Fatih Seker1, Johannes Alex Rolf Pfaff1, Yahia Mokli2, Anne Berberich2, Rafael Namias3, Steven Gerry4, Simon Nagel2, Martin Bendszus1, Christian Herweh1.
Abstract
BACKGROUND AND AIM: The aim of this study was to assess the diagnostic accuracy of e-CTA (product name) (Brainomix) in the automatic detection of large vessel occlusions in anterior circulation stroke.Entities:
Keywords: Artificial intelligence; CT angiography; CT scan; ischemic stroke; large vessel occlusion; radiology
Mesh:
Year: 2021 PMID: 33527886 PMCID: PMC8739618 DOI: 10.1177/1747493021992592
Source DB: PubMed Journal: Int J Stroke ISSN: 1747-4930 Impact factor: 5.266
Figure 1.e-CTA demonstrating a distal occlusion of the right middle cerebral artery.
Detection of any LVO (proximal or distal) on correct side
| Sensitivity | Specificity | PPV | NPV | Accuracy | Kappa | |
|---|---|---|---|---|---|---|
| e-CTA ( | 0.84 (0.77–0.89) | 0.96 (0.91–0.98) | 0.96 (0.91–0.98) | 0.84 (0.77–0.89) | 0.89 (0.85–0.93) | 0.83 (0.77–0.89) |
| e-CTA ( | 0.84 (0.74–0.92) | 0.96 (0.89–0.99) | 0.95 (0.87–0.99) | 0.87 (0.77–0.93) | 0.90 (0.84–0.95) | 0.84 (0.76–0.92) |
| Neuroradiologist ( | 0.97 (0.90–1.00) | 0.99 (0.93–1.00) | 0.99 (0.92–1.00) | 0.97 (0.91–1.00) | 0.98 (0.94–1.00) | 0.97 (0.93–1.00) |
| Radiology resident ( | 0.96 (0.88–0.99) | 0.92 (0.83–0.97) | 0.92 (0.83–0.97) | 0.92 (0.83–0.97) | 0.94 (0.88–0.97) | 0.90 (0.84–0.96) |
| Neurology resident 1 ( | 0.86 (0.75–0.93) | 0.91 (0.81–0.96) | 0.90 (0.80–0.96) | 0.87 (0.77–0.94) | 0.88 (0.82–0.93) | 0.81 (0.72–0.89) |
| Neurology resident 2 ( | 0.91 (0.82–0.97) | 1.00 (0.95–1.00) | 1.00 (0.94–1.00) | 0.93 (0.84–0.97) | 0.96 (0.91–0.98) | 0.93 (0.88–0.98) |
NPV: negative predictive value; PPV: positive predictive value.
Confidence intervals are provided in brackets. The whole cohort (n = 301) was analyzed by e-CTA only. A subgroup (n = 144) was additionally analyzed by four human readers.
Detection of proximal LVO on correct side
| Sensitivity | Specificity | PPV | NPV | Accuracy | Kappa | |
|---|---|---|---|---|---|---|
| e-CTA ( | 0.92 (0.84–0.96) | 0.98 (0.94–1.00) | 0.97 (0.91–0.99) | 0.95 (0.89–0.98) | 0.95 (0.92–0.98) | 0.92 (0.87–0.96) |
| e-CTA ( | 0.93 (0.80–0.98) | 1.00 (0.95–1.00) | 1.00 (0.91–1.00) | 0.96 (0.89–0.99) | 0.97 (0.93–0.99) | 0.94 (0.88–1.00) |
| Neuroradiologist ( | 0.97 (0.87–1.00) | 1.00 (0.95–1.00) | 1.00 (0.91–1.00) | 0.99 (0.93–1.00) | 0.99 (0.95–1.00) | 0.98 (0.94–1.00) |
| Radiology resident ( | 0.95 (0.83–0.99) | 1.00 (0.95–1.00) | 1.00 (0.91–1.00) | 0.97 (0.91–1.00) | 0.98 (0.94–1.00) | 0.96 (0.91–1.00) |
| Neurology resident 1 ( | 0.88 (0.73–0.96) | 0.97 (0.91–1.00) | 0.95 (0.82–0.99) | 0.94 (0.85, 0.98) | 0.94 (0.88–0.98) | 0.86 (0.76–0.96) |
| Neurology resident 2 ( | 0.80 (0.64–0.91) | 1.00 (0.95–1.00) | 1.00 (0.89–1.00) | 0.90 (0.82–0.96) | 0.93 (0.87–0.97) | 0.84 (0.73–0.95) |
NPV: negative predictive value; PPV: positive predictive value.
Confidence intervals are provided in brackets. The whole cohort after exclusion of distal occlusions (n = 236) was analyzed by e-CTA only. A subgroup (n = 114) was additionally analyzed by four human readers.