| Literature DB >> 33576118 |
Cathryn R Cadwell1, Sarah Bowman1, Zoltan G Laszik1, Melike Pekmezci1.
Abstract
Conversion of glass slides to digital images is necessary to capitalize on advances in computational pathology and could potentially transform our approach to primary diagnosis, research, and medical education. Most slide scanners have a limited maximum scannable area and utilize proprietary tissue detection algorithms to selectively scan regions that contain tissue, allowing for increased scanning speed and reduced file size compared to scanning the entire slide at high resolution. However, very small and faintly stained tissue fragments may not be recognized by these algorithms, leading to loss of fidelity in the digital image compared to the glass slides. Cavitron ultrasonic surgical aspirator (CUSA) is frequently used in brain tumor resections, resulting in highly fragmented specimens that are used for primary diagnosis. Here we evaluated the rate of loss of fidelity in 296 digital images from 40 CUSA-resected brain tumors scanned using a Philips Ultra Fast Scanner. Overall, 54% of the slides (at least one from every case) showed loss of fidelity, with at least one tissue fragment not scanned at high resolution. The majority of the missed tissue fragments were small (<0.5 mm), but rare slides were missing fragments greater than 5 mm in greatest dimension. In addition, 19% of the slides with missing tissue showed no indication of loss of fidelity in the digital image itself; the missing tissue could only be appreciated upon review of the glass slides. These results highlight a potential liability in the use of digital images for primary diagnosis in CUSA-resected brain tumor specimens.Entities:
Keywords: digital pathology; scanning infidelity; tissue detection; whole-slide image
Mesh:
Year: 2021 PMID: 33576118 PMCID: PMC8412125 DOI: 10.1111/bpa.12938
Source DB: PubMed Journal: Brain Pathol ISSN: 1015-6305 Impact factor: 6.508
FIGURE 1Loss of fidelity in scanned whole‐slide images (WSI) is frequent and cannot always be appreciated without review of the glass slide. (A) Example WSI showing loss of fidelity with a large amount of unscanned tissue attributed to ROI selection. (B) Example WSI showing loss of fidelity, with multiple unscanned tissue fragments, attributed to a combination of ROI selection and MSA. (C) Summary of the frequency of infidelity (unscanned tissue) across all WSI examined (n = 296). Slides with missing tissue only beyond the coverslipped area were not counted as a loss of fidelity. (D) The manufacturer‐defined the maximum scannable area (MSA, gray striped area) of each slide as the area at least 3mm from all edges of the coverslip (dashed line) and 5mm from the edge closest to the slide label. (E) Summary of the likely cause of infidelity (n = 160 WSI with unscanned tissue). (F) Percent of WSI with unscanned small (<0.5 mm) or large (>=0.5 mm) tissue fragments that are attributed to MSA only (n = 3/108 [2.8%] for small fragments and n = 7/53 [13.5%] for large fragments; *p < 0.05, Fisher’s exact test). (G) Example WSI showing loss of fidelity with unscanned tissue that can only be detected on the glass slide. (H) Summary of the frequency of different modes of identification of scanning infidelity across all slides (n = 296). (I) Comparison of the frequency of different modes of identification of scanning infidelity between small and large unscanned tissue fragments (n = 39, 42, and 27 out of 108 total WSI with small unscanned tissue fragments, n = 27, 21, and 4 out of 52 total WSI with large unscanned tissue fragments; p = 0.023, overall χ 2 for 3 × 2 contingency table; post‐hoc comparisons for each modality compared to the other two combined using χ 2 with Bonferroni correction for multiple comparisons; *p < 0.05 compared to the frequency of small unscanned fragments being detected on glass only). All stains are hematoxylin and eosin. In (A), (B) and (G), black boxes denote areas scanned at high resolution; red lines highlight unscanned fragments that could be seen on the WSI; blue ellipses indicate unscanned fragments that could only be seen on review of the glass slides; scale bars are 5 mm. Error bars in (F) and (I) are 90% Clopper‐Pearson confidence intervals