Automatic dosimetric verification of online adapted plans on the Unity MR-Linac using 3D EPID dosimetry.

Igor Olaciregui-Ruiz1, Begoña Vivas-Maiques2, Sandra van der Velden2, Marlies E Nowee2, Ben Mijnheer2, Anton Mans2.   

Abstract

BACKGROUND AND
PURPOSE: The Unity MR-Linac is equipped with an EPID, the images from which contain information about the dose delivered to the patient. The purpose of this study was to introduce a framework for the automatic dosimetric verification of online adapted plans using 3D EPID dosimetry and to present the obtained dosimetric results.
MATERIALS AND METHODS: The framework was active during the delivery of 1207 online adapted plans corresponding to 127 clinical IMRT treatments (74 prostate, 19 rectum, 19 liver and 15 lymph node oligometastases). EPID reconstructed dose distributions in the patient geometry were calculated automatically and then compared to the dose distributions calculated online by the treatment planning system (TPS). The comparison was performed by γ-analysis (3% global/2mm/10% threshold) and by the difference in median dose to the high-dose volume (ΔHDVD50). 85% for γ-pass rate and 5% for ΔHDVD50 were used as tolerance limit values.
RESULTS: 93% of the online plans were verified automatically by the framework. Missing EPID data was the reason for automation failure. 91% of the verified plans were within tolerance.
CONCLUSION: Automatic dosimetric verification of online adapted plans on the Unity MR-Linac is feasible using in vivo 3D EPID dosimetry. Almost all online adapted plans were approved automatically by the framework. This newly developed framework is a major step forward towards the clinical implementation of a permanent safety net for the entire online adaptive workflow.
Copyright © 2021 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  EPID; In vivo dosimetry; MR-Linac; Online adaptation; QA; Unity

Mesh:

Year:  2021        PMID: 33582193     DOI: 10.1016/j.radonc.2021.01.037

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  1 in total

1.  Clinical rationale for in vivo portal dosimetry in magnetic resonance guided online adaptive radiotherapy.

Authors:  Begoña Vivas Maiques; Igor Olaciregui Ruiz; Tomas Janssen; Anton Mans
Journal:  Phys Imaging Radiat Oncol       Date:  2022-06-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.