| Literature DB >> 33541032 |
Ga-Young Kim1, Jong Sei Kim2, Mini Jo1, Hye Yoon Seol1, Young Sang Cho2, Il Joon Moon2.
Abstract
OBJECTIVES: This study was conducted to investigate the electroacoustic characteristics of personal sound amplification products (PSAPs), to identify whether PSAPs provide adequate gain and output for three common hearing loss (HL) configurations, and to compare the benefits of a representative PSAP (RPSAP) and a conventional hearing aid (HA) for clinical hearing outcomes as a pilot study.Entities:
Keywords: Hearing Aids; Hearing Loss; Sensorineural Hearing Loss; Wearable Electronic Devices
Year: 2021 PMID: 33541032 PMCID: PMC8901943 DOI: 10.21053/ceo.2020.02313
Source DB: PubMed Journal: Clin Exp Otorhinolaryngol ISSN: 1976-8710 Impact factor: 3.372
Characteristics of the PSAPs included in this study
| Device | Model No. | Manufacturer | Retail cost | Price group | Bluetooth | Smartphone application | Channel | Feedback cancellation | Noise reduction | Self-fitting | Preset mode | Other feature |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| BeethoSOL | EM-C110 | EM-Tech | $210/pair | Basic | ○ | ○ | 5 | ○ | ○ | ○ | ○ | |
| EarJJang | VA-3000 | ESONIC | $167/pair | Basic | × | × | NS | NS | NS | × | ○ | |
| Geniesori2 | HA1000 | UWICOM | $198/pair | Basic | ○ | × | NS | NS | NS | × | ○ | Smart voice sender mode, dual microphone |
| Hearing Able | IRIS20 | Microtech | ||||||||||
| System | $244/single | High-end | ○ | ○ | 20 | NS | ○ | ○ | ○ | |||
| Olive Smart Ear | DHFA2FBK | Olive | ||||||||||
| Union | $253/single | High-end | ○ | ○ | 16 | ○ | ○ | ○ | ○ | Equalizer | ||
| SoriIn | YDH-B1000 | The Yeolrim | $139/single | High-end | × | × | NS | ○ | ○ | × | × |
Price groups were determined based on the retail cost for a pair.
PSAP, personal sound amplification product; NS, not stated.
Fig. 1.(A-F) Individuals audiograms of the participants.
Demographic and audiological characteristics of the individual patients
| Patient No. | Sex | Age (yr) | PTA[ | PTA[ | WRS (right) | WRS (left) | Tinnitus | HA user |
|---|---|---|---|---|---|---|---|---|
| 1 | F | 60 | 53.75 | 55.00 | 80 | 76 | Both | No |
| 2 | F | 56 | 41.25 | 51.25 | 100 | 98 | Both | No |
| 3 | F | 51 | 46.25 | 42.50 | 88 | 96 | Both | No |
| 4 | M | 68 | 48.75 | 46.25 | 92 | 88 | None | No |
| 5 | F | 53 | 50.00 | 51.25 | 96 | 96 | Left | Yes |
| 6 | F | 61 | 41.25 | 45.00 | 84 | 96 | Both | No |
PTA, pure tone average; WRS, word recognition score; HA, hearing aid.
Calculated from air conduction hearing level (dB) at 0.5, 1, 2, and 4 KZ.
Electroacoustic analysis of the PSAPs included in this study
| Device | Electroacoustic analysis result | |||||
|---|---|---|---|---|---|---|
| OSPL90 Max (dB SPL) | Frequency range (Hz) | EIN (dB SPL) | THD (%) | |||
| 500 Hz | 800 Hz | 1,600 Hz | ||||
| BeethoSOL (basic) | 115.9 | 293–8,030[ | 32.4[ | 2.4 | 4.5[ | 2.5 |
| EarJJang (basic) | 119.6 | 100–7,178 | 39.0[ | 0.4 | 0.2 | 2.5 |
| Geniesori2 (basic) | 123.7[ | 100–7,735 | 35.4[ | 0.2 | 0.1 | 0.1 |
| Hearing-Able (high-end) | 103.0 | 100–3,717[ | 26.7 | 0.9 | 1.0 | 0.1 |
| Olive Smart Ear (high-end) | 97.2 | 215–6,650 | 27.4 | 1.8 | 0.6 | 0.1 |
| SoriIn (high-end) | 109.6 | 100–7,024 | 26.9 | 0.7 | 0.5 | 0.1 |
PSAP, personal sound amplification product; OSPL90 Max, maximum output sound pressure level at 90 dB SPL; EIN, equivalent input noise; THD, total harmonic distortion.
