Vince J Lai1, Alan W Reynolds1, Michael Kindya2, John Konicek3, Sam Akhavan1. 1. Department of Orthopaedic Surgery, Allegheny Health Network, Pittsburgh, Pennsylvania, U.S.A. 2. Orthopaedic Associates of Osceola, Kissimmee, Florida, U.S.A. 3. Arthrex, Inc., Naples, Florida, U.S.A.
Abstract
PURPOSE: To biomechanically evaluate the use of the suture augmentation construct at time 0 of ACL reconstruction. METHODS: Eighty porcine knees underwent ACL reconstruction using 2 techniques for graft fixation: a single suspensory construct (SSC), performed with a femoral button and tibial interference screw; and a double suspensory construct (DSC), with a femoral and tibial button. Each fixation technique was performed on 40 porcine knees divided into 4 subgroups. The first group had a nonaugmented ACL reconstruction, the second group had an ACL reconstruction with suture augmentation, and the third and fourth groups were the same as the first and second groups, with the graft resected 80% to simulate graft weakening. Ultimate load, yield load, stiffness, cyclic displacement values, and mode of failure were recorded for each graft. RESULTS: In a weakened graft model with 80% graft resection, there was a significant increase in ultimate strength (P < .001), yield strength (P < .001), and cyclic displacement (P < .001) with suture augmentation. There was no significant increase in stiffness with suture augmentation with either construct (P = .278). In the setting of an intact graft, there were no differences in either SSC or DCS groups with or without suture augmentation. CONCLUSIONS: The addition of a suture to ACL reconstruction techniques resulted in minimal changes in baseline biomechanical characteristics while improving ultimate load, yield load, and cyclic displacement in a weakened graft model. CLINICAL RELEVANCE: Suture augmentation of ACL reconstruction may confer improved integrity of the graft and is worth consideration and future clinical study.
PURPOSE: To biomechanically evaluate the use of the suture augmentation construct at time 0 of ACL reconstruction. METHODS: Eighty porcine knees underwent ACL reconstruction using 2 techniques for graft fixation: a single suspensory construct (SSC), performed with a femoral button and tibial interference screw; and a double suspensory construct (DSC), with a femoral and tibial button. Each fixation technique was performed on 40 porcine knees divided into 4 subgroups. The first group had a nonaugmented ACL reconstruction, the second group had an ACL reconstruction with suture augmentation, and the third and fourth groups were the same as the first and second groups, with the graft resected 80% to simulate graft weakening. Ultimate load, yield load, stiffness, cyclic displacement values, and mode of failure were recorded for each graft. RESULTS: In a weakened graft model with 80% graft resection, there was a significant increase in ultimate strength (P < .001), yield strength (P < .001), and cyclic displacement (P < .001) with suture augmentation. There was no significant increase in stiffness with suture augmentation with either construct (P = .278). In the setting of an intact graft, there were no differences in either SSC or DCS groups with or without suture augmentation. CONCLUSIONS: The addition of a suture to ACL reconstruction techniques resulted in minimal changes in baseline biomechanical characteristics while improving ultimate load, yield load, and cyclic displacement in a weakened graft model. CLINICAL RELEVANCE: Suture augmentation of ACL reconstruction may confer improved integrity of the graft and is worth consideration and future clinical study.
Anterior cruciate ligament (ACL) injuries have an incidence of 200,000 annually in the United States, making ACL reconstruction one of the most common ligament reconstruction procedures. The percentage of ACL reconstructions requiring revision surgeries is relatively high, ∼10% to 15%.2, 3, 4, 5, 6 Prior research from the Multicenter ACL Revision Study (MARS) Group has shown the factors leading to ACL revision reconstructions to include trauma (32%), technical factors (24%), or biologic factors (7%). One concern that may contribute to failure of the graft is trauma to the graft during the initial phases of ligamentization.The ligamentization process can be prolonged, lasting ≤6 months until its final stages. During this time, the graft will undergo extensive remodeling, with a significant decrease in its mechanical properties.9, 10, 11, 12, 13 The ultimate strength of the graft during these phases may decrease by as much as 80% of its original strength, putting the graft at risk for plastic deformation or failure.,,Suture augmentation of ACL reconstructions has been used to increase the strength of the graft construct during the early revascularization and remodeling phase.. Cook et al. looked at using synthetic augmentation for ACL reconstructions in a live canine model and concluded that the biologic-synthetic load-sharing ACL construct prevented early failure and allowed for functional graft remodeling while avoiding problems with use of all-synthetic grafts. Furthermore, concerns exist that suture augmentation may stress shield the graft during the initial phases of healing.The purpose of this study was to biomechanically evaluate the use of the suture augmentation construct at time 0 of ACL reconstruction. It was hypothesized that the use of suture augmentation would be protective to the graft, especially in the setting of a compromised graft.
