Young Ahn Yoon1, Yong-Wha Lee2, Sollip Kim3, Kyunghoon Lee4, Hyung-Doo Park5, Sail Chun6, Won-Ki Min6.
Abstract
BACKGROUND: Total cholesterol concentration measurement is important in the diagnosis of dyslipidemia and evaluation of cardiovascular disease risk factors. Measurement reliability for obtaining an accurate total cholesterol concentration requires procedure standardization. We evaluated the standardization status for total cholesterol concentration measurement through Korean external quality assessment (EQA) data analysis.Entities:
Keywords: External quality assessment; Inaccuracy; Standardization; Total cholesterol
Mesh:
Substances:
Year: 2021 PMID: 33536354 PMCID: PMC7884189 DOI: 10.3343/alm.2021.41.4.366
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Statistics of laboratories participating in total cholesterol concentration measurement
| Manufacturer | Laboratories (N) | Sample A[ | Sample B[ | Sample C[ |
|---|---|---|---|---|
| Mean ± SD (mmol/L) | ||||
| Total[ | 1,670 | 6.58 ± 0.21 | 2.68 ± 0.13 | 6.60 ± 0.20 |
| Sekisui Medical, Tokyo, Japan | 285 | 6.61 ± 0.16 | 2.74 ± 0.07 | 6.62 ± 0.13 |
| Roche Diagnostics, Mannheim, Germany | 273 | 6.44 ± 0.12 | 2.58 ± 0.06 | 6.46 ± 0.14 |
| Beckman Coulter Inc., Brea, CA, USA | 224 | 6.66 ± 0.16 | 2.57 ± 0.09 | 6.68 ± 0.17 |
| Wako Diagnostics, Osaka, Japan | 98 | 6.60 ± 0.18 | 2.76 ± 0.06 | 6.58 ± 0.16 |
| Shenzhen Mindray Bio-Medical Electronics, Shenzhen, China | 91 | 6.65 ± 0.12 | 2.72 ± 0.06 | 6.65 ± 0.15 |
| Stanbio Laboratory, Boerne, TX, USA | 82 | 6.36 ± 0.18 | 2.62 ± 0.14 | 6.45 ± 0.24 |
| JW Medical, Seoul, Korea | 76 | 6.65 ± 0.26 | 2.81 ± 0.11 | 6.66 ± 0.20 |
| Biosystems, Barcelona, Spain | 64 | 6.79 ± 0.23 | 2.69 ± 0.09 | 6.90 ± 0.18 |
| Kyowa Medex, Tokyo, Japan | 58 | 6.57 ± 0.09 | 2.75 ± 0.05 | 6.57 ± 0.10 |
| Siemens Healthineers, Erlangen, Germany | 49 | 6.39 ± 0.14 | 2.73 ± 0.06 | 6.38 ± 0.13 |
| GDLKorea, Anyang, Korea | 47 | 6.62 ± 0.19 | 2.74 ± 0.11 | 6.57 ± 0.13 |
| ELITechGroup, Puteaux, France | 33 | 6.54 ± 0.30 | 2.72 ± 0.14 | 6.67 ± 0.35 |
| DiaSys Diagnostic Systems, Holzheim, Germany | 31 | 6.78 ± 0.28 | 2.68 ± 0.15 | 6.80 ± 0.25 |
| Hanlab, Chungju, Korea | 27 | 6.61 ± 0.10 | 2.75 ± 0.05 | 6.61 ± 0.09 |
| IVDLab, Uiwang, Korea | 27 | 6.66 ± 0.22 | 2.79 ± 0.13 | 6.75 ± 0.26 |
| LC Diagnostics, Quezon, The Philippines | 22 | 6.78 ± 0.16 | 2.78 ± 0.11 | 6.82 ± 0.24 |
| Asan Pharm, Seoul, Korea | 21 | 6.67 ± 0.21 | 2.82 ± 0.09 | 6.75 ± 0.27 |
| Spinreact, Girona, Spain | 15 | 6.76 ± 0.21 | 2.75 ± 0.06 | 6.86 ± 0.16 |
| Shinyang Diagnostics, Seoul, Korea | 14 | 6.45 ± 0.22 | 2.55 ± 0.03 | 6.48 ± 0.21 |
| Erba diagnostics, Miami Lakes, FL, USA | 14 | 6.49 ± 0.21 | 2.70 ± 0.08 | 6.51 ± 0.16 |
| Shino-Test Corporation, Tokyo, Japan | 13 | 6.33 ± 0.13 | 2.65 ± 0.06 | 6.34 ± 0.12 |
| Embiel, Gunpo, Korea | 12 | 6.49 ± 0.26 | 2.61 ± 0.12 | 6.58 ± 0.24 |
| GeneMatrix, Yongin, Korea | 12 | 6.74 ± 0.29 | 2.88 ± 0.10 | 6.78 ± 0.19 |
| Shinsung Pharm, Suwon, Korea | 10 | 6.71 ± 0.36 | 2.94 ± 0.19 | 6.79 ± 0.21 |
*The target value and performance criteria of sample A, B, and C: 6.59±0.20, 2.80±0.08, and 6.63±0.20 mmol/L. †Manufacturers with <10 laboratories are not listed in this table but were counted in the total number.
Abbreviation: SD, standard deviation.
Fig. 1Distribution of total cholesterol concentrations according to manufacturers, and target values and performance criteria of quality control materials (samples A, B, and C). The peer group mean is the average value of all laboratories in the same manufacturer group. Numbers from 1 to 24 were given to each manufacturer group in ascending order of peer group mean. (A) Data obtained using sample A. (B) Data obtained using sample B. (C) Data obtained using sample C.
Abbreviation: SD, standard deviation.