Indicates measured values that deviated from the tolerance defined in the Methods section.
Simulated real-ear measurement results of the PSAPs included in this study
| Input level (65 dB SPL) | Mild-to-moderate high-frequency hearing loss (frequency, kHz) | Appropriate fit | Moderate to moderately severe sloping hearing loss (frequency, kHz) | Appropriate fit | Moderate flat hearing loss (frequency, kHz) | Appropriate fit | |||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 0.25 | 0.5 | 1 | 2 | 3 | 4 | 6 | 0.25 | 0.5 | 1 | 2 | 3 | 4 | 6 | 0.25 | 0.5 | 1 | 2 | 3 | 4 | 6 | |||||
| BeethoSOL (basic) | |||||||||||||||||||||||||
| Targeted (dB SPL) | 57 | 57 | 53 | 60 | 62 | 57 | 51 | 6/7 | 66 | 63 | 66 | 75 | 77 | 71 | 66 | 5/7 | 66 | 64 | 62 | 65 | 64 | 59 | 54 | 6/7 | |
| Responded (dB SPL) | 59 | 65 | 54 | 53 | 66 | 65 | 34 | 62 | 68 | 61 | 60 | 74 | 73 | 41 | 58 | 64 | 64 | 59 | 64 | 63 | 34 | ||||
| Fit to prescription | 2 | 8 | 1 | –7 | 4 | 8 | –17 | –4 | 5 | –5 | –15[ | –3 | 2 | –25[ | –8 | 0 | 2 | –6 | 0 | 4 | –20[ | ||||
| EarJJang (basic) | |||||||||||||||||||||||||
| Targeted (dB SPL) | 57 | 57 | 53 | 60 | 62 | 57 | 51 | 4/7 | 66 | 63 | 66 | 75 | 77 | 71 | 66 | 2/7 | 66 | 64 | 62 | 65 | 64 | 59 | 54 | 4/7 | |
| Responded (dB SPL) | 85 | 93 | 74 | 65 | 56 | 53 | 54 | 85 | 93 | 74 | 65 | 56 | 53 | 54 | 84 | 93 | 74 | 65 | 56 | 53 | 54 | ||||
| Fit to prescription | 28[ | 36[ | 21[ | 5 | –6 | –4 | 3 | 19[ | 30[ | 8 | –10 | –21[ | –18[ | –12[ | 18[ | 29[ | 12[ | 0 | –8 | –6 | 0 | ||||
| Geniesori2 (basic) | |||||||||||||||||||||||||
| Targeted (dB SPL) | 57 | 57 | 53 | 60 | 62 | 57 | 51 | 4/7 | 66 | 63 | 66 | 75 | 77 | 71 | 66 | 2/7 | 66 | 64 | 62 | 65 | 64 | 59 | 54 | 5/7 | |
| Responded (dB SPL) | 84 | 83 | 68 | 64 | 56 | 50 | 49 | 90 | 89 | 74 | 69 | 62 | 55 | 54 | 87 | 86 | 71 | 66 | 59 | 52 | 51 | ||||
| Fit to prescription | 27[ | 26[ | 15[ | 4 | –6 | –7 | –2 | 24[ | 26[ | 8 | –6 | –15[ | –16[ | –12[ | 21[ | 22[ | 9 | 1 | –5 | –7 | –3 | ||||
| Hearing Able (high-end) | |||||||||||||||||||||||||
| Targeted (dB SPL) | 57 | 57 | 54 | 61 | 64 | 61 | 53 | 6/7 | 66 | 64 | 67 | 76 | 79 | 76 | 67 | 6/7 | 66 | 64 | 63 | 66 | 67 | 63 | 55 | 6/7 | |
| Responded (dB SPL) | 55 | 58 | 52 | 73 | 66 | 59 | 52 | 60 | 64 | 69 | 79 | 71 | 64 | 57 | 54 | 62 | 61 | 71 | 64 | 55 | 48 | ||||
| Fit to prescription | –2 | 1 | –2 | 12[ | 2 | –2 | –1 | –6 | 0 | 2 | 3 | –8 | –12[ | –10 | –12[ | –2 | –2 | 5 | –3 | –8 | –7 | ||||
