Management of proximal shunt obstruction. Technical note.

C C Duncan1.   

Abstract

Proximal shunt obstruction or obstruction of the ventricular catheter may present with signs and symptoms of shunt failure with either no cerebrospinal fluid flow or a falsely low intracranial pressure (ICP) upon shunt tap. The author reports a technique for lowering the ICP and for measuring the pressure in patients with such obstruction by cannulation of the reservoir and ventricular catheter to penetrate into the ventricle with a 3 1/2-in. No. 22 spinal needle. The findings in 20 cases in which this approach was utilized are summarized.