Literature DB >> 33622544

Guidelines for the choice of circumferential wrought wire and cast clasp arms for removable partial dentures.

C Peter Owen1, Noland Naidoo2.   

Abstract

PURPOSE: The aim of this study was to establish guidelines for the selection of cast and wrought-wire clasps for removable partial dentures (RPDs) that would be appropriate for clinically encountered undercuts and facial curvatures.
METHODS: Randomly selected discarded casts were collected and 30 premolars and 30 molars were surveyed, sectioned to a line representing the clasp and scanned using a flatbed scanner. The average clasp curvature and length for each group was determined and a three-dimensional model printed, to which wrought wire clasps of 0.9- and 1.0-mm diameter were adapted. Standard wax clasp patterns were adapted and cast in a stellite alloy. Each clasp was deformed beyond its proportional limit; and the forces exerted at that limit and at deflections of 0.25 mm, 0.5 mm, and 0.75 mm were measured, and a safety limit was calculated that would ensure elastic deformation at the required undercuts.
RESULTS: A table was produced with guidelines for those clasps that would provide the highest retentive force within the proposed safety limit. The highest forces were provided by cast clasps in a 0.25-mm undercut. Wrought round wire of 1-mm diameter provided the next highest retentive forces, in a 0.25-mm undercut for premolar clasps arms and 0.5-mm for molar clasps.
CONCLUSIONS: The results provide valid guidelines for the use of combinations of clasp material and undercut that would exert the maximum retentive force without deformation for both short (premolar) as well as long (molar) clasps, for wrought and cast clasps.
Copyright © 2021 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Clasps; Forces; Removable partial denture; Safety limit; Undercuts

Mesh:

Substances:

Year:  2021        PMID: 33622544      PMCID: PMC9275311          DOI: 10.1016/j.identj.2021.01.005

Source DB:  PubMed          Journal:  Int Dent J        ISSN: 0020-6539            Impact factor:   2.607


