A DGA-arm-grafted macrocyclic aza-crown ether ligand (Cr6DGA) was synthesized, and its solvent extraction behavior toward trivalent americium and europium in nitric acid medium was studied. The effects of various parameters such as the contact time, temperature, concentration of the extractant, and acidity on the extraction by Cr6DGA were investigated. It was found that in 3 mol/L HNO3, the SFEu/Am value was about 2. The complexation energies calculated by DFT showed that the Eu(III) complexes were more stable than the corresponding Am(III) complexes in gas, aqueous, and organic phases. Furthermore, the coordination study showed that the metal/ligand ratio of the extracted species was 1:2 by mass spectrometry (MS) analysis. The time-resolved laser-induced fluorescence spectra (TRLFS) further proved that the extracted species contained one water molecule, and so the composition of the extracted complexes may be [EuL2NO3(H2O)]2+ or [EuL2(NO3)2(H2O)]+. Finally, DFT calculations revealed that [EuL2(NO3)2(H2O)]+ is a more stable species and the binding energy of Eu(III) with the DGA unit is lower than that with the crown unit.
A DGA-arm-grafted macrocyclic aza-crown ether ligand (Cr6DGA) was synthesized, and its solvent extraction behavior toward trivalent americium and europium in nitric acid medium was studied. The effects of various parameters such as the contact time, temperature, concentration of the extractant, and acidity on the extraction by Cr6DGA were investigated. It was found that in 3 mol/L HNO3, the SFEu/Am value was about 2. The complexation energies calculated by DFT showed that the Eu(III) complexes were more stable than the corresponding Am(III) complexes in gas, aqueous, and organic phases. Furthermore, the coordination study showed that the metal/ligand ratio of the extracted species was 1:2 by mass spectrometry (MS) analysis. The time-resolved laser-induced fluorescence spectra (TRLFS) further proved that the extracted species contained one water molecule, and so the composition of the extracted complexes may be [EuL2NO3(H2O)]2+ or [EuL2(NO3)2(H2O)]+. Finally, DFT calculations revealed that [EuL2(NO3)2(H2O)]+ is a more stable species and the binding energy of Eu(III) with the DGA unit is lower than that with the crown unit.
Since the first use
of nuclear energy, its development has spanned
more than half a century. At present, nuclear energy has become synonymous
with “economic, green, and efficient” energy. With the
continuous increase of population and the rapid development of the
economy, the energy demand is increasing rapidly and therefore more
attention needs to be paid to the safe use of nuclear energy.For the safe development and effective utilization of nuclear energy,
the disposal of high-level radioactive waste is crucial. In particular,
the long-lived fission products and actinides urgently need to be
remediated because of their strong radioactivity, strong toxicity,
and high heat release rate. Among them, the separation of trivalent
actinides and lanthanides has become one of the most challenging topics
due to their very similar physicochemical properties.[1,2]In recent years, a series of important ligands have been applied
to the partitioning of lanthanides/actinides. These ligands include
organic phosphorus compounds (for the TALSPEAK process),[3] diglycolamide (DGA),[4] malonamide (for the DIAMEX process),[5] and N-heterocyclic ligands (for the SANEX process).[6] Among them, it was found that the DGA extractant rapidly
developed due to its advantages of stable chemical properties, radiation
resistance, flammability, and absence of secondary pollution. Several
DGA extractants were tested on a large scale and “thermal engineering
operation” was performed.[7] In particular, N,N,N,N′-tetraoctyl diglycolamide (TODGA) was found to be the optimal
ligand in terms of solubility and extraction capacity,[8] and it was certified that the stoichiometry of the extraction
complexes depended on the nature of diluents.[9] In polar diluents, such as 1,2-dichloroethane, the dominant chemical
species was the metal/TODGA in 1:2 ratio, while in nonpolar diluents,
the 1:3 or 1:4 species was the predominant complex.[10] Obviously, the complexation of several TODGA molecules
with a trivalent metal ion is not favorable to the entropy change
during extraction.[11] In nonpolar diluents,
such as n-dodecane, three to four TODGA molecules
are involved in the formation of reverse micelles, the core of which
shows an affinity for trivalent lanthanide and actinides in a dimensionally
selective manner.[12] It can be predicted
that the preattachment of three to four DGA molecules to a molecular
platform will improve the affinity of the ligand to trivalent lanthanide
and actinide ions, and this ligand may demonstrate a higher ability
to extract trivalent lanthanide and actinide ions than TODGA. Therefore,
some researchers have developed a series of multiple-DGA-functionalized
ligands by attaching multiple DGA arms to a central platform to make
the extraction process more thermodynamically favorable and independent
of diluents, such as 2,2′-((9-methyl-9-(11-octyl-6,10-dioxo-2,8-dioxa-5,11-diazanonadecyl)-2,16-dioxo-7,11-dioxa-3,15-diazaheptadecane-1,17-diyl)bis(oxy))bis(N,N-dioctylacetamide) (T-DGA), 2,2′,2″-((((nitrilotris(ethane-2,1-diyl))tris(azanediyl))tris(2-oxoethane-2,1-diyl))tris(oxy))tris(N,N-dioctylacetamide) (TREN-DGA), and 2,2′,2″-(((((2,4,6-triethylbenzene-1,3,5-triyl)tris(ethane-2,1-diyl))tris(azanediyl))tris(2-oxoethane-2,1-diyl))tris(oxy))tris(N,N-dioctylacetamide) (Bz-T-DGA), and the
M/L ratio of the extracted species by these ligands is 1:2, and only
one DGA unit in a multiple-DGA extractant molecule participates in
the coordination and extraction of metal ions.[13−15] Recently, 2,2′,2″-(((1,4,7-triazonane-1,4,7-triyl)tris(2-oxoethane-2,1-diyl))tris(oxy))tris(N,N-dioctylacetamide) (T9C3ODGA, Figure ) was designed by
connecting three DGA arms to the macrocyclic aza-crown ether to reduce
the steric hindrance of the ligand and make full use of the advantages
of multiple-DGA-functionalized structures, and the distribution ratios
of Eu(III)/Am(III) by T9C3ODGA were significantly higher than those
of T-DGA and TREN-DGA.[16] Although the protonation
of macrocyclic aza-crown ether occurs under strong acid conditions,
the extraction experiments showed that the distribution ratio and
the selectivity for Eu(III)/Am(III) are not significantly affected.[17] Subsequently, 2,2′,2″,2‴-(((1,4,7,10-tetraazacyclododecane-1,4,7,10-tetrayl)-tetrakis(2-oxoethane-2,1-diyl))tetrakis(oxy))tetrakis(N,N-dioctylacetamide) (T12C4ODGA, Figure ) has been reported.
