| Literature DB >> 33536401 |
Tomoyuki Fujioka1, Mio Mori1, Jun Oyama1, Kazunori Kubota1,2, Emi Yamaga1, Yuka Yashima1, Leona Katsuta1, Kyoko Nomura1, Miyako Nara1,3, Goshi Oda4, Tsuyoshi Nakagawa4, Ukihide Tateishi1.
Abstract
PURPOSE: Synthetic MRI reconstructs multiple sequences in a single acquisition. In the present study, we aimed to compare the image quality and utility of synthetic MRI with that of conventional MRI in the breast.Entities:
Keywords: breast; breast cancer; magnetic resonance imaging; synthetic magnetic resonance imaging
Mesh:
Year: 2021 PMID: 33536401 PMCID: PMC8922358 DOI: 10.2463/mrms.mp.2020-0132
Source DB: PubMed Journal: Magn Reson Med Sci ISSN: 1347-3182 Impact factor: 2.471
Fig. 1Multidynamic multiecho sequence for synthetic MRI. The figure shows the Gm, Gp, and Gs and the RF pulse amplitude over time. There are two phases in each block. In first phase (saturation), the 120° saturation pulse and subsequent spoiling acts. In second phase (acquisition), the multi-echo spin-echo acquisition is performed, using the 90° excitation pulse and multiple 180° refocusing pulses. Gm, gradient in measurement; Gp, gradient in phase-encoding; Gs, gradient in slice-selection. This image is reprinted with permission from GE Healthcare Japan.
Fig. 2Real image and imaginary image. Using different combinations of echo time and saturation delay, each acquisition produced (a) real images and (b) imaginary images per section for the quantification of the physical properties of tissue.
Fig. 3T1, T2, and PD maps. Quantitative images of (a) T1, (b) T2, and (c) PD maps were generated by quantification of the physical properties of tissue. PD, proton density.
MRI acquisition parameters.
| Sequence | Sy T1-WI | Sy T2-WI | Sy FS-T2-WI | T1-WI | T2-WI | FS-T2-WIb |
|---|---|---|---|---|---|---|
| TR (ms) | 500 | 5000 | 15000 | 5.4 | 3000 | 3000 |
| TE (ms) | 10 | 85 | 100 | 2.6 | 85 | 85 |
| TI (ms) | — | — | 260 | — | — | — |
| Flip angle (°) | — | — | — | 10 | — | — |
| Field of view (mm) | 360 | 360 | 360 | 360 | 360 | 360 |
| Section thickness (mm) | 4 | 4 | 4 | 2(4)a | 4 | 4 |
| Intersection gap (mm) | 4 | 4 | 4 | 2(4)a | 4 | 4 |
| Bandwidth (Hz/pixel) | 462.9 | 462.9 | 462.9 | 559 | 125 | 125 |
| Echo-train length | 16 | 16 | 16 | — | 14 | 14 |
| Acceleration factor | 2 | 2 | 2 | 2.25 | 2 | 2 |
| Acquisition time (min:s) | 3:28 | 3:28 | 3:28 | 0:43 | 2:21 | 2:21 |
T1-WI were acquired with a section thickness of 2 mm and an intersection gap of 2 mm and reconstructed into a section thickness of 4 mm and an intersection gap of 4 mm. FS was conducted using the two-point DIXON sequence. FS, fat-suppressed; Sy, synthetic; WI, weighted images.