Introduction

Tooth loss has a multifactorial aetiology and can have varying effects on an individual's quality of life. The epidemiological literature suggests that there may be a decreasing number of patients afflicted with complete edentulism,, but that tooth loss continues to occur and may result in clinicians having to treat a higher incidence of patients who are partially edentulous. Treatment modalities include removable partial dentures (RPDs), fixed partial dentures (FPDs), implant supported prostheses (ISPs), or no treatment. The last option is, however, dependent on the patient's ability to adequately function in their current state or their preferences. Globally, a variety of surveys have assessed need and usage of RPDs. A survey in the United States in 2002 found an increasing percentage of RPD wearing with age up to the >70 years cohort where there was a decrease, which was related to an increase in complete dentures. The authors estimated that at least 250,000 people younger than age 40 had RPDs. A study of self-reported data from 10,902 elderly subjects older than 60 years of age in 7 Latin American and Caribbean cities reported only on whether they had “bridges/dentures” and found that overall, 70% reported positively. Data from surveys in Europe and Israel found disparities between countries in terms of the average number of natural teeth remaining, which ranged from 15 to 27. Similarly there were country variations in the numbers of teeth replaced, related to the variable health services options available. A Chinese national oral health survey found that 27% in the 55 to 64 age group and 31% in the 65 to 74 age group wore RPDs. In a suburban Nigerian community, a survey of subjects older than 65 years of age revealed that 48% were partially edentulous but only 7% had RPDs, mainly because of financial constraints. In a national survey of a Polish population older than 65 years of age, 46% were partially edentulous, but 31% of those did not have RPDs. The replacement of missing teeth by means of RPDs is a well-established and cost-effective option. The responsibility for designing an RPD rests solely with the clinician and a detailed diagrammatic prescription should be provided to the dental technician., This prescription should be based on evidence or clinical guidelines. However, much of the design principles for RPDs are based on anecdotal reports or clinical experience rather than high levels of scientific evidence. This is also true for the selection of clasps for RPDs. Clasps used in RPD designs are derived from wrought wire or are cast components of an all-metal framework. These clasps need to be flexible enough to allow the RPD to be seated and removed numerous times without permanently deforming the clasp and without damaging the tooth.,15, 16, 17, 18, 19 The flexibility of a clasp is influenced by its length, diameter, cross-sectional form, and by its material.,18, 19, 20, 21, 22, 23 If the amount of force required to overcome an undercut is beyond the proportional limit of the clasp arm, either the tooth will be affected or the clasp arm will be permanently deformed or will fracture.,, Davenport et al reported that cast chrome-cobalt (Cr-Co) clasps needed to be at least 15mm to flex 0.25mm without permanently deforming and that wrought wire clasps needed to be at least 7 mm to overcome 0.5-mm undercuts without deforming. Their conclusions were, however, formulated from questionnaires that were sent to a selection of clinicians. Hence, their statements were a consensus of expert opinions and were not validated or substantiated. Warr used a mathematical analysis of clasp behaviour to predict loading forces relative to the proportional limit and suggested that the load exerted by moving in and out of an undercut should allow for a margin of safety and, in 1961, proposed that the reason why clasps deformed or fractured was that they function too close to their proportional limit. However this ‘safety factor,’ or margin from the proportional limit, was never identified or tested. Bates approached this from a statistical point of view and stated that a clasp should be selected to function where the force required to overcome the undercut is equal to the proportional limit of that material less 2 standard deviations and termed this the ‘realistic limit.’ It is logical that some form of limit should be applied to all clasp materials to avoid breakage, which is the likely outcome in cast metals, and permanent deformation, which is more likely in the more flexible wrought metals. Although there have been attempts in the literature to provide guidelines for RPD designs and clasp selection, the studies generally fall short of simulating the clinical situation, and extrapolation from these studies must be done with caution.,,20, 21, 22,, It was therefore decided to conduct a study on clasps that reflected clinically encountered tooth curvatures and lengths to ascertain the force exerted on deflection at clinically encountered undercuts. Additionally, the proportional limit in relation to this deflection was calculated, to test whether Bates's realistic limit was feasible and had clinical validity, or whether some other more clinically valid factor could act as a safety limit. The hypothesis was that there would be combinations of clasp materials, clasp lengths, and undercuts that would exceed the realistic limit as well as the proportional limit.

Methods and materials

Experimental design

This was an in vitro study designed to test a variety of clasp materials in terms of the forces they would exert when flexed at a deflection corresponding to the accepted undercuts present on premolar and molar teeth. The undercuts used were 0.25, 0.5, and 0.75 mm. Thirty premolars and molars were used and the clasp materials are shown in Table 1.
Table 1

Clasp materials tested in the study.

MaterialDiameter (mm)Number of specimens
PremolarMolar
Wrought wire
Leowire*0.91010
1.01010
Remanium Hard0.91010
1.01010
Noninium§0.91010
Cast
Vitallium1010

Leowire (Leone), stainless steel.

Remanium Hard (Dentaurum), stainless steel.

Noninium (Dentaurum), nickel-free stainless steel.

Only 0.9 mm available.

Vitallium (Dentsply), stellite alloy.

Clasp materials tested in the study. Leowire (Leone), stainless steel. Remanium Hard (Dentaurum), stainless steel. Noninium (Dentaurum), nickel-free stainless steel. Only 0.9 mm available. Vitallium (Dentsply), stellite alloy.

Sample size calculation

The relative margin of error was calculated and found to be acceptable for 30 premolars and 30 molars. The sample size for the different clasp materials and diameters was 10 per group. This was chosen in line with previous research, where the differences between the materials tested provided acceptable precision for the use of 10 samples per material ,,

Determining the average curvatures

Discarded casts from the dental laboratory attached to our prosthetics clinic were randomly and anonymously collected. Inclusion criteria were that they had undamaged first premolars and molars with no evidence of cavitation or fracture and excluded were casts where the premolars or molars showed any sign of damage. The casts were placed onto a dental surveyor and the maximum facial curvature (survey line) of each tooth was established in the normal manner. A line was then drawn to represent the position of a normal clasp arm. The teeth were then sectioned into separate dies and each die was trimmed to the surveyed line representing the clasp arm such that it would stand on a flat-bed scanner (Canon MG3540). The midpoint was marked to line up with the midpoint of a scanning template to establish a scale reference to ensure that magnification errors did not occur. Each die was scanned at 600 dpi. The images of the premolar and molar groups were then imported individually into the software package CorelDraw (Corel) and converted to mathematically derived vector graphics. The curves were then placed over each other, each onto a separate virtual layer. The homogeneity of the curves was such that an average curvature could be mapped, representing the determined average curvature for the teeth in that group.