Compared with T9C3ODGA, the ability of this ligand to extract trivalent
lanthanide/actinidemetal ions was significantly improved (n-dodecane, DAm > 500, DEu > 103).[18] According to the experimental and theoretical results,
the ligands
T9C3ODGA and T12C4ODGA formed the same type of extraction complex
(M/L = 1:1 and 1:2), and the 1:2 species was the optimal structure
during the extraction process. It was found that the coordination
modes of the two ligands were almost identical due to their similar
structures. It needs to be noted that only one DGA arm in one ligand
molecule participates in the coordination with one trivalent metal
ion. In other words, during the extraction process, most of the DGA
arms of these multiple-DGA ligands do not directly participate in
the coordination of trivalent metal ions, which seems to be inconsistent
with the atom economy principle of green chemistry.[19] Diglycolamide-functionalized calix[4]arenes (C4DGA) can
significantly improve the selectivity of the metal Am3+ compared with UO22+, while the selectivity
of the Eu ion is slightly decreased. Using this extractant, it is
possible to quantitatively extract trivalent lanthanide and actinide
ions (distributive ratio > 300) with little or no extraction of
UO22+ and Sr2+.[4]arenes showing unusual complexation of actinide ions in room
temperature ionic liquids: Role of ligand structure, radiolytic stability,
emission pectroscopy, and thermodynamic studies. Inorg. Chem.. 2013 ">20]
Figure 1
Chemical structures of some multiple-DGA-functionalized ligands
used to extract lanthanides and actinides.
Chemical structures of some multiple-DGA-functionalized ligands
used to extract lanthanides and actinides.In recent years, ionic liquids have been widely used in the extraction
and separation of trivalent lanthanides and actinides.[21−28] Under the same conditions, the ionic liquid can affect the extraction
ability and separation factor of the extractant on the trivalent lanthanide
and actinide ions. Previous studies on T-DGA ligands showed that,
under the same conditions, the DAm value
was high in molecular diluents (DAm =
11.1) but significantly lower in room-temperature ionic liquids (DAm = 0.08). In an ionic liquid medium, TREN-DGA
has a strong ability to extract trivalent lanthanide and actinidemetal ions.[14] The solvent extraction studies
suggested a unique selectivity reversal in the extraction of trivalent
actinides versus trivalent lanthanides, which was observed when extraction
studies in an ionic liquid was performed. Therefore, it is necessary
to develop novel DGA extractants with a higher extraction efficiency
for trivalent metal ions.To investigate the effects of macrocyclic
crown ether compounds
on the extraction and separation of trivalent lanthanides/actinides,
some extractants and synergistic agents derived from crown ethers
were developed. For example, combined with 4-benzoyl-3-phenyl-5-isoxazolone
(HPBI), the synergistic extraction D values of 15-crown-5
(15C5), benzo-15-crown-5 (B15C5, Figure S1), 1-benzyl-1-aza-15-crown-5 (BA15C5, Figure S1), and 2-[(1-aza-15-crown-5)-1-ylmethyl)]-4-(phenyldiazenyl)-naphthalen-1-ol
(PDN1A15C5, Figure S1) for Eu(III) ions
are 1.0, 0.5, 2.5, and 6.3 at pH 2.4, respectively.[29] Specifically, the La(III), Pr(III), and Eu(III) ions are
extracted as Ln(PBI)3·2S complexes (S = PDN1A15C5),
while the light lanthanoid(III) ions of the 4f-series (La–Gd)
form the extraction complex as Ln(PBI)3·S species
(S = B15C5 and 15C5). It should be noted that the addition of crown
ether increases the extraction efficiency and the separation factors
between the heavier Ln(III) and light Ln(III) are higher than those
found for their extraction with the extractant alone. Further modification
of the macrocyclic crown ether, N,N′-bis[(6-carboxy-2-pyridyl)methyl]-1,10-diaza-18-crown-6 (H2BP18C6, Figure ), showed that it has good separation performance for trivalent lanthanide
ions (La–Gd), and the extraction distribution ratios are distributed
within the range of 5.0 × 10–3–2 ×
102 (conditions: 0.05 mol/L bis-2-ethyl(hexyl)phosphoric
acid in o-xylene/0.001 mol/L H2BP18C6
in 1 mol/L NaNO3, 0.05 mol/L lactate at pH 3).[30] The above results can be explained based on
a comparison of the following parameters: the size of the crown cavity,
the diameter of cations, and the species of the donor groups in the
crown ether host. Macrocyclic crown ethers have potential research
prospects in the coordination and separation of Ln(III) and An(III).