| Olive Smart Ear (high-end) | |||||||||||||||||||||||||
| Targeted (dB SPL) | 57 | 57 | 54 | 62 | 65 | 61 | 54 | 7/7 | 66 | 63 | 66 | 75 | 77 | 73 | 66 | 5/7 | 66 | 64 | 62 | 66 | 65 | 60 | 54 | 7/7 | |
| Responded (dB SPL) | 51 | 60 | 57 | 62 | 63 | 56 | 52 | 56 | 68 | 68 | 70 | 67 | 56 | 53 | 57 | 67 | 65 | 67 | 64 | 56 | 53 | ||||
| Fit to prescription | –6 | 3 | 3 | 0 | –2 | –5 | –2 | –10 | 5 | 2 | –5 | –10 | –17[ | –13 | –9 | 3 | 3 | 1 | –1 | –4 | –1 | ||||
| SoriIn (high-end) | |||||||||||||||||||||||||
| Targeted (dB SPL) | 57 | 57 | 53 | 62 | 69 | 69 | 59 | 4/7 | 65 | 63 | 66 | 76 | 84 | 84 | 74 | 4/7 | 66 | 64 | 62 | 66 | 71 | 71 | 62 | 5/7 | |
| Responded (dB SPL) | 55 | 59 | 62 | 58 | 57 | 51 | 39 | 67 | 71 | 74 | 70 | 68 | 64 | 50 | 63 | 66 | 70 | 66 | 64 | 59 | 46 | ||||
| Fit to prescription | –2 | 2 | 9 | –4 | –12[ | –18[ | –20[ | 2 | 8 | 8 | –6 | –16[ | –20[ | –24[ | –3 | 2 | 8 | 0 | –7 | –12[ | –16[ | ||||
PSAP, personal sound amplification product.
Indicates that the difference between the target and response levels was more than ±10 dB SPL.
Fig. 2.Functional gain of a hearing aid (HA) and a representative personal sound amplification product (RPSAP) for six participants by frequency: (A) 250 Hz, (B) 500 Hz, (C) 1,000 Hz, (D) 2,000 Hz, (E) 3,000 Hz, (F) 4,000 Hz, (G) 6,000 Hz, (H) 8,000 Hz. Functional gain was defined as the difference in dB HL between the aided and unaided thresholds. The overlapping dots and gray lines indicate individual changes. The rhombus shape indicates the mean values of the results.
Fig. 3.Average results of the word recognition score (WRS; A) and Korean version of the Hearing in Noise Test (K-HINT; B) for the hearing aid (HA) and representative personal sound amplification product (PSAP). The overlapping dots and gray lines indicate individual changes. The rhombus shape indicates the mean values of the results.
Fig. 4.Results of the Korean version of the Hearing in Noise Test (K-HINT) for each patient. (A) Individual results of the K-HINT in a quiet setting. (B) Average results for each condition of the K-HINT in a quiet setting. (C) Individual results of the K-HINT in a noisy setting. (D) Average results for each condition of the K-HINT in a noisy setting. The rhombus shape indicates the mean values of the results. HA, hearing aid; RPSAP, representative personal sound amplification product; S/N, signal-to-noise.