Literature DB >> 33582146

High-resolution, non-contrast-enhanced magnetic resonance angiography of the wrist, hand and digital arteries using optimized implementation of Cartesian quiescent interval slice selective (QISS) at 1.5 T.

Mona Salehi Ravesh1, Annett Lebenatus2, Alexandra Bonietzki2, Johannes Hensler2, Ioannis Koktzoglou3, Robert R Edelman4, Joachim Graessner5, Olav Jansen2, Marcus Both2.   

Abstract

PURPOSE: Non-contrast-enhanced (CE) magnetic resonance angiography (MRA) techniques are of considerable interest for diagnosing vascular diseases in the upper extremities owing to the possibility of repeated examinations, sufficient coverage of the measurement volume, and because possible side effects of administering iodine- or gadolinium-based contrast agents and radiation exposure can be avoided. The aim of this study was to investigate the feasibility of an optimized electrocardiogram (ECG) triggered Cartesian quiescent interval slice selective (QISS) technique for MRA of hand arteries.
MATERIAL AND METHODS: Both hands of 20 healthy volunteers (HVs) were examined using an optimized QISS-MRA pulse sequence at 1.5 Tesla. The wrist and hand arterial trees were divided into 36 segments. Cross-sectional areas (CSA) of all arterial segments were measured. For the technical evaluation of the pulse sequence, the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were computed and six imaging artifacts were graded. Two experienced observers used an ordinal scoring system to assess the image quality of each arterial segment. Interobserver agreement was determined.
RESULTS: The median CSA was 7.3 mm2 in the ulnar and radial artery, 3.2 mm2 in the four common digital arteries, and 1.5 mm2 in five proper digital arteries. The median SNR and CNR of the third common proper arteries were 45.9 and 20.3, respectively. None of the arterial segments were contaminated by venous enhancement. The image quality of arterial segments for both hands was considered as diagnostic in 87.2% of all 1440 segments. An interobserver agreement of 0.67 for both hands was determined for image quality of arterial segments using a five-grade scoring system. Optimized QISS-MRA allows as the first MRA technique the classification of superficial palmar arch (SPA) and deep palmar arch (DPA) variants. 5 new SPA and 6 new DPA variants could be classified using QISS-MRA in comparison with previous studies using CE computed tomography angiography and using fixed cadaver hands.
CONCLUSIONS: By using this optimized 2D Cartesian QISS-MRA protocol, contrast agent-free angiography of the wrist and hand arteries provided a high in-plane spatial resolution and an excellent visualization of small digital arteries.
Copyright © 2021 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Cartesian quiescent interval slice selective (QISS); Deep palmar arch; Non-contrast enhanced magnetic resonance angiography; Superficial palmar arch; Wrist, hand and digital arteries

Mesh:

Substances:

Year:  2021        PMID: 33582146      PMCID: PMC7979532          DOI: 10.1016/j.mri.2021.02.008

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  42 in total

1.  Feasibility of non-contrast-enhanced magnetic resonance angiography for imaging upper extremity vasculature prior to vascular access creation.

Authors:  A S Bode; R N Planken; M A G Merkx; F M van der Sande; L Geerts; J H M Tordoir; T Leiner
Journal:  Eur J Vasc Endovasc Surg       Date:  2011-11-08       Impact factor: 7.069

2.  Bilateral ce-MR angiography of the hands at 3.0 T and 1.5 T: intraindividual comparison of quantitative and qualitative image parameters in healthy volunteers.

Authors:  Jan Thorsten Winterer; Olaf Moske-Eick; Michael Markl; Alexander Frydrychowicz; Thorsten Alexander Bley; Mathias Langer
Journal:  Eur Radiol       Date:  2007-11-27       Impact factor: 5.315

Review 3.  Nonenhanced MR angiography.

Authors:  Mitsue Miyazaki; Vivian S Lee
Journal:  Radiology       Date:  2008-07       Impact factor: 11.105

4.  Upper Extremity Angiographic Patterns in Systemic Sclerosis: Implications for Surgical Treatment.

Authors:  Jacinta Leyden; Matthew B Burn; Victor Wong; Daniel Sotelo Leon; Yukitoshi Kaizawa; Lorinda Chung; James Chang
Journal:  J Hand Surg Am       Date:  2019-02-21       Impact factor: 2.230

5.  High permittivity ceramics improve the transmit field and receive efficiency of a commercial extremity coil at 1.5 Tesla.

Authors:  Irena Zivkovic; Wouter Teeuwisse; Alexey Slobozhanyuk; Elizaveta Nenasheva; Andrew Webb
Journal:  J Magn Reson       Date:  2018-12-15       Impact factor: 2.229

6.  Evaluation of anatomy and variations of superficial palmar arch and upper extremity arteries with CT angiography.

Authors:  Hatice Kaplanoglu; Osman Beton
Journal:  Surg Radiol Anat       Date:  2016-09-19       Impact factor: 1.246

7.  Rice pads: novel devices for homogeneous fat suppression in the knee.

Authors:  Susumu Moriya; Yukio Miki; Tsuneo Yokobayashi; Mitsunori Kanagaki; Yoshiaki Komori; Mitsunori Ishikawa
Journal:  Acta Radiol       Date:  2010-03       Impact factor: 1.990

Review 8.  MR angiography of the hand and wrist.

Authors:  N M Rofsky
Journal:  Magn Reson Imaging Clin N Am       Date:  1995-05       Impact factor: 2.266

9.  Handedness may be related to variations in palmar arterial arches in humans.

Authors:  Aniruddha Sarkar; Sumita Dutta; Kana Bal; Jayanta Biswas
Journal:  Singapore Med J       Date:  2012-06       Impact factor: 1.858

Review 10.  Transradial Artery Access Complications.

Authors:  Yader Sandoval; Malcolm R Bell; Rajiv Gulati
Journal:  Circ Cardiovasc Interv       Date:  2019-11-01       Impact factor: 6.546

View more
  1 in total

1.  Quiescent-Interval Single-Shot Magnetic Resonance Angiography May Outperform Carbon-Dioxide Digital Subtraction Angiography in Chronic Lower Extremity Peripheral Arterial Disease.

Authors:  Judit Csőre; Ferenc Imre Suhai; Marcell Gyánó; Ákos András Pataki; Georgina Juhász; Milán Vecsey-Nagy; Dániel Pál; Daniele Mariastefano Fontanini; Ákos Bérczi; Csaba Csobay-Novák
Journal:  J Clin Med       Date:  2022-08-01       Impact factor: 4.964

  1 in total