Measurement of arterial pressure using catheter-transducer systems. Improvement using the Accudynamic.

M W Allan1, W M Gray, A J Asbury.   

Abstract

The dynamic performance of arterial manometers depends upon the values of the natural frequency fo and the damping factor beta. Satisfactory reproduction of all arterial waveforms requires that fo should be greater than about 13 Hz; for an fo of 13 Hz, beta should have a value of 0.5, but, as fo increases, an increasing range of beta values is permissible. We measured fo and beta for two pressure transducers (Gould P23 Statham and Gould Disposable) in combination with Portex or Gould catheter tubing of lengths varying from 100 cm to 300 cm. All combinations were considerably underdamped, with the best having an fo of 30 Hz and a beta of 0.1. We also investigated the performance of a commercial device, the Accudynamic, designed to allow beta to be increased without affecting fo. We found that the Accudynamic worked well, allowing acceptable performance to be obtained from the Gould Disposable transducer with catheters up to 200 cm in length.

Mesh:

Year:  1988        PMID: 3355737     DOI: 10.1093/bja/60.4.413

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  4 in total

1.  The mean prehospital machine; accurate prehospital non-invasive blood pressure measurement in the critically ill patient.

Authors:  Sandy Muecke; Andrew Bersten; John Plummer
Journal:  J Clin Monit Comput       Date:  2010-06-08       Impact factor: 2.502

2.  A comparison of measurements of blood pressure, heart-rate and oxygenation during inter-hospital transport of the critically ill.

Authors:  C J Runcie; W Reeve; J Reidy; J R Dougall
Journal:  Intensive Care Med       Date:  1990       Impact factor: 17.440

Review 3.  Practical points in the application of oxygen transport principles.

Authors:  P Nightingale
Journal:  Intensive Care Med       Date:  1990       Impact factor: 17.440

4.  Catheter-induced errors in pressure measurements in vessels: an in-vitro and numerical study.

Authors:  Adelaide de Vecchi; Rachel E Clough; Nicholas R Gaddum; Marcel C M Rutten; Pablo Lamata; Tobias Schaeffter; David A Nordsletten; Nicolas P Smith
Journal:  IEEE Trans Biomed Eng       Date:  2014-06       Impact factor: 4.538

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.