Comparison of scores for synthetic MRI and conventional MRI
| Reader 1 | Reader 2 | ||||||
|---|---|---|---|---|---|---|---|
| T1-WI | Sy T1-WI | T1-WI | Sy T1-WI | ||||
| Overall image quality | 4.97 ± 0.16 | 4.95 ± 0.23 | 0.773 | 4.86 ± 0.35 | 4.95 ± 0.23 | 0.149 | |
| Anatomical sharpness | 5.00 ± 0.00 | 4.95 ± 0.23 | 0.346 | 4.92 ± 0.28 | 4.89 ± 0.31 | 0.777 | |
| Contrast between tissues | 4.97 ± 0.16 | 5.00 ± 0.00 | 1.000 | 4.92 ± 0.28 | 4.84 ± 0.37 | 0.233 | |
| Image homogeneity | 4.86 ± 0.42 | 4.86 ± 0.35 | 0.351 | 4.78 ± 0.42 | 4.73 ± 0.45 | 0.530 | |
| Presence of artifacts | 4.57 ± 0.50 | 4.49 ± 0.51 | 0.299 | 4.62 ± 0.49 | 4.73 ± 0.45 | 0.267 | |
| T2-WI | Sy T2-WI | T2-WI | Sy T2-WI | ||||
| Overall image quality | 5.00 ± 0.00 | 4.97 ± 0.16 | 1.000 | 4.95 ± 0.23 | 4.97 ± 0.16 | 0.773 | |
| Anatomical sharpness | 5.00 ± 0.00 | 4.97 ± 0.16 | 1.000 | 4.81 ± 0.40 | 4.95 ± 0.23 | 0.073 | |
| Contrast between tissues | 5.00 ± 0.00 | 5.00 ± 0.00 | 1.000 | 4.62 ± 0.49 | 4.97 ± 0.16 | ||
| Image homogeneity | 5.00 ± 0.00 | 4.81 ± 0.40 | 4.95 ± 0.23 | 4.81 ± 0.40 | 0.110 | ||
| Presence of artifacts | 4.95 ± 0.23 | 4.78 ± 0.42 | 4.95 ± 0.23 | 4.86 ± 0.35 | 0.299 | ||
| FS-T2-WI | Sy FS-T2-WI | FS T2-WI | Sy FS-T2-WI | ||||
| Overall image quality | 4.97 ± 0.16 | 3.81 ± 0.40 | 4.92 0.28 | 4.22 ± 0.63 | |||
| Anatomical sharpness | 4.97 ± 0.16 | 4.16 ± 0.76 | 4.81 0.40 | 4.41 ± 0.50 | |||
| Contrast between tissues | 5.00 ± 0.00 | 4.89 ± 0.39 | 0.174 | 4.89 0.31 | 4.76 ± 0.86 | 0.608 | |
| Image homogeneity | 5.00 ± 0.00 | 3.51 ± 0.51 | 4.81 0.40 | 4.00 ± 0.53 | |||
| Presence of artifacts | 4.68 ± 0.47 | 3.97 ± 0.29 | 4.81 0.40 | 4.35 ± 0.48 | |||
Values represent means ± standard deviation. Wilcoxon’s signed-rank test was conducted. FS, fat-suppressed; Sy, synthetic; WI, weighted images.
Weighted kappa for evaluating image quality between the two readers
| Conventional MRI | Synthetic MRI | |||
|---|---|---|---|---|
| Kappa coefficient (95% CI) | Kappa coefficient (95% CI) | |||
| T1-WI | 0.56 (0.39–0.73) | < 0.001 | 0.52 (0.35–0.70) | < 0.001 |
| T2-WI | 0.52 (0.34–0.69) | < 0.001 | 0.57 (0.39–0.74) | < 0.001 |
| FS-T2-WI | 0.52 (0.35–0.69) | < 0.001 | 0.37 (0.21–0.53) | < 0.001 |
| Average | 0.53 | 0.46 | ||
CI, confidence interval; FS, fat-suppressed; WI, weighted images.
Fig. 4Case 1 of Sy MRI and conventional MRI in the breast. A 69 year-old woman with noninfiltrating intracystic carcinoma (18 mm) in the right breast. Sy MRI (a c) and conventional MRI (d f) clearly show solid component as low signal intensity of the T1-WI and iso-signal intensity of T2- and FS-T2-WI and cystic components as low signal intensity of T1-WI and high signal intensity of T2- and FS-T2-WI. Contrast-enhanced T1-WI (g) and ultrasound image (h) show a mass composed of solid and cystic components. Sy T1-WI (a), Sy T2-WI (b), Sy FS-T2-WI (c), T1-WI (d), T2-WI (e), FS-T2-WI (f), contrast-enhanced T1-WI (g), and ultrasound image (h). FS, fat-suppressed; Sy, synthetic; T1-WI, T1-weighted image; T2-WI, T2-weighted image.