Methods
Specimen Preparation
We performed ACL reconstruction using 2 commonly used constructs in porcine knee specimens. Proximal femurs and distal tibias were potted in fiberglass resin. Our first model was a single suspensory construct (SSC) using button fixation in the femur and an interference screw for fixation in the tibia. Our second model was a double suspensory construct (DSC) using button fixation in both the femur and tibia. All porcine knees were stripped of all soft tissues before performing the ACL reconstructions.
Femoral Tunnel and Intra-Articular Length Determination
In both constructs, the femoral tunnel was made as a 20-mm socket using an outside-guide and Flip-Cutter drill (Arthrex, Naples, FL). We ensured that femoral intraosseous length was ≥30 mm to ensure a stout cortical wall for button fixation. Femoral fixation was performed using a TightRope button (Arthrex). The intra-articular graft length was determined by fixing the porcine leg in independent femoral and tibial clamps and using a ruler to make sure the tunnels were 30 mm apart. Demonstration of this intra-articular measurement is shown in Figure 1.
Fig 1
Demonstration of measurement of the intra-articular anterior cruciate ligament (ACL) length, as was performed on all porcine knees. The dissected specimen in this image includes the distal femur on the left of the image and proximal tibia in the bottom right.
Demonstration of measurement of the intra-articular anterior cruciate ligament (ACL) length, as was performed on all porcine knees. The dissected specimen in this image includes the distal femur on the left of the image and proximal tibia in the bottom right.
Tibial Tunnel
In the SCC group, the full tibial tunnel was drilled using a guide wire followed by a reamer. The graft was passed and fixed with a line-to-line Biocomposite interference screw (Arthrex).In the DSC group, a socket was drilled using a flip cutter drill. The socket was drilled to the step bullet on the guide, ensuring a 7-mm bone bridge for the TightRope button. This ensured that the socket would be long enough to allow for graft tensioning. Once the graft was secured on both the femoral and tibial side, the graft was then cycled through flexion-extension motion and retightened, and the sutures on the button were tied to secure the loop.
Graft Preparation
For the SSC group, we used a quadrupled bovine extensor tendon construct. The ends of each graft were sutured using 2-0 FiberWire (Arthrex). The graft was passed through the loop of the TightRope RT button, while ensuring equal length of the graft limbs.For the DSC group, a GraftLink (Arthrex) construct was made using bovine extensor tendon. The graft was mounted with TightRope RT button for the femur and a TightRope ABS for the tibia.
Suture Augmentation
In cases in which suture augmentation was used, FiberTape suture (Arthrex) was passed through the femoral button, ensuring equal length of sutures on each side of the graft. In the SSC model, the suture was brought through the cannulation of the tibial interference screw. Once the graft was fixed on the femoral and tibial side, the FiberTape suture was tensioned separately and fixed on the tibia with a 4.75-mm SwiveLock anchor (Arthrex). Demonstration of tibia fixation is shown in Figure 2 for this group.
Fig 2
Demonstration of the tibial fixation in a single suspensory construct, in which fixation was achieved with an interference screw and the suture augmentation with a more distally placed anchor. In this image, the entire proximal tibia has been dissected, with the fixation and tunnel on the medial side.
Demonstration of the tibial fixation in a single suspensory construct, in which fixation was achieved with an interference screw and the suture augmentation with a more distally placed anchor. In this image, the entire proximal tibia has been dissected, with the fixation and tunnel on the medial side.In the DSC model, the suture was tensioned separately and fixed on the tibia with a 4.75-mm SwiveLock anchor after the graft was tensioned and fixed on the femur and tibia. This technique of tibial fixation can be seen in Figure 3. A view of the complete reconstruction is shown in Figure 4.
Fig 3
Demonstration of tibial fixation in a double suspensory construct using a suture button. In this image, the tibia shaft is on the left of the image with the plateau on the right; it is viewed from the medial side.
Fig 4
The completed soft tissue anterior cruciate ligament (ACL) reconstruction with addition of suture augmentation. In this image as viewed from the lateral side, the distal femur is at the left of the image.
Demonstration of tibial fixation in a double suspensory construct using a suture button. In this image, the tibia shaft is on the left of the image with the plateau on the right; it is viewed from the medial side.The completed soft tissue anterior cruciate ligament (ACL) reconstruction with addition of suture augmentation. In this image as viewed from the lateral side, the distal femur is at the left of the image.