Construction of the average curvature model

The average curvature images for the premolars and molars were imported into the software Solidworks (Dassault Systèmes Solidworks Corporation) and three-dimensional (3D) models of each curvature were created (Figure 1).
Fig. 1

Three-dimensional models of the average curvatures. (A) premolar and (B) molar.

Three-dimensional models of the average curvatures. (A) premolar and (B) molar. To adapt the wire samples, a solid model was required; so in the software, additional primitive objects were added, and the combined shape was extruded to allow adequate space for 10 wire samples to be adapted to the curve. A 5-mm ledge was created on the curve corresponding to the clasp tip to ensure that each sample terminated at the same point (Figure 2). Standard Tessellated Language (stl) files of these models were created and exported to a 3D printer (Objet 350v, Objet Inc.) and printed using the material Objet FullCure720 RGD720.
Fig. 2

Examples of (A) wrought wire clasps and (B) cast clasps adapted to the three-dimensional models.

Examples of (A) wrought wire clasps and (B) cast clasps adapted to the three-dimensional models.

Constructing the clasp samples

The various wrought wire and cast clasp samples were adapted to the 3D-printed models by a single dental laboratory technician. The wrought wires were adapted in the same manner as they would be on a cast for use in a normal RPD. The cast clasps were all from a standard wax pattern with identical taper and cross-sectional shape. They were invested in the same flask, and each clasp sample was analysed under magnification by a single operator to identify any defects. The clasps were then incorporated into acrylic resin blocks measuring 38 mm x 25 mm x 6 mm using a prefabricated mould; 10 samples were prepared at the same time. All the samples were checked again by a single operator to ensure that there was no mobility of the wires in the acrylic resin blocks.

Tensile testing

The clasps embedded in acrylic blocks were tested in a tensile testing machine (Instron Corporation) using the Bluehill Lite software program (Instron). The clasp tips were engaged in a custom-made jig and displaced at a cross-head speed of 0.5 mm/min using a 2kN load cell, which was auto-balanced before each test to eliminate any measurement caused by the clasp tip touching the jig. Each clasp arm was then displaced until its proportional limit was exceeded. All the testing was performed by a single operator; all 120 samples were tested on the same day under the same conditions. The software was programmed to record the proportional limit for each specimen as well as the load exerted at deflections of 0.25 mm, 0.5 mm, and 0.75 mm and was able to account for any slippage that occurred between the clasp tip and the jig platform during the initial movement of the load cell. This ensured that any deviation from this would represent the true proportional limit of the sample (Figure 3).
Fig. 3

Example of a graph generated during the tensile testing for a premolar cast clasp.

Example of a graph generated during the tensile testing for a premolar cast clasp.

Data analysis

For each experiment (combination of tooth type and clasp type), the univariate statistics (mean, standard deviation) for the force at 0.25-, 0.50-, and 0.75-mm deflection, as well as at the proportional limit, were calculated. The realistic limit (mean proportional limit minus 2 standard deviations) was calculated for each experiment and expressed as a percentage of the proportional limit. Similarly, a factor of the proportional limit less 1 standard deviation was calculated and, again, expressed as a percentage of the proportional limit. A 1-sample t-test was used to determine the extent of the difference between the mean loads at the different deflections and these limits.

Results

The lengths of the determined average curvature for the premolar and molar dies selected were 9 mm and 14.5 mm, respectively. None of the wrought wire samples fractured during testing, but 1 molar cast clasp and 5 of the premolar cast clasps fractured. Table 2 shows the results of the load tests for each combination of tooth type and clasp type. Table 3 shows the results for the calculations of Bates's realistic limit and is expressed as a percentage of the proportional limit for each wire. The mean percentage of the proportional limit was 63%, and it can be seen that at this limit that no cast clasps (Vitallium, Table 2) would be below this limit. As it is known clinically that cast clasps have been placed successfully on both premolars and molars, this realistic limit would appear to be too restrictive. Therefore, Table 3 also shows a proposed safety limit of 1 standard deviation from the proportional limit. The average difference of this limit from the proportional limit is 82%, and so it would seem reasonable to suggest that this now defines the safety limit. Table 4 then shows the effect of applying this safety limit to identify those materials that would fall within this limit and, therefore, could be used clinically. Fig. 4, Fig. 5 illustrate the force exerted at undercuts of 0.25 and 0.5 mm relative to the proportional limit and this safety limit.
Table 2

Tabulated results from the load tests for each experiment.

Clasp typeTooth typeProportional limit (g)
Deflection 0.25 mm (g)
Deflection 0.5 mm (g)
Deflection 0.75 mm (g)
nmeanSDRSD(%)nmeanSDRSD(%)nmeanSDRSD (%)nmeanSDRSD (%)
0.9 mm Leowirepremolar101003215211030345151058396161084915118
molar10121938832101062827102034924102886723
1.0 mm Leowirepremolar1011302332110676118171012133052510175542424
molar10856811010359621710657116181091615717
0.9 mm Remanium Hardpremolar1089621224104205613108001211510116217215
molar106997411101242117102353816103345416
1.0 mm Remanium Hardpremolar10858140161053566121010271141110147517312
molar1091514216102192712104175112106047212
0.9 mm Noniniumpremolar1057454910360651810619109181081813617
molar1070512317101603824102744717103635415
Vitalliumpremolar10145733323101179287241019274292210236442218
molar101310271211077396121014641671110198824512

RSD = relative standard deviation; SD = standard deviation.

Table 3

Mean PL less 2 SD (Bates's ‘RL’), and the mean PL less 1 SD, both expressed as percentages of the PL. The last column creates a ‘safety limit’ based on the mean percentage of the PL less 1 SD as being clinically valid.

Clasp typeTooth typeBates RL (g) (= mean PL less 2 SD)RL as % of PLMean PL less 1 SDPL less 1 SD as % of PLSafety limit: 82% of mean PL
0.9-mm Leowirepremolar5725778779822
molar44336831681000
1.0-mm Leowirepremolar6645989779927
molar6938177490702
0.9-mm Remanium Hardpremolar4735368576735
molar5507962589573
1.0-mm Remanium Hardpremolar5786771884703
molar6316977384751
0.9-mm Noniniumpremolar4678152091470
molar4586558283578
Vitalliumpremolar792541125771195
molar768591039791074
Mean63Mean82

PL = proportional limit; RL = realistic limit; SD = standard deviation.

Table 4

Analysis of force at deflection relative to the safety limit. Forces higher than the safety limit are shaded.

Clasp TypeTooth TypePLDeflection (g)
SL: 82% of PLP value for mean Defl < SL
0.25-mm0.5-mm0.75-mmDefl .25 mmDefl .50 mmDefl .75 mm
meanmeanmean
0.9-mm Leowirepremolar1003303583849822.00000.0000Defl > SL
molar12191062032881000.0000.0000.0000
1.0-mm Leowirepremolar113067612131755927.0001Defl > SLDefl > SL
molar856359657916702.0000.2486*Defl > SL
0.9-mm Remanium Hardpremolar8964208001162735.0000Defl > SLDefl > SL
molar699124235334573.0000.0000.0000
1.0-mm Remanium Hardpremolar85853510271475703.0000Defl > SLDefl > SL
molar915219417604751.0000.0000.0001
0.9-mm Noniniumpremolar574360619818470.0005Defl > SLDefl > SL
molar705160274363578.0000.0000.0000
Vitalliumpremolar14571179192723641195.8625Defl > SLDefl > SL
molar1310773146419881074.0000Defl > SLDefl > SL

Defl = deflection; PL = proportional limit; SL = safety limit.

