Literature DB >> 33619639

The Interface Motion and Hydrodynamic Shear of the Liquid Slosh in Syringes.

Yuchen Zhang1, Dingding Han1, Zhongwang Dou1, Jean-Christophe Veilleux2, Galen H Shi2, David S Collins2, Pavlos P Vlachos1, Arezoo M Ardekani3.   

Abstract

PURPOSE: Interface motion and hydrodynamic shear of the liquid slosh during the insertion of syringes upon autoinjector activation may damage the protein drug molecules. Experimentally validated computational fluid dynamics simulations are used in this study to investigate the interfacial motion and hydrodynamic shear due to acceleration and deceleration of syringes. The goal is to explore the role of fluid viscosity, air gap size, syringe acceleration, syringe tilt angle, liquid-wall contact angle, surface tension and fill volume on the interface dynamics caused by autoinjector activation.
METHODS: A simplified autoinjector platform submerged in water is built to record the syringe and liquid motion without obstruction of view. The syringe kinematics is imported to the simulations based on OpenFOAM InterIsoFoam solver, which is used to study the effects of various physical parameters.
RESULTS: The simulations agree with experiments on the air-liquid interface profile and interface area. The interfacial area and the volume of fluid subject to high strain rate decrease with the solution viscosity, increase with the air gap height, syringe velocity, tilt angle and syringe wall hydrophobicity, and hardly change with the surface tension and liquid column height. The hydrodynamic shear mainly occurs near the syringe wall and entrained bubbles.
CONCLUSION: For a given dose of drug solution, the syringe with smaller radius and larger length will generate less liquid slosh. Reducing the air volume and syringe wall hydrophobicity are also helpful to reduce interface area and effective shear. The interface motion is reduced when the syringe axis is aligned with the gravitational direction.

Keywords:  air-liquid interface; autoinjector; liquid slosh; protein denaturation

Year:  2021        PMID: 33619639     DOI: 10.1007/s11095-021-02992-3

Source DB:  PubMed          Journal:  Pharm Res        ISSN: 0724-8741            Impact factor:   4.200


  2 in total

Review 1.  The effects of shear flow on protein structure and function.

Authors:  Innocent B Bekard; Peter Asimakis; Joseph Bertolini; Dave E Dunstan
Journal:  Biopolymers       Date:  2011-05-04       Impact factor: 2.505

2.  Distribution of silicone oil in prefilled glass syringes probed with optical and spectroscopic methods.

Authors:  Zai-Qing Wen; Aylin Vance; Fabian Vega; Xiaolin Cao; Bruce Eu; Robert Schulthesis
Journal:  PDA J Pharm Sci Technol       Date:  2009 Mar-Apr
  2 in total
  1 in total

Review 1.  Computational models for studying physical instabilities in high concentration biotherapeutic formulations.

Authors:  Marco A Blanco
Journal:  MAbs       Date:  2022 Jan-Dec       Impact factor: 5.857

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.