Literature DB >> 33533089

Intracardiac echocardiography-guided implantation of the Watchman FLX left atrial appendage closure device.

Mohit K Turagam1, Petr Neuzil2, Jan Petru2, Pavel Hala2, Tomas Mraz2, Jiri Baroch2, Veronika Lekesova2, Milena Prokopova2, Srinivas R Dukkipati1, Vivek Y Reddy1,2.   

Abstract

BACKGROUND: The next-generation Watchman FLX left atrial appendage closure (LAAC) device has: (1) an atraumatic closed distal end, (2) reduced height, (3) a recessed screw hub to decrease device-related thrombus (DRT), (4) two rows of J-shape anchors so redeployment is possible after full recapture, and (5) ability to treat a greater size range of LAA ostia.
OBJECTIVE: To report, for the first time, the feasibility and safety of intracardiac echocardiography (ICE)-guided Watchman FLX implantation.
METHODS: A single-center prospective registry of atrial fibrillation patients planned for LAAC with the FLX device underwent ICE-guided implantation with conscious sedation. Transesophageal echocardiography (TEE) imaging was done preprocedure (to assess LAA size and exclude thrombus) and at clinical follow-up at 6-12 weeks. Clinical outcomes were LAA closure success, complications, leak, or DRT on follow-up TEE and major safety events.
RESULTS: The study included 30 patients: age 75 ± 8 years, 53% men, CHA2 DS2 -VASc 4.6 ± 1.6, and HAS-BLED 3.4 ± 1.1. The primary indication was prior bleeding in 60% (72% GI bleeding). The LAA orifice width and length were 22.7 ± 3.1 and 25.7 ± 5.7 mm, respectively. Technical success was 100% (the first-choice device was used in 28 of 30). Procedure time was less than 30 min in 27 of 30 cases, with 36 ± 15 ml contrast used. The final device size was 29.2 ± 4.7 mm with 21.6 ± 4.5% compression. There were no procedure-related complications. Follow-up TEE at a median 47 days follow-up showed 100% device success with no DRT or peridevice leak ≥5 mm. Major safety events occurred in 6.6% (2/30).
CONCLUSION: The Watchman FLX device can be safely implanted with intraprocedural ICE imaging instead of TEE.
© 2021 Wiley Periodicals LLC.

Entities:  

Keywords:  Watchman FLX; intracardiac echocardiography; left atrial appendage occlusion; stroke prevention

Year:  2021        PMID: 33533089     DOI: 10.1111/jce.14927

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  2 in total

1.  Experience of Combined Procedure during Percutaneous LAA Closure.

Authors:  Guillaume Domain; Nicolas Dognin; Gilles O'Hara; Josep Rodès-Cabau; Jean-Michel Paradis; Camille Strubé; Mathieu Bernier; Kim O'Connor; Jonathan Beaudoin; François Philippon; Erwan Salaun; Jean Champagne
Journal:  J Clin Med       Date:  2022-06-08       Impact factor: 4.964

2.  Snare and track (SNACK) technique: a novel approach for successful placement of an intracardiac echocardiogram catheter in the left atrium: a case report.

Authors:  Akram Kawsara; Bryan Raybuck
Journal:  Eur Heart J Case Rep       Date:  2022-09-05
  2 in total