Not significant: 77% of PL.

Not significant: 81% of PL.

Fig. 4

The force exerted by a short (premolar) 1.0-mm diameter stainless steel wire at undercuts of 0.25 and 0.5 mm relative to the MPL of the wire and the proposed MSL as being 82% of the MPL. MPL= mean proportional limit; MSL = mean safety limit.

Fig. 5

The force exerted by a short (premolar) cast wire at undercuts of 0.25 and 0.5 mm relative to the MPL of the wire and the proposed MSL as being 82% of the MPL. MPL= mean proportional limit; MSL = mean safety limit.

Tabulated results from the load tests for each experiment. RSD = relative standard deviation; SD = standard deviation. Mean PL less 2 SD (Bates's ‘RL’), and the mean PL less 1 SD, both expressed as percentages of the PL. The last column creates a ‘safety limit’ based on the mean percentage of the PL less 1 SD as being clinically valid. PL = proportional limit; RL = realistic limit; SD = standard deviation. Analysis of force at deflection relative to the safety limit. Forces higher than the safety limit are shaded. Defl = deflection; PL = proportional limit; SL = safety limit. Not significant: 77% of PL. Not significant: 81% of PL. The force exerted by a short (premolar) 1.0-mm diameter stainless steel wire at undercuts of 0.25 and 0.5 mm relative to the MPL of the wire and the proposed MSL as being 82% of the MPL. MPL= mean proportional limit; MSL = mean safety limit. The force exerted by a short (premolar) cast wire at undercuts of 0.25 and 0.5 mm relative to the MPL of the wire and the proposed MSL as being 82% of the MPL. MPL= mean proportional limit; MSL = mean safety limit. Two of the values were not statistically significantly different from the mean proportional limit: the deflection of the 1.0-mm Leowire at an undercut of 0.5 mm for a molar clasp and the deflection of the Vitallium (cast) clasp at an undercut of 0.25 mm for a premolar clasp. However the former was within 5% of the safety limit and the latter within 1% of the safety limit. Therefore, it is possible to construct a table as in Table 5 that provides the clinical guidelines for the highest forces exerted within the safety limit. In addition, although nickel sensitivity is rare, it nevertheless does exist, and so the nickel-free wire Noninium is included in Table 5 at the appropriate diameter for premolar and molar clasps.
Table 5

Clinical guidelines for the clasp selection for molars and premolars based on the highest loads exerted within the safety limit and on the wires tested.

PremolarsMolars
Undercut0.25 mm0.25 mm0.5 mm0.75 mm
Clasp material (mean force in grams)Vitallium (1179)Leowire 1.0 mm (676)Vitallium (773)Leowire 1.0-mm (657)
For nickel-sensitive patients (mean force in grams)Noninium 0.9 mm (360)Noninium 0.9 mm (363)
Clinical guidelines for the clasp selection for molars and premolars based on the highest loads exerted within the safety limit and on the wires tested.