Graft Resection
For the groups undergoing graft resection, an 80% graft resection of each individual limb of the graft was performed. The limbs of the graft were measured with a digital caliper (Mitutoyo, Aurora, IL), and 80% of its diameter was resected using a scalpel. This resection technique is shown in Figure 5. Prior studies in the literature have shown that the ACL graft has significantly diminished strength properties in the postoperative stages. The amount of graft resection was chosen to allow for a definitive decrease in graft strength, beyond what has been shown in the literature.,,,
Fig 5
Demonstration of soft tissue anterior cruciate ligament (ACL) reconstruction undergoing measured resection to 80% of width. In the left image, the width of the graft is measured with calipers after reconstruction is complete. As shown in the image on the right, 80% of the width was then resected for this group. In these images, the distal femur is at the bottom of the image and the knee is slightly flexed.
Demonstration of soft tissue anterior cruciate ligament (ACL) reconstruction undergoing measured resection to 80% of width. In the left image, the width of the graft is measured with calipers after reconstruction is complete. As shown in the image on the right, 80% of the width was then resected for this group. In these images, the distal femur is at the bottom of the image and the knee is slightly flexed.
Group Allocation
A total of 80 porcine knees were tested, with 40 specimens allocated to each model. For each model, we divided the 40 specimens into 4 groups of 10 specimens each.The first group underwent a standard SSC or DSC reconstruction. The second group underwent a SSC or DSC reconstruction with a Fibertape suture augmentation. The third group underwent a SSC or DSC reconstruction with 80% resection, and the fourth group underwent a SSC or DSC reconstruction with 80% resection and Fibertape suture augmentation.
Specimen Testing
All specimens were mounted on a materials testing machine (E10000; Instron Corp., Norwood, MA) with the femur held vertically to the cross-head and the tibia held at 30° flexion to the testing surface. Samples were preconditioned for 10 sinusoidal cycles, at 10 to 50 N and 1 Hz. Sinusoidal cyclic loading of the samples was conducted for 500 cycles, at 50 to 250 N and 1 Hz, followed by a single cycle load to failure at 20 mm/min. Load and displacement data were recorded at 1000 Hz. Plastic cyclic displacement, stiffness, yield load, and ultimate load were measured directly from the load–displacement curves. Stiffness was measured from the slope of the linear portion of the load–displacement curve immediately after cyclic loading.
Power and Statistical Analysis
A post hoc power analysis of the ultimate load comparisons was performed using SigmaPlot version 11.0 (Systat Software, San Jose, CA). For α = 0.05, the power of the two-way ANOVA was 1.000, for both the presence of an internal brace and graft condition. Furthermore, the interaction between the presence of an internal brace and the condition of the graft had a power of .856.Statistical analysis was performed using two-way ANOVA, with P < .05 considered significant. Analysis of failure modes was performed using a Fisher exact test, with P < .05 considered significant.
Results
Single Suspensory Construct
Intact Graft Versus Intact Graft With Suture Augmentation
In the SSC model, augmentation with a suture did not result in a significant increase from baseline in biomechanical properties. There was no significant difference in ultimate strength (892 ± 116 versus 1000 ± 139 N; P = .057), yield strength (826 ± 122 versus 820 ± 186 N; P = .927), or cyclic displacement (5.3 ± 1.1 versus 4.4 ± 0.7 mm; P = .057). There was no significant difference in stiffness when a suture augmentation was performed (146 ± 28 versus 139 ± 28 N/mm; P = .172).
Resected Graft Versus Resected Graft With Suture Augmentation
When the graft was resected 80%, there was a significant difference in ultimate load (263 ± 82 versus 655 ± 149 N; P < .001), yield strength (263 ± 82 versus 623 ± 124 N; P < .001), and cyclic displacement (9.5 ± 1.6 versus 6.0 ± 1.0 mm; P < .001). There was no significant difference in stiffness when a suture augmentation was performed (85 ± 20 versus 121 ± 9 N/mm; P = .148).
Intact Graft Versus Resected Graft With Suture Augmentation
Adding a suture augmentation to an 80% resected graft did not fully restore the biomechanical properties of the construct in the SCC model. There was a significant difference in terms of ultimate strength (891 ± 116 versus 655 ± 149 N, P < .001) , yield strength (826 ± 122 versus 623 ± 124 N, P = .002), and cyclic displacement (5.3 ± 1.1 versus 6.0 ± 1.0 mm, P = .001) compared with an intact state. There was no significant difference in terms of stiffness (146 ± 28 versus 121 ± 9 N/mm). Single suspensory results are summarized in Table 1.