Figure 2
Chemical
structure of the crown ether-derived ligand H2BP18C6 used
to separate lanthanides and actinides.
Chemical
structure of the crown ether-derived ligand H2BP18C6 used
to separate lanthanides and actinides.Based on the above research background, we designed a single DGA-arm-grafted
aza-18-crown-6 macrocyclic ligand (Cr6DGA, Scheme ), reducing the number of DGA arms, to explore
its extraction performance for trivalent lanthanide/actinide elements.
In this paper, we first synthesized and characterized Cr6DGA and systematically
studied its extraction behavior and the coordination mode with the
trivalent metal ions. The possible extraction mechanism of Ln(III)/An(III)
by this macrocyclic crown ether was experimentally and theoretically
studied.
Scheme 1
Synthesis Route to the Cr6DGA Ligand
Results
and Discussion
Preparation of the Ligand
The ligand,
Cr6DGA, was prepared
in a 25% yield according to Scheme . Note that the reaction conditions need to be strictly
controlled, and an appropriate eluent and elution order are very important
for purification of the crude product of Cr6DGA by column chromatography.
After two consecutive silica gel column purifications (300–400
mesh, CH2Cl2/MeOH (v:v) = 250:2; CH2Cl2/MeOH (v:v) = 250:1), pure Cr6DGA was obtained as a
pale yellow solid.
Solvent Extraction
First, the extraction
equilibrium
time of Am(III) and Eu(III) was investigated. It was found that the
extraction equilibrium (Figure a) was reached within 10 min. Additionally, the extended contact
time had also no adverse effect on extraction efficiency. Thus, to
ensure that the measured samples had reached equilibrium, all the
experiment time of extraction were carried out with the balanced contact
time of 1 h. According to the acidity extraction results in Figure b, it was found that
as the concentration of nitric acid increased from 1.0 × 10–1 to 3.0 mol/L, the extraction distribution ratios
of Am(III) and Eu(III) were increased from 0.3 to 9 and 0.5 to 18
in the 2.0 × 10–2 mol/L Cr6DGA/cyclohexanone
system, respectively. Increasing the concentration of nitric acid
caused the increase of NO3– concentration,
and the homo-ionic effect of NO3– promoted
the extraction efficiency of metal ions. This trend is similar to
the concentration of nitric acid extraction curves of DGA and multiple-DGA-functionalized
ligands. While in this concentration range of nitric acid, the crown
ether substructure reduced the extraction efficiency of metal ions
due to protonation.[17] This suggested that
the binding site of the metal ion to the Cr6DGA may be mainly in the
amide pocket, whereas the crown ether unit is most likely not involved
in the coordination.
Figure 3
(a) Effect of contact time on the extraction of Am3+ and Eu3+. Organic phase: 1.0 × 10–2 mol/L Cr6DGA in cyclohexanone. Aqueous phase: 3.0 mol/L HNO3. (b) Effect of the concentration of HNO3 on the
extraction of Am3+ and Eu3+. Organic phase:
2.0 × 10–2 mol/L Cr6DGA in cyclohexanone. (c)
Effect of the concentration of Cr6DGA (in mmol/L) on the extraction
of Am3+ and Eu3+. Organic phase: Cr6DGA in cyclohexanone.
Aqueous phase: 3.0 × 10 mol/L HNO3. (d) Effect of
the concentration of NaNO3 on the extraction of Am3+ and Eu3+. Organic phase: 2.0 × 10–2 mol/L Cr6DGA in cyclohexanone. Temperature: 25 ± 0.5 °C.
(a) Effect of contact time on the extraction of Am3+ and Eu3+. Organic phase: 1.0 × 10–2 mol/L Cr6DGA in cyclohexanone. Aqueous phase: 3.0 mol/L HNO3. (b) Effect of the concentration of HNO3 on the
extraction of Am3+ and Eu3+. Organic phase:
2.0 × 10–2 mol/L Cr6DGA in cyclohexanone. (c)
Effect of the concentration of Cr6DGA (in mmol/L) on the extraction
of Am3+ and Eu3+. Organic phase: Cr6DGA in cyclohexanone.