Discussion

Permanent deformation and fatigue fracture are among the most common mechanical complications that can affect RPD clasps.31, 32, 33 The resultant loss of retention and reduced stability can compromise the comfort of the patient. An RPD clasp design based on sound knowledge of the behavioural characteristics of the various clasps materials and diameters should decrease the incidence of these mechanical complications. Studies have shown that clasps tend to lose their efficiency over time. In 1 study, clasps exposed to repeated stress underwent permanent deformation and fatigue over a 36-month period. This raises the question of whether constant deflection of the clasp during insertion and removal causes fatigue of the clasp or whether the clasp functioned too close to its proportional limit. Deformation of the clasp arm may lead to unfavourable stresses on the abutments and the RPD itself. Vallittu and Kokonen used cyclic fatigue testing on straight wires by deflecting them 0.6 mm. They found a decrease in load (but gave no figures) and fatigue fracture occurring at about 25 loading cycles. However, the clinical applicability of such tests is questionable. Saito et al, in a retrospective clinical evaluation of a variety of RPDs, found that clasp-retained dentures showed an increase in clasp fracture after 6 years but did not quantify this or identify the clasps. Cheng et al tested 3 different cobalt-chromium cast alloys in a circumferential clasp assembly configured on a metal molar tooth with 0.25- and 0.5-mm undercuts and subjected them to an insertion and removal test for 7200 cycles to simulate 5 years of clinical use. They found a “marked decrease” in retentive force after 360 cycles and, thereafter, a gradual decrease. However, the amount of decrease was not provided, and the load at deflection was only determined by displacing a straight wire and not a wire corresponding to a circumferential clasp's curve. Nevertheless, the load at an undercut of 0.5 mm exceeded the proportional limit but not that at an undercut of 0.25 mm Previous studies have made assumptions as to the length of the facial curvature of teeth. It has been reported that wrought wire clasps need to be at least 7 mm in length to overcome 0.5-mm undercuts without deforming, and that a cast clasp needs to be at least 15 mm because anything shorter would be too rigid to disengage 0.25-mm undercuts without permanently deforming the clasp or damaging the tooth. However, these statements were unsubstantiated by scientific evidence and were based on the clinical opinion of colleagues. The models created to represent the average curvature of molars and premolars in this study simulated the clinical length and curvature to which clasps for premolars and molars would be adapted. The average lengths were 9 mm for premolars and 14.5 mm for molars, and both length and diameter proved to be significant factors in the flexibility of the clasps. Not surprisingly, the retentive force of the clasps for a given undercut were also affected by the length of the clasp because the shorter the clasp and stiffer the clasp, the greater the force required for its flexure. The variations noted in the load exerted at the different deflections and in the proportional limits for the different materials and diameters indicate that there are inconsistencies in the manufacturing of these materials. Therefore, a margin of safety does need to be applied, but the concept of the realistic limit, was found to be clinically unrealistic. For this reason, it is proposed that a safety limit of 82% of the mean proportional limit be adopted as being clinically appropriate. Understanding the nature of forces that are involved in dislodging a denture is essential in determining the appropriate clasp and denture design for a given situation. However, these forces have not been measured clinically, so it would seem wise to choose clasp arms that will provide the greatest retention at all times, provided that those clasp arms can function elastically, within the safety limit proposed here. The clasp materials and the undercuts that result from this analysis are shown in Table 5 as a clinical guideline for clasp selection. The design referred to throughout is that of a circumferential clasp arm, so these guidelines would not apply to gingivally approaching clasps. The hypothesis was accepted in that it was found that several material, clasp length, and undercut combinations produced forces that exceeded the realistic limit and the proportional limit. The realistic limit was found to be too stringent, and a new safety limit was proposed as the average of the proportional limit minus 1 standard deviation. This study has shown that cast clasps should not exceed a 0.25-mm undercut and that even that was close to the proposed safety limit for premolars (or teeth of equivalent size). Casting the cast clasps together in 1 casting was a limitation of this study because this did not allow for natural variations that occur during casting. A follow-up, although labor-intensive study, where each clasp is part of a cast framework would allow for the natural variations that occur during the casting process and may be more clinically applicable. In addition, this study should be replicated with other makes of wrought and cast materials. The flexibility of wrought wire is influenced by its alloy type, diameter, length, and depth of undercut. In this study, the highest force exerted by a wrought wire and within the safety limit, was 676 g from a stainless steel wire of 1.0-mm diameter for a 0.25-mm undercut. For a molar clasp length, the highest force within the safety limit was 657 g from a stainless steel wire of 1.0-mm diameter engaging a 0.5-mm undercut. These results are difficult to compare with other studies, which have reported either clinician opinions, removal force without recording deformation, standardised curvatures, or straight wires, and often just to determine the proportional limit.,20, 21, 22 This is the first study to use clasps conforming to the natural curvatures of premolars and molars. In addition, wrought wire free of nickel was tested and found to be applicable but was more flexible than stainless steel and, therefore, exerted less retentive force. Finally, acrylic resin base RPDs are often regarded as interim prostheses, but with the provision of both anterior tooth support through acrylic resin rests, posterior tooth support through the use of, for example, half-round wire, and the use in particular of guide plane retention, wrought wire clasps need provide additional retention only when required. In a world characterised by increasing inequality and an economic order where the majority of patients who are partially edentulous are likely to have limited access to the more expensive treatment methods and even to cast metal framework RPDs, this type of RPD could then become a more permanent and cost-effective prosthesis.