Table 1
Single Suspensory Construct Data and Comparisons
Factor
Intact Graft Only (a)
Intact Graft Suture Augmentation (b)
Graft 80% Resected (c)
Graft 80% Resected With Augmentation (d)
P Value
a vs. b
c vs. d
a vs. d
Ultimate strength (N)
891 ± 116
1000 ± 139
263 ± 82
655 ± 149
.057
<.001
<.001
Yield strength (N)
826 ± 122
820 ± 186
263 ± 82
623 ± 124
.927
<.001
.005
Cyclic displacement (mm)
5.3 ± 1.1
4.4 ± 0.7
9.5 ± 1.6
6 ± 1.0
.111
<.001
.117
Stiffness (N/mm)
146 ± 28
139 ± 28
85 ± 20
121 ± 9
.428
.148
.09
Data are mean ± standard deviation.
Single Suspensory Construct Data and ComparisonsData are mean ± standard deviation.
Mode of Failure
With an intact graft and no suture augmentation, in 80% of the cases, the mode of failure was the graft slipping past the interference screw. These results are summarized in Table 2. With the addition of suture augmentation, in 80% of cases, the mode of failure was the button breaking through the cortex. This finding was statistically significant (P = .023).
Table 2
Single Suspensory Construct Mode of Failure
Failure Mode
Intact Graft, No Suture
Intact Graft, Suture Augmentation
Resected Graft, No Suture
Resected Graft, Suture Augmentation
Graft slippage past screw
8
2
1
1
Button breaking/button pulled through bone
2
8
0
1
Graft failure
0
0
9
8
Data are n.
Single Suspensory Construct Mode of FailureData are n.With a resected graft and no suture augmentation, in 90% of cases, the failure mode was failure and stretching of the graft. With the addition of suture augmentation, in 80% of cases, the mode of failure was failure and stretching of the graft. This finding was not significant (P = 1.0).
Double Suspensory Construct
In the DSC model, augmentation with a suture resulted in no significant difference in ultimate strength, yield strength, or cyclic displacement. There was no significant difference in stiffness with suture augmentation.When the graft was resected 80%, there was a significant difference in ultimate strength, yield strength, and cyclic displacement with suture augmentation. There was no significant difference in stiffness when a suture augmentation was performed.Adding a suture augmentation to an 80% resected graft fully restored the biomechanical properties of the construct in the DSC model. There was no significant difference in terms of ultimate strength, yield strength, and cyclic displacement compared with an intact state. There was also no significant difference in terms of stiffness. Double suspensory results are summarized in Table 3.
Table 3
Double Suspensory Construct Data and Comparisons
Factor
Intact Graft Only (a)
Intact Graft Suture Augmentation (b)
Graft 80% Resected (c)
Graft 80% Resected With Augmentation (d)
P Value
a vs. b
c vs. d
a vs. d
Ultimate strength (N)
747 ± 86
850 ± 216
372 ± 139
798 ± 116
.327
<.001
.441
Yield strength (N)
680 ± 147
815 ± 212
371 ± 139
697 ± 115
.181
<.001
.810
Cyclic displacement (mm)
5.2 ± 1.9
3.8 ± 0.8
8.9 ± 3.3
4.9 ± 1.2
.286
.001
.735
Stiffness (N/mm)
118 ± 14
128 ± 22
79± 18
108 ± 20
.237
.022
.398
Data are mean ± standard deviation.
Double Suspensory Construct Data and ComparisonsData are mean ± standard deviation.With an intact graft and no suture augmentation, in 100% of the cases, the mode of failure was button breakage on the tibia. With the addition of suture augmentation, in 90% of cases, the mode of failure was the button breaking on the tibia. This finding was not statistically significant (P = 1.0). Listing of these findings can be found in Table 4.
Table 4
Double Suspensory Construct Mode of Failure
Failure Mode
Intact Graft, No Suture
Intact Graft, Suture Augmentation
Resected Graft, No Suture
Resected Graft, Suture Augmentation
Button breaking/button pulled through
10
9
0
5
Graft failure
0
0
10
5
Suture loop of TightRope broke at tibial button
0
1
0
0
Data are n.
Double Suspensory Construct Mode of FailureData are n.With a resected graft and no suture augmentation, in 100% of cases, the failure mode was failure and stretching of the GraftLink construct. With the addition of suture augmentation, in 50% of cases, the mode of failure was failure and stretching of the Graftlink. In the remaining 50%, the failure was the button breaking on the tibia. This finding was statistically significant (P = .033).