Aqueous phase: 3.0 × 10 mol/L HNO3. (d) Effect of
the concentration of NaNO3 on the extraction of Am3+ and Eu3+. Organic phase: 2.0 × 10–2 mol/L Cr6DGA in cyclohexanone. Temperature: 25 ± 0.5 °C.The influence of the ligand concentration on the
extraction of
Am(III) and Eu(III) is shown in Figure c; the logarithmic curve of DM and the concentration of the ligand has a straight slope
of 1.76 and 1.82, indicating that Cr6DGA with Am(III) and Eu(III)
mainly exist in a 2:1 extraction species, which is very similar to
other DGAs.[15,16]The effect of a salting-out
agent on the extraction efficiency
is shown in Figure d, which is similar to that of the TOGDA ligand. In the range of
0.5–3.0 mol/L, the extraction efficiency of Cr6DGA for metal
ions is positively correlated with the concentration of NaNO3. The results suggested that nitrate ions would promote the extraction
of metal ions by Cr6DGA. The logarithmic curve of DM and the concentration of NO3– ions had a straight slope of 2.2 and 1.9, indicating that the extraction
species mainly existed in the 2:1 form. However, it is unknown whetherNO3– ions participate in the inner coordination
of Am(III) and Eu(III) ions, which will be further discussed below.The distribution ratios of various DGA ligands to Eu(III) and Am(III)
ions in the 3 mol/L HNO3 solution are summarized in Table . As can be seen from Table , Cr6DGA has a high
selectivity to Eu(III), DEu is higher
than DAm, and the SFEu/Am value
is about 2. However, under the same conditions, TODGA has a slightly
higher selectivity for Am. It is worth noting that the selectivity
trend of Cr6DGA is consistent with that of other DGA extractants,
which all have better selectivity to Eu. The details of the extraction
mechanism of these trivalent metal ions by Cr6DGA will be discussed
further below.
Table 1
Extraction Behavior of Am(III)/Eu(III)
in the 3 mol/L HNO3 Solution Using Different DGA Ligands
at a Concentration of 2.0 × 10–2 mol/L (Temperature:
25 ± 0.5 °C)
metal ion
DCr6DGAa
DTODGAa
DT-DGAb,(16)
DTREN-DGAb,(16)
DBz-T-DGAb,(16)
DT9C3ODGAc,(18)
DT12C4ODGAc,(18)
Am(III)
9
105
11
0.4
235
71
255
Eu(III)
18
93
99
1
390
300
>103
Organic phase: cyclohexanone.
1.0 × 10–3 mol/L solutions
in a 10% IDA + 90% n-dodecane mixture.
1.0 × 10–3 mol/L
solutions in 5% IDA/n-dodecane medium.
Organic phase: cyclohexanone.1.0 × 10–3 mol/L solutions
in a 10% IDA + 90% n-dodecane mixture.1.0 × 10–3 mol/L
solutions in 5% IDA/n-dodecane medium.
ESI-MS Analysis
ESI-MS is particularly
suitable for
the mass determination of complexes because it can directly analyze
nonvolatile macromolecules in the liquid phase.[31] Methanol solutions with Eu(NO3)3 concentrations
of 5.0 × 10–3, 1.0 × 10–2, and 1.5 × 10–2 mol/L were stirred with 5.0
× 10–3 mol/L Cr6DGA for 1.0 h. In Figure , the peaks at m/z 709.8961 and 494.6093 were attributed
to the complex species [Eu(Cr6DGA)2NO3]2+ and [Eu(Cr6DGA)2(NO3)2 +
2H]3+, respectively, which were consistent with the isotopic
simulated cluster peaks. ESI-MS analysis showed that the binding ratio
of Eu(III) to Cr6DGA was 1:2 and the extracted complex contained the
nitrate ions, which were consistent with the results of extraction
experiments.
Figure 4
ESI-MS analysis of Eu(III) complexes with the Cr6DGA ligand
in
methanol ([Eu(NO3)3] = 5.0 × 10–3 mol/L; [ligand] = 5.0 × 10–3 mol/L): (a)
[EuL2NO3]2+ and (b) [EuL2(NO3)2 + 2H]3+. Temperature: 25
± 0.5 °C.
ESI-MS analysis of Eu(III) complexes with the Cr6DGA ligand
in
methanol ([Eu(NO3)3] = 5.0 × 10–3 mol/L; [ligand] = 5.0 × 10–3 mol/L): (a)
[EuL2NO3]2+ and (b) [EuL2(NO3)2 + 2H]3+. Temperature: 25
± 0.5 °C.
Luminescence Spectral Analysis
Solvent extraction results
showed that Cr6DGA and Eu(III) formed a 2:1 complex in the cyclohexanone
diluent, but it was unclear whether there were water molecules in
the inner layer of the Eu ion. The emission intensities for the 5D0 → 7F1 (λ
= 594 nm) and 5D0 → 7F2 (λ = 619 nm) transitions are shown in Figure .[32,33] The complexation constants for spectral titration were fitted using
the HyperSpec program,[34] and the complexation
constants of ML and ML2 types of complexes were 4.91 and
9.88, respectively, which were consistent with the results of extraction
experiments and ESI-MS analysis. However, in the case of T9C3ODGA,
the stability constants of ML and ML2 types of complexes
were 4.75 and 9.12, respectively.[16]
Figure 5
Fluorescence
spectroscopic titrations of Cr6DGA with Eu(NO3)3 in acetonitrile (conditions: [Eu(NO3)3] =
4.0 × 10–3 mol/L, volume
= 2.0 mL; [Cr6DGA] = 4.0 × 10–2 mol/L; temperature
= 25 ± 0.5 °C).
Fluorescence
spectroscopic titrations of Cr6DGA with Eu(NO3)3 in acetonitrile (conditions: [Eu(NO3)3] =
4.0 × 10–3 mol/L, volume
= 2.0 mL; [Cr6DGA] = 4.0 × 10–2 mol/L; temperature
= 25 ± 0.5 °C).
TRLFS Analysis
Time-resolved laser-induced fluorescence
spectra can provide information about the first coordination sphere
of Eu(III) by the efficient energy transfer from the excited states
of the metal ions to the ligand. There is a linear correlation between
the fluorescence lifetime and the number of water molecules (eq ). The fluorescence lifetime
τ (in ms) can be determined from the curve fitting with single
exponential functions, and the values are listed in Table . The fluorescence decay curves
with Cr6DGA/Eu(III) = 1:1 in methanol is shown in Figure . It can be seen that when
the hydration number decreases from 1.25 to 0.3, the molar ratio of
metal/ligand decreases from 1 to 0.25, indicating that the ligand
gradually replaces the water molecule entering the central Eu(III)
ion.
Table 2
Lifetime and Number of Water Molecules
of Cr6DGA with Eu(NO3)3 in Methanol at Different
M/L Molar Ratios (Initial Conditions: [Eu(NO3)3] = 2.5 × 10–3 mol/L, Volume = 1.5 mL; [Cr6DGA]
= 4.0 × 10–2 mol/L; Temperature = 25 ±
0.5 °C)
M/L ratio
1
0.5
0.25
τ
(ms)
0.57
0.78
1.16
N (H2O)
1.25
0.75
0.3
Figure 6
Fluorescence
decay curve of Eu(NO3)3 with
Cr6DGA in methanol solution (initial conditions: [Eu(NO3)3] = 2.5 × 10–3 mol/L, volume
= 1.5 mL; [Cr6DGA] = 4.0 × 10–2 mol/L; temperature
= 25 ± 0.5 °C).
Fluorescence
decay curve of Eu(NO3)3 with
Cr6DGA in methanol solution (initial conditions: [Eu(NO3)3] = 2.5 × 10–3 mol/L, volume
= 1.5 mL; [Cr6DGA] = 4.0 × 10–2 mol/L; temperature
= 25 ± 0.5 °C).Generally, for most DGA-derived ligands,
the internal coordination
of a metal ion usually does not involve water molecules.[35] However, for the ligand in this paper, there
is one water molecule in the inner coordination layer of the metal
ion. We speculate that for the 9-coordinated Eu(III) complex, its
internal coordination layer contains two ligands and one water molecule,
and there may also be 1–3 NO3– ions involved in the inner layer coordination. Based on the results
of the above experiments, it can be inferred that the composition
of the complex structure may be [Eu(Cr6DGA)2(NO3)(H2O)]2+ and [Eu(Cr6DGA)2(NO3)2(H2O)]+.
DFT Calculations
Although we have confirmed the composition
of the metal–ligand complexes, the specific binding site of
the ligand and metal ion is unclear since both the oxygen atom of
the crown ether and the amideoxygen atom may bind to the metal ion.
To identify the binding sites of the ligand with metal ions, we carried
out DFT calculations to optimize the structure of these two possible
complexes as shown in Figure S5 and found
that the binding energy to the DGA unit was 9.9 kcal/mol lower than
that at the “crown ether” unit (−11.5:–11.1,
water/cyclohexanone). Correspondingly, in the follow-up research,
we will focus on the binding mode of metal ions with the DGA unit
for further discussion.To elucidate the binding differences
between the coordination of the ligand with Am(III) and Eu(III), the
structures of the 1:2 species of [ML′2(NO3)(H2O)]2+ and [ML′2(NO3)2(H2O)]+ (M = Am and Eu)
complexes (Figure ) were optimized; these structures were based on the results of the
extraction study, spectrophotometric studies, and MS analysis. The
relevant structural parameters of Eu(III) and Am(III) complexes are
shown in Table , i.e.,
bond distance (d) and the Mayer bond order (MBO).
The metal–ligand distances (M–O(CO) and M–O(ether))
and metal–oxygen (M–O(nitrate) and M–O(water))
bond lengths are provided as average values. The Eu–O(CO) bond
distances were in agreement with the experimental values in Eu(TODGA)33+ (2.401 Å, EXAFS) complexes[36] and [Eu(TEDGA)3]3+ (2.389 Å,
crystal structure) complexes.[37] Additionally,
the Am–O(CO) bond lengths were also well consistent with those
in the Am(TMOGA)3(ClO4)3 (2.459 Å)
crystal structure.[38] Though the calculated
M–O(ether) bond distances were longer than the experimental
one by about 0.1 Å, this is quite understandable. The previous
computational values reported the bond lengths between the Eu(III)
and the ether O atom of TODGA within 2.56–2.65 Å, which
was in line with the calculated bond lengths.[39−41] It should be
noted that the M–O(CO) bond lengths in [ML′2(NO3)(H2O)]2+ complexes are shorter
than the M–O(CO) bond lengths in [ML′2(NO3)2(H2O)]+ complexes. Accordingly,
the bond orders of M–O(CO) in [ML′2(NO3)(H2O)]2+ complexes are higher than
those of M–O(CO) in [ML′2(NO3)2(H2O)]+ complexes.
Figure 7
Optimized structures
of metal complexes [ML′2(NO3)(H2O)]2+ and [ML′2(NO3)2(H2O)]+ (M
= Am (a, b), Eu (c, d)) in the gas phase at the B3LYP/def2-SVP/RECP
level.
Table 3
Calculated M–O
Distances (dM–X, in Å) and
Mayer Bond Order
(MBO) in [ML′2(NO3)(H2O)]2+ and [ML′2(NO3)2(H2O)]+ (M = Eu and Am) Complexes in the Gas Phase
Optimized structures
of metal complexes [ML′2(NO3)(H2O)]2+ and [ML′2(NO3)2(H2O)]+ (M
= Am (a, b), Eu (c, d)) in the gas phase at the B3LYP/def2-SVP/RECP
level.[Eu(TEDGA)3][Ln(NO3)6] crystal
structure (ref (36)).Am(TMOGA)3(ClO4)3 crystal structure (ref (37)).The Mulliken’s charges of Am(III) and Eu(III)
ions are significantly
smaller in comparison with the formal “+3” oxidation
state. This implies a strong ligand to metal atom charge transfer
(Table S1). In the metal complexes, the
net charge of Eu(III) is smaller than that of Am(III), which means
that the charge transfer from the ligand to Eu(III) is greater than
that of Am(III). In addition, the Mulliken’s spin population
of the metal (ρM) in the metal complexes obtained
in the gas phase is also provided (Table S1) because it was shown that ρM is an important parameter
for evaluating the bonding properties of the f-block
complexes. The spin population values (ρM) of the
metal atom of the discussed complexes were obtained by the Multiwfn
code using Mulliken’s[42,43] and Löwdin’s[44] methods. When the deviation between ρM and 6.00 is greater, the bonding interaction between the
ligand and the metal tends to be more covalent. Comparing the spin
population values (ρM) between Am and Eu complexes,
it was found that the Eu(III) complexes had a greater ρM value than the Am complexes by both Mulliken’s and
Löwdin’s methods.To understand the extraction
mechanism further, the possible complexation
reactions of the ligand and their complexation products [ML′2(NO3)(H2O)]2+ and [ML′2(NO3)2(H2O)]+ (M
= Eu and Am) were investigated. The liquid extraction system is a
very complicated process. For the process of liquid–liquid
extraction in the HNO3 medium, eq can be used to describe the simplified extraction
processes of Am(III) and Eu(III) with the studied Cr6DGA ligand.The single-point
energy calculations of the optimized Eu(III) and
Am(III) complexes were performed in the aqueous phase and the organic
(cyclohexanone) phase at the B3LYP/def2-SVP/RECP//def-TZVP level.
The changes in the complex formation energy (ΔEcf) of reaction in the gas, aqueous, and organic phases were calculated and
are given in Table . Obviously, the ΔEcf values for
the reaction with [ML′2(NO3)2(H2O)]+ complexes turn out to be more negative
than those of [M(L′)2(NO3)(H2O)]2+ complexes in all gas, aqueous, and organic phases,
which suggests that the extraction metal complexes are more likely
to form [ML′2(NO3)2(H2O)]+ species during the extraction process. As
expected, the Eu(III) complexes are more stable than the corresponding
Am(III) complexes in the gas phase (Δ(ΔEcf)Am/Eu = +4.1 and +3.7 kcal/mol for [ML′2(NO3)(H2O)]2+ and [ML′2(NO3)2(H2O)]+,
respectively). By contrast, when the solvent effect of water (PCM)
is taken into account, the differences between Am and Eu complexes
decrease to +1.0 and +0.7 kcal/mol, respectively. When changing to
cyclohexanone as a solvent, this trend is similar. The positive Δ(ΔEcf)Am/Eu values (+1.2 and +0.9 kcal/mol
for [M(L′)2(NO3)(H2O)]2+ and [ML′2(NO3)2(H2O)]+, respectively) indicate that the Cr6DGA ligand
has better selectivity for Eu(III) over Am(III). These results are
consistent with the extraction results.
Table 4
Complex
Formation Energies, ΔEcf (kcal/mol),
and the Energy Differences, Δ(ΔEcf)Am/Eu, of the Formation of Eu(III)
and Am(III) Complexes in the Gas Phase, Water, and Cyclohexanone,
Calculated at the B3LYP/def2-SVP/RECP//def-TZVP Level
complexes
metal ion
phase
ΔEcf
Δ(ΔEcf)Am/Eu
[ML′2(NO3)(H2O)]2+
Eu3+
gas
–334.4
water
–26.1
cyclohexanone
–42.3
Am3+
gas
–330.3
4.1
water
–25.1
1.0
cyclohexanone
–41.1
1.2
[ML′2(NO3)2(H2O)]+
Eu3+
gas
–469.3
water
–34.1
cyclohexanone
–56.6
Am3+
gas
–465.6
3.7
water
–33.4
0.7
cyclohexanone
–55.7
0.9
Conclusions
In this paper, a novel
single DGA-arm-grafted macrocyclic crown
ether ligand was synthesized to investigate its extraction behavior
for the Eu(III)/Am(III) cations. It was found that Cr6DGA was more
selective than TODGA for the trivalent lanthanide ions in a 3 mol/L
HNO3 solution, which may be explained by the introduction
of the crown ether structure in the ligand. The coordination behavior
of Cr6DGA with Eu(III) and Am(III) was identified by both slope analysis
and luminescence spectroscopy, suggesting an ML2 complex
with an inner-sphere water molecule. ESI-MS analysis indicated that
the extraction complexes may exist in the form of [Eu(Cr6DGA)2NO3(H2O)]2+ or [Eu(Cr6DGA)2(NO3)2(H2O)]+.