Conclusion

Several combinations of clasp material, length, and undercut were found to be unsuitable because they resulted in clasp deformation at the required undercut that exceeded the proportional limit as well as the originally suggested realistic limit. The hypothesis that this would be the case was therefore accepted. A proposed safety limit of 82% of the proportional limit was proposed to ensure that clasp deflection would remain elastic, although a cast clasp at an undercut of 0.25 mm for a premolar clasp was within 1% of the safety limit and should be used with caution. The combinations of clasp material and undercut that would exert the maximum retentive force within this safety limit were found for both short (eg, premolar) and long (molar) wrought wire and cast clasps. These were produced in a summary table to provide clinical guidelines for clasp selection for RPDs.

Conflict of interest

None disclosed.
  34 in total

1.  Retention.

Authors:  J C Davenport; R M Basker; J R Heath; J P Ralph; P O Glantz
Journal:  Br Dent J       Date:  2000-12-23       Impact factor: 1.626

2.  Cyclic fatigue properties of cobalt-chromium alloy clasps for partial removable dental prostheses.

Authors:  Hui Cheng; Minrui Xu; Hai Zhang; Weiqing Wu; Ming Zheng; Xiurong Li
Journal:  J Prosthet Dent       Date:  2010-12       Impact factor: 3.426

3.  Complications and failures in removable partial dentures: a clinical evaluation.

Authors:  M Saito; K Notani; Y Miura; T Kawasaki
Journal:  J Oral Rehabil       Date:  2002-07       Impact factor: 3.837

4.  A comparison of tooth retention and replacement across 15 countries in the over-50s.

Authors:  Christian Stock; Hendrik Jürges; Jing Shen; Kayvan Bozorgmehr; Stefan Listl
Journal:  Community Dent Oral Epidemiol       Date:  2015-12-28       Impact factor: 3.383

5.  Data Resource Profile: the Survey of Health, Ageing and Retirement in Europe (SHARE).

Authors:  Axel Börsch-Supan; Martina Brandt; Christian Hunkler; Thorsten Kneip; Julie Korbmacher; Frederic Malter; Barbara Schaan; Stephanie Stuck; Sabrina Zuber
Journal:  Int J Epidemiol       Date:  2013-06-18       Impact factor: 7.196

6.  Status of Tooth Loss and Denture Restoration in Chinese Adult Population: Findings from the 4th National Oral Health Survey.

Authors:  Jing Guo; Jing Hao Ban; Gang Li; Xing Wang; Xi Ping Feng; Bao Jun Tai; De Yu Hu; Huan Cai Lin; Bo Wang; Yan Si; Chun Xiao Wang; Wen Sheng Rong; Wei Jian Wang; Shu Guo Zheng; Xue Nan Liu; Sheng Chao Wang
Journal:  Chin J Dent Res       Date:  2018

7.  Deflection fatigue of cobalt-chromium, titanium, and gold alloy cast denture clasp.

Authors:  P K Vallittu; M Kokkonen
Journal:  J Prosthet Dent       Date:  1995-10       Impact factor: 3.426

8.  Stress-relaxation testing. Part IV: Clasp pattern dimensions and their influence on clasp behavior.

Authors:  H F Morris; K Asgar; J S Brudvik; S Winkler; E P Roberts
Journal:  J Prosthet Dent       Date:  1983-09       Impact factor: 3.426

9.  A study of the flexibility of wrought wire clasps.

Authors:  R P Frank; J I Nicholls
Journal:  J Prosthet Dent       Date:  1981-03       Impact factor: 3.426

10.  Quality of removable partial dentures worn by the adult U.S. population.

Authors:  Susan K Hummel; Margaret A Wilson; Victoria A Marker; Martha E Nunn
Journal:  J Prosthet Dent       Date:  2002-07       Impact factor: 3.426

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.