Literature DB >> 33551440

Quantification of 23 Volatile Organic Compounds with a Single Reference Material Using Post-column Reaction Gas Chromatography Combined with a Stainless-steel Heating Furnace.

Yuko Kitamaki1, Naoki Saito1, Nanako Sasaki2, Mariko Morita2, Tomohiro Sasaki2, Hideaki Miyamoto2, Masahiko Numata1, Toshihide Ihara1.   

Abstract

We built a heating furnace using stainless-steel instead of aluminum in gas chromatography combined with an oxidation/reduction system; it increased the oxidation temperature to 650°C. At 600°C, it completely oxidized five organochlorine compounds. This system was applied to a standard solution of 23 volatile organic compounds. The analytical results of 20 hydrocarbon and organochlorine compounds showed good agreement with the expanded uncertainty (k = 2) of the reference values. Three organobromine compounds obtained values higher than the reference; this was investigated further.

Keywords:  Post-column reaction gas chromatography; organobromine compounds; organochlorine compounds; volatile organic compound

Year:  2021        PMID: 33551440     DOI: 10.2116/analsci.20N036

Source DB:  PubMed          Journal:  Anal Sci        ISSN: 0910-6340            Impact factor:   2.081