Discussion
The results of the current study demonstrate the biomechanical protective effect of suture augmentation in the setting of ACL reconstruction, especially in a weakened graft. This “weakened state” can be seen during the ligamentization phase of ACL reconstruction, which can last ≥6 months after surgery.,,The process of ligamentization has been well described in the literature and includes early-healing, proliferative, and maturation phases. Progression through these stages will include graft necrosis, revascularization, cellular repopulation, deposition of collagen fibers, and remodeling.,,,17, 18, 19, 20 When the graft is in the early healing stages (3 to 8 weeks postoperatively), there are regions of central necrosis leading to graft weakening., The amount of central necrosis in the ACL graft is more evident in animal studies as opposed to the human studies.,9, 10, 11, Human ligamentization studies have shown a less robust process of graft necrosis and remodeling; the time frame, however, has been shown to extend longer than in an animal model.,,,The reconstructions in the current study were performed using 2 commonly used reconstruction techniques. Our first model was a single-suspensory construct (SSC) model, using a femoral button and a tibial interference screw. When the graft was intact, suture augmentation of the graft did not result in significant increases of ultimate strength, yield strength, and cyclic displacement. Analysis of the mode of failure in suture augmentation with a tibial interference screw, however, demonstrated that suture augmentation may reduce graft slippage past the screw when the construct is exposed to tension forces.Our second model was a double-suspensory construct (DSC) model, using a button for both femoral and tibial fixation. When the graft was intact, there was a no significant difference in the biomechanical properties when suture augmentation was used. The limiting factor for failure is the strength of the button in both graft conditions.When the graft was resected, the addition of suture augmentation to both the SSC and DSC models significantly increased the biomechanical properties of the construct. In the setting of the DSC model, these properties were essentially normalized to the intact state. Most significantly, the suture augmentation significantly reduced the cyclic displacement of the construct, acting to protect the graft in the weakened state.One of the biggest concerns with the use of suture augmentation is the potential for stress shielding of the graft., Our current study demonstrated that at initial implantation, the use of suture augmentation did not increase the overall stiffness of the construct, and thus theoretically should not lead to stress shielding of the graft. This is important, because stress shielding is a concern for any ACL augmentation because of delayed graft integration. We believe an important aspect to prevent stress shielding is to tension and fix the ACL graft independently from the suture augmentation. With independently and appropriately tensioned graft and augmentation, the graft will see the bulk of the stresses, with the suture augmentation becoming relevant only when the forces may overwhelm the graft. Theoretically, this should protect the ACL graft during its weakest stage, preventing plastic deformation of the graft and allowing the patient to proceed through the early phase of ligamentization with a biomechanically stronger reconstruction.
Limitations
There are several limitations to the current study. First, this study used a cadaveric porcine model stripped of soft tissue attachments. The soft tissue envelope, notably the muscles surrounding the knee, can provide additional stability and strength to a graft despite the weakening during the ligamentization stage. Second, we simulated the weakened graft state through a simple resection of the graft, versus an intrinsic replacement of graft tissue by immature tissue starting with the peripheral moving centrally. Finally, our model only tested the construct at initial implantation and does not take into account the in vivo surroundings of the graft during the healing phases of ACL reconstruction. Further research should evaluate the clinical relevance and importance of use of a simple suture augment to protect the ACL reconstruction during the ligamentization phase in an in vivo model.
Conclusions
The addition of a suture to ACL reconstruction techniques resulted in minimal changes in baseline biomechanical characteristics while improving ultimate load, yield load, and cyclic displacement in a weakened graft model.
Authors: Andreas Weiler; Ricarda Peine; Alireza Pashmineh-Azar; Clemens Abel; Norbert P Südkamp; Reinhard F G Hoffmann Journal: Arthroscopy Date: 2002-02 Impact factor: 4.772
Authors: James L Cook; Pat Smith; James P Stannard; Ferris Pfeiffer; Keiichi Kuroki; Chantelle C Bozynski; Cristi Cook Journal: J Knee Surg Date: 2016-12-26 Impact factor: 2.757
Authors: W G Clancy; R G Narechania; T D Rosenberg; J G Gmeiner; D D Wisnefske; T A Lange Journal: J Bone Joint Surg Am Date: 1981-10 Impact factor: 5.284
Authors: Toan D Duong; Dung T Tran; Bich N T Do; Tai T Nguyen; Son M Le; Ha H Le Journal: Asia Pac J Sports Med Arthrosc Rehabil Technol Date: 2022-07-01