Literature DB >> 33487928

Objective Assessment of the Quality and Accuracy of Deformable Image Registration.

Ines-Ana Jurkovic1, Nikos Papanikolