Survivorship and Patient-Reported Outcomes of an Uncemented Vitamin E-Infused Monoblock Acetabular Cup: A Multicenter Prospective Cohort Study.

Fahd F Mahmood1, Martin Beck2, Arthur de Gast3, Philipp Rehbein4, Gary J French5, Roland Becker6, Martin Dominkus7, Naeder Helmy8, Lutz Hollmann9, Joe Baines1.   

Abstract

BACKGROUND: Addition of vitamin E to polyethylene is theorized to reduce the potential for oxidative wear in acetabular components. This paper presents a multicenter prospective cohort study that reports on outcomes from use of a Vitamin E-infused highly cross-linked polyethylene acetabular cup.
METHODS: Patients were recruited across nine medical institutions. Clinical outcome measures recorded were the Harris Hip Score, visual analogue score for pain and satisfaction. Evidence of implant loosening or osteolysis was collected radiologically. Cup survival and reasons for revision in relevant cases were also recorded. Data collection was undertaken preoperatively, at 6-12 weeks, 6 months, 1 year, 2 years, and 5 years. A total of 675 patients were recruited, with 450 cases available at final review. Data regarding cup survival was available to 8 years and 9 months postoperatively.
RESULTS: Improvements in both the Harris Hip Score and visual analogue score for pain and satisfaction were recorded at all time points, with these being maintained through the length of follow-up. In total, 89% of cups were implanted within the Lewinnek safe zone. A lucent line was identified in one case, with no evidence of acetabular osteolysis observed throughout the follow-up period. Cup survival was 98.9% at 8 years and 9 months. No revisions for aseptic loosening were observed.
CONCLUSIONS: The use of a vitamin E-infused polyethylene acetabular cup demonstrates reassuring patient-reported outcomes, radiological measures, and cup survival at medium to long-term follow-up.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  aseptic loosening; polyethylene; survival; total hip replacement; vitamin E

Year:  2020        PMID: 33516632     DOI: 10.1016/j.arth.2020.12.032

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  1 in total

1.  Vitamin E-Enhanced Liners in Primary Total Hip Arthroplasty: A Systematic Review and Meta-Analysis.

Authors:  Qian-Yue Cheng; Bin-Fei Zhang; Peng-Fei Wen; Jun Wang; Lin-Jie Hao; Tao Wang; Hui-Guang Cheng; Ya-Kang Wang; Jian-Bin Guo; Yu-Min Zhang
Journal:  Biomed Res Int       Date:  2021-12-06       Impact factor: 3.411

  1 in total