DFT calculations revealed that the binding ability of Eu(III) to the
DGA unit of Cr6DGA is stronger than that of the crown ether unit,
and [ML′2(NO3)2(H2O)]+ is more energy favorable. The calculation results
also show that the Eu(III) complexes are more stable than the corresponding
Am(III) complexes in gas, aqueous, and organic phases, which is in
good agreement with the selectivity of the extraction results. These
results can help us to understand the extraction mechanism of Am(III)
and Eu(III) ions by the DGA-type ligands and provide some insights
for the design of novel DGA-derived ligands for extraction of lanthanides
and actinides. However, the role of the crown ether unit during the
extraction process is quite complicate, which might be explained from
the comparison of the cavity size of crown ethers and the diameter
of cations, and the types of donor groups in the crown ether host.
Since the specific function of the crown ether structure is not clear
here, our group will continue to systematically study the selectivity
of crown ethers and their derivatives to trivalent lanthanide and
actinidemetal ions.
Experimental Section
Materials and Instruments
The reagents mentioned were
purchased from Aladdin (Shanghai Aladdin Biochemical Technology Co.
Ltd, China) and were of analytical or chromatographic grade. CH2Cl2 and THF were distilled from CaH2. The reactions were conducted under N2 unless otherwise
stated. 241Am and 152,154Eu, provided by the
China Institute of Atomic Energy, were used as radioactive tracers
in solvent extraction.The NMR spectra were recorded on a Varian
Inova NMR spectrometer (Bruker Inc., Switzerland). Electrospray ionization
mass spectrometry (ESI-MS) data were obtained by an LCMS-IT-TOF spectrometer
(Shimadzu, Japan). The concentration of Eu was determined using an
inductively coupled plasma optical emission spectrometer (ICP-OES,
Optima8000, PerkinElmer).The activities of 241Am
and 152,154Eu were
measured by a NaI (TI) scintillation counter. Luminescence spectroscopic
titration was obtained on a fluorescence spectrophotometer (F96pro,
China). The Transient Fluorescence spectra was measured on a Fluorolog-3
spectrofluorometer (JobinYvon, Horiba).
Synthesis of Cr6DGA
Compounds 1 and 2 were synthesized according
to the reported methods.[45,46] The synthesis of Cr6DGA
was as follows: compound 1 (0.1
g, 0.56 mmol) was dissolved in 25 mL of anhydrous CH2Cl2. Then, compound 2 (0.36 g, 0.56 mmol), chlorotripyrrolidinyl
hexafluorophosphate (0.24 g, 0.56 mmol), and diisopropylethyl amine
(DIEA, 0.16 g, 1.25 mmol) were added in one portion. The mixture was
stirred at room temperature (RT) in a nitrogen atmosphere for 4 days.
Subsequently, it was washed three times with 10 mL of 5% hydrochloric
acid. The crude product was collected by removing the organic solvent
under vacuum and was purified by a silica gel column (300–400
mesh; CH2Cl2/MeOH (v:v) = 250:2; CH2Cl2/MeOH (v:v) = 250:1) twice. Finally, the desired product
(Cr6DGA) was obtained as a pale-yellow solid with a Rf value of 0.5 (CH2Cl2/MeOH (v:v)
= 10:1) and a yield of 25%. m.p.: 39–41 °C. HRMS: m/z calcd for Cr6DGA: 603.4579 [M + H]+; found: 603.4568. 1H NMR (400 MHz, CDCl3): δ = 4.48 (s, 4H), 3.88–3.46 (m, 24H), 3.23 (t, 2H),
3.09 (t, 2H), 1.50 (bs, 4H), 1.40–1.15 (m, 20H), 0.88 (t, 6H). 13C NMR (100 MHz, CDCl3): δ = 14.0, 22.94,
26.72, 27.29, 27.47, 28.78, 29.22, 29.38, 29.47, 29.57, 31.80, 31.92,
46.23, 47.07, 68.44, 76.70, 77.02, 77.33, 168.77.The organic phases were prepared
in cyclohexanone. For a single metal ion solution, it was prepared
by dissolving metal ions in nitric acid solutions of different concentrations.
