Literature DB >> 33614603

Cyclic Damage Accumulation in the Femoral Constructs Made With Cephalomedullary Nails.

Farah Hamandi1, Alyssa Whitney1, Mark H Stouffer2, Michael J Prayson2, Jörn Rittweger3, Tarun Goswami1.   

Abstract

Background: The purpose of this study was to evaluate the risk of peri-prosthetic fracture of constructs made with cephalomedullary (CM) long and short nails. The nails were made with titanium alloy (Ti-6Al-4V) and stainless steel (SS 316L).
Methods: Biomechanical evaluation of CM nail constructs was carried out with regard to post-primary healing to determine the risk of peri-implant/peri-prosthetic fractures. Therefore, this research comprised of, non-fractured, twenty-eight pairs of cadaveric femora that were randomized and implanted with four types of fixation CM nails resulting in four groups. These constructs were cyclically tested in bi-axial mode for up to 30,000 cycles. All the samples were then loaded to failure to measure failure loads. Three frameworks were carried out through this investigation, Michaelis-Menten, phenomenological, and probabilistic Monte Carlo simulation to model and predict damage accumulation. Findings: Damage accumulation resulting from bi-axial cyclic loading in terms of construct stiffness was represented by Michaelis-Menten equation, and the statistical analysis demonstrated that one model can explain the damage accumulation during cyclic load for all four groups of constructs (P > 0.05). A two-stage stiffness drop was observed. The short stainless steel had a significantly higher average damage (0.94) than the short titanium nails (0.90, P < 0.05). Long titanium nail group did not differ substantially from the short stainless steel nails (P > 0.05). Results showed gender had a significant effect on load to failure in both torsional and bending tests (P < 0.05 and P < 0.001, respectively). Interpretation: Kaplan-Meier survival analysis supports the use of short titanium CM nail. We recommend that clinical decisions should take age and gender into consideration in the selection of implants.
Copyright © 2021 Hamandi, Whitney, Stouffer, Prayson, Rittweger and Goswami.

Entities:  

Keywords:  biomechanics; cephalomedullary nail; damage accumulation; femur; gender

Year:  2021        PMID: 33614603      PMCID: PMC7894258          DOI: 10.3389/fbioe.2020.593609

Source DB:  PubMed          Journal:  Front Bioeng Biotechnol        ISSN: 2296-4185