Literature DB >> 33606622

Path Replanning for Orientation-Constrained Needle Steering.

Marlene Pinzi, Thomas Watts, Riccardo Secoli, Stefano Galvan, Ferdinando Rodriguez Y Baena.   

Abstract

INTRODUCTION: Needle-based neurosurgical procedures require high accuracy in catheter positioning to achieve high clinical efficacy. Significant challenges for achieving accurate targeting are (i) tissue deformation (ii) clinical obstacles along the insertion path (iii) catheter control.
OBJECTIVE: We propose a novel path-replanner able to generate an obstacle-free and curvature bounded three-dimensional (3D) path at each time step during insertion, accounting for a constrained target pose and intraoperative anatomical deformation. Additionally, our solution is sufficiently fast to be used in a closed-loop system: needle tip tracking via electromagnetic sensors is used by the path-replanner to automatically guide the programmable bevel-tip needle (PBN) while surgical constraints on sensitive structures avoidance are met.
METHODS: The generated path is achieved by combining the "Bubble Bending" method for online path deformation and a 3D extension of a convex optimisation method for path smoothing.
RESULTS: Simulation results performed on a realistic dataset show that our replanning method can guide a PBN with bounded curvature to a predefined target pose with an average targeting error of 0.65  ± 0.46 mm in position and 3.25  ± 5.23 degrees in orientation under a deformable simulated environment. The proposed algorithm was also assessed in-vitro on a brain-like gelatin phantom, achieving a target error of 1.81  ± 0.51 mm in position and 5.9  ± 1.42 degrees in orientation.
CONCLUSION: The presented work assessed the performance of a new online steerable needle path-planner able to avoid anatomical obstacles while optimizing surgical criteria. SIGNIFICANCE: This method is particularly suited for surgical procedures demanding high accuracy on the desired goal pose under tissue deformations and real-world inaccuracies.

Year:  2021        PMID: 33606622     DOI: 10.1109/TBME.2021.3060470

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  2 in total

1.  Modular robotic platform for precision neurosurgery with a bio-inspired needle: System overview and first in-vivo deployment.

Authors:  Riccardo Secoli; Eloise Matheson; Marlene Pinzi; Stefano Galvan; Abdulhamit Donder; Thomas Watts; Marco Riva; Davide Danilo Zani; Lorenzo Bello; Ferdinando Rodriguez Y Baena
Journal:  PLoS One       Date:  2022-10-19       Impact factor: 3.752

Review 2.  Insights into Infusion-Based Targeted Drug Delivery in the Brain: Perspectives, Challenges and Opportunities.

Authors:  Asad Jamal; Tian Yuan; Stefano Galvan; Antonella Castellano; Marco Riva; Riccardo Secoli; Andrea Falini; Lorenzo Bello; Ferdinando Rodriguez Y Baena; Daniele Dini
Journal:  Int J Mol Sci       Date:  2022-03-15       Impact factor: 5.923

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.