The effects of various parameters on the extraction distribution ratio,
including the contact time, nitric acid concentration (1.0 ×
10–1–3.0 mol/L HNO3), extractant
concentration (4.0 × 10–3–2.0 ×
10–2 mol/L), and NaNO3 concentration
(0.5–3.0 mol/L), were evaluated using radioactive tracers of
the metal ion. The selectivity experiment was performed with 2.0 ×
10–2 mol/L extractant in the presence of 152,154Eu and 241Am in various concentrations of HNO3 solution. The experiments were carried out using the following standard
protocol: first, the organic phase was pre-equilibrated with HNO3 solution at least three times in a phase ratio of 1:1 at
25.0 ± 0.5 °C. Second, the aqueous phases were spiked with
trace amounts of 241Am and 152,154Eu in nitric
acid solution, and equal volumes of two phases were mixed thoroughly
for 1 h in a thermostated water bath (25 ± 0.5 °C). After
the phase separation by centrifugation, the count rates of 241Am and 152,154Eu were determined by a NaI (TI) scintillation
counter. DM is defined as the ratio between
the radioisotope activity or the element concentration in the organic
and aqueous phases. The selectivity for 152,154Eu over 241Am is represented by the separation factor, SFEu/Am, which is defined as the ratio of the distribution ratios DEu/DAm. All of the
extraction experiments were conducted twice under the same conditions,
and the reported values are the average of these values.LCMS-IT-TOF (Shimadzu, Japan) combines
QIT (ion trap) and TOF (time-of-flight) technology, which can effectively
introduce ions into QIT and spray to capture ion TOF at the same time.
LCMS-IT-TOF can greatly assist in the identification of target compounds
using high-speed/high-precision MSn (n ≥ 10). In the high-resolution mode, the detection range of
the instrument is 50–5000 m/z. Methanol solutions with Cr6DGA concentrations of 5.0 × 10–3, 1.0 × 10–2, and 1.5 ×
10–2 mol/L were stirred with 5 × 10–3 mol/L Eu(NO3)3 for 1.0 h, and the Eu(III)
complexes were analyzed.
Luminescence Spectroscopic Titration
Luminescence spectroscopic
titration was realized by changing the ratio of metal/ligand in the
acetonitrile medium. Acetonitrile is a suitable solvent because of
the solubility of both Eu(NO3)3 and Cr6DGA.
The fluorescence titration was performed in a 10 mm quartz cell at
25 ± 0.5 °C. The initial concentrations of Cr6DGA and Eu(NO3)3 were 4.0 × 10–2 and 4.0
× 10–3 mol/L, respectively, controlling the
titration ratio of metal/ligand from 0.2:1 to 4:1. The emission spectrum
of Eu(III) was obtained in the range of 575–630 nm with an
excitation wavelength of 395 nm and a bandwidth of 10 nm. The stability
constants of the complexes were calculated by the HyperSpec program.[34]Time-resolved laser-induced
fluorescence
spectra (TRLFS) were measured on a Fluorolog-3 spectrofluorometer
with a SpectraLED (390 nm, S-390) as the excitation source and a picosecond
photon detection module (PPD-850) as the detector. The TRLFS analysis
test was carried out in a 10 mm quartz cell, and a methanol solution
with Eu(III) at concentrations of 2.5 × 10–3, 5.0 × 10–3, and 1 × 10–2 mol/L was mixed with a methanol solution of Cr6DGA (1 × 10–3 mol/L) with a phase ratio of 1:1 at 25 °C.All geometries were optimized with
the hybrid B3LYP[47,48] functional implemented in Gaussian
09D software.[49] For Eu and Am metal atoms,
relativistic effects were taken into consideration with the quasi-relativistic
effective core potential (RECP). The corresponding ECP60MWB-SEG[50−52] and ECP28MWB-SEG[53,54] valence basis sets were used
to describe Am and Eu atoms, respectively. The electronic configuration
of Eu(III) and Am(III) in their septet state was chosen as the ground
state of their complexes.[55−57] For all other light atoms (C,
H, O, and N), the def2-SVP basis set[58,59] was employed
for optimization. Frequency calculations were performed to ensure
that the obtained stationary points were the minima on the potential
energy surface. In summary, all of the geometries were optimized at
the B3LYP/def2-SVP/RECP level in the gas phase. Many reported studies
stated that the structures of the optimized complexes in aqueous and
organic solutions do not change obviously in comparison with those
in the gas phase.[60,61] Therefore, solvation energies
in the aqueous and organic phases were estimated by single-point calculations
with the def-TZVP[62] basis set for light
atoms and the same basis set for Am and Eu with the corresponding
ECP based on the structures optimized in the gas phase. The self-consistent
reaction field (SCRF) polarizable continuum model (PCM)[63] was used to take into account the solvent effect
of water and cyclohexanone (dielectric constants of 78.4 for water
and 15.6 for cyclohexanone) according to the experimental report.
The two solvents considered in this work, water and cyclohexanone,
were used in experimental studies. To reduce the calculation cost,
all alkyl chains connected with the amidenitrogen atom of the ligand
were replaced by a methyl group, and this simplified ligand is represented
as L′. We believe that this approximation is acceptable, because
the length of the alkyl chain in DGA has no decisive effect on the
separation behavior of Am(III) and Eu(III).[64] The Mayer bond order (MBO), Mulliken atomic charges, and Löwdin’s
and Mulliken’s spin populations (ρM) of the
metals in complexes were determined using Multiwfn software.[65] Visualizations were created with CYLview software.[66]