In vitro assessment of a novel additive manufactured titanium implant abutment.

Les Kalman1.   

Abstract

BACKGROUND: Dental implant temporization remains a clinical challenge. A novel abutment simplifying the procedure was developed, but conventional fabrication was costly and unpredictable. A workflow was developed to fabricate the novel abutment using additive manufacturing. This in vitro investigation assessed the additive manufactured (AM) novel abutments to conventionally manufactured (CM) abutments.
MATERIAL AND METHODS: The AM abutments were fabricated in dental-grade titanium (Ti-6Al-4V) using Selective Laser Melting and were post-processed. The CM abutments were milled and subsequently laser welded manually. Pin strength of the abutments was assessed using a universal loading machine. Torque was measured by tightening the AM and CM abutments into dental implants within artificial bone.
RESULTS: Average pin strength was 364.4 N for the AM abutments and 62.5 N for the CM abutments. Average torque was 49.9 Ncm for the AM abutments and 62.9 Ncm for the CM abutments. AM abutment's pin strength was higher than the CM abutments. CM abutments measured a higher torque than the AM abutments.
CONCLUSIONS: Additive manufacturing with titanium using SLM provided an alternative fabrication pathway of a novel implant abutment. The AM approach was cost-effective, predictable, efficient and demonstrated pin strength and torque suitable for temporization procedures in implant dentistry. Key words:Abutment, dental implant, temporization, medical device, prototype design, additive manufacturing, 3D printing, titanium. Copyright:
© 2021 Medicina Oral S.L.

Entities:  

Year:  2021        PMID: 33574993      PMCID: PMC7864368          DOI: 10.4317/jced.57389

Source DB:  PubMed          Journal:  J Clin Exp Dent        ISSN: 1989-5488


  12 in total

1.  Damping behavior of implant-supported restorations.

Authors:  Pascal Magne; Michael Silva; Elisa Oderich; Luis Leonildo Boff; Reyes Enciso
Journal:  Clin Oral Implants Res       Date:  2011-10-03       Impact factor: 5.977

2.  A digital approach for one-step formation of the supra-implant emergence profile with an individualized CAD/CAM healing abutment.

Authors:  Tim Joda; Marco Ferrari; Urs Braegger
Journal:  J Prosthodont Res       Date:  2016-02-09       Impact factor: 4.642

3.  Success and failure rates of osseointegrated implants in function in regenerated bone for 72 to 133 months.

Authors:  Paul A Fugazzotto
Journal:  Int J Oral Maxillofac Implants       Date:  2005 Jan-Feb       Impact factor: 2.804

Review 4.  Fixed and removable provisional options for patients undergoing implant treatment.

Authors:  Sang-Choon Cho; Saphal Shetty; Stuart Froum; Nicolas Elian; Dennis Tarnow
Journal:  Compend Contin Educ Dent       Date:  2007-11

5.  Immediate provisionalization of single-tooth implants in fresh-extraction sites at the maxillary esthetic zone: up to 6 years of follow-up.

Authors:  Eitan Mijiritsky; Ofer Mardinger; Ziv Mazor; Gavriel Chaushu
Journal:  Implant Dent       Date:  2009-08       Impact factor: 2.454

6.  Relation between initial implant stability quotient and bone-implant contact percentage: an in vitro model study.

Authors:  Heng-Li Huang; Ming-Tzu Tsai; Kuo-Chih Su; Yu-Fen Li; Jui-Ting Hsu; Chih-Han Chang; Lih-Jyh Fuh; Aaron Yu-Jen Wu
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2012-08-24

7.  An apparatus for measuring the forces of dental extraction.

Authors:  A J MacGregor; G E Tomlinson
Journal:  Br J Oral Surg       Date:  1979-07

8.  The influence of thread geometry on implant osseointegration under immediate loading: a literature review.

Authors:  Hyo-Sook Ryu; Cheol Namgung; Jong-Ho Lee; Young-Jun Lim
Journal:  J Adv Prosthodont       Date:  2014-12-17       Impact factor: 1.904

9.  Placement of single tooth implant in healed socket with immediate temporization: Clinical study.

Authors:  Jeevan Lata; Monika Parmar
Journal:  Contemp Clin Dent       Date:  2012-10

Review 10.  Dealing with dental implant failures.

Authors:  Liran Levin
Journal:  J Appl Oral Sci       Date:  2008 May-Jun       Impact factor: 2.698

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.