curtain (left) and rectangular patch to enlarge aortic root (right). A, The valve position was marked on the Dacron patch to guide the placement of valve sutures with the 2-size larger sizer touching 3 nadirs of the aortic annulus of 3 aortic cusps. B, 2-0 ETHIBOND sutures were placed along the aortic annulus in a non-everting fashion and from outside in on the patch (left panel, white arrow: mitral annulus; white stars: aortic annulus sewn to the patch). The transitional stitches were placed with one arm of the suture through the aortic annulus (non-everting: from ventricle to aorta), and the other arm through the patch (one bite: inside-outside-inside) (right panel: illustration). C, The upsized bioprosthetic valve was seated right above the aortic annulus and the Dacron patch with one strut facing the left-right commissure post to position the coronary ostia above the nadir of left and right coronary cusps of the bioprosthesis. D, The left ventricular outflow tract was wide open underneath the upsized bioprosthesis (view from inside of the valve). E, The partial transverse aortotomy was closed incorporating the Dacron patch.

Discussion

A small aortic root can cause patient–prosthesis mismatch and exclude patients from transcatheter AVR. Our concept is similar to the Nicks and Manouguian procedures—to enlarge the fibrous portion of the root and subsequently the aortic annulus. The difference is our technique radically enlarges the fibrous portion of the aortic root by replacing the entire aortomitral curtain with a rectangular patch to accommodate an increase of 2 valve sizes from the initial assessment without violating the LA or MV. This enlargement technique originated from our experience of reconstructing the aortic root after the aortomitral curtain was completely destroyed by removing an old mechanical valve or endocarditis in small roots. In those cases, once the aortic annulus was incised, a large patch to repair the aortomitral curtain significantly enlarged the aortic annulus to allow a bigger prosthetic valve. Sewing the patch to the mitral and aortic annulus is more secure compared with the Nicks and Manouguian procedures, which sew the patch to the aortomitral curtain—a thin layer of fibrous tissue. We recommended a “Y” incision instead of a “T” incision to avoid accidental incising the LA or MV since sometimes it is difficult to differentiate the aortomitral curtain, mitral annulus, and MV. The “Y” incision could be easily extended below the nadirs of aortic annulus from aortomitral curtain without requiring dissecting out of aortic root but should not pass the fibrous tissue under nadirs of both left and noncoronary sinuses. The rectangular patch could push on the left coronary sinus with a potential of distortion of left coronary artery. However, the transverse sinus behind the rectangular patch allows the patch and left coronary sinus to expand posteriorly without significant distortion of left coronary artery. Because of this potential risk, we do not recommend anything larger than the rectangular patch. In summary, our enlargement technique is simple and effective for avoiding patient–prosthesis mismatch and preparing patients for future valve-in-valve transcatheter AVR.
  3 in total

1.  Mitral regurgitation late after Manouguian's anulus enlargement and aortic valve replacement.

Authors:  K Imanaka; S Takamoto; A Furuse
Journal:  J Thorac Cardiovasc Surg       Date:  1998-03       Impact factor: 5.209

2.  Hypoplasia of the aortic root. The problem of aortic valve replacement.

Authors:  R Nicks; T Cartmill; L Bernstein
Journal:  Thorax       Date:  1970-05       Impact factor: 9.139

3.  Patch enlargement of the aortic valve ring by extending the aortic incision into the anterior mitral leaflet. New operative technique.

Authors:  S Manouguian; W Seybold-Epting
Journal:  J Thorac Cardiovasc Surg       Date:  1979-09       Impact factor: 5.209

  3 in total
  8 in total

1.  A Y Incision and Rectangular Patch to Enlarge the Aortic Annulus by Three Valve Sizes.

Authors:  Bo Yang; Aroma Naeem
Journal:  Ann Thorac Surg       Date:  2021-03-06       Impact factor: 5.102

2.  Commentary: Aortic root enlargement-optimizing for today and preparing for the future.

Authors:  Elizabeth L Norton; Kendra J Grubb
Journal:  JTCVS Tech       Date:  2022-02-21

3.  "Roof" technique-a modified aortotomy closure in Y-incision aortic root enlargement upsizing 3-4 valve sizes.

Authors:  Bo Yang; Aroma Naeem; Sarah Palmer
Journal:  JTCVS Tech       Date:  2022-01-26

4.  Transcatheter aortic valve replacement valve-in-valve: Future implications for the surgeon.

Authors:  Robert J Steffen; Vinayak N Bapat
Journal:  JTCVS Open       Date:  2022-02-24

5.  Stentless valves for bicuspid and tricuspid aortic valve disease.

Authors:  Bailey Brown; Tan Le; Aroma Naeem; Aroosa Malik; Elizabeth L Norton; Xiaoting Wu; Himanshu J Patel; G Michael Deeb; Karen M Kim; Bo Yang
Journal:  JTCVS Open       Date:  2021-09-28

Review 6.  Management of Failed Bioprosthetic Aortic Valves: Mitigating Complications and Optimizing Outcomes.

Authors:  Elizabeth L Norton; Alison F Ward; Adam Greenbaum; Kendra J Grubb
Journal:  J Interv Cardiol       Date:  2022-09-02       Impact factor: 1.776

7.  Commentary: Expertise in fitting square pegs in round holes.

Authors:  Fraser D Rubens
Journal:  JTCVS Tech       Date:  2020-11-25

8.  Commentary: Enlargement of the aortic annulus: Always a technical price to pay.

Authors:  Joel Price
Journal:  JTCVS Tech       Date:  2020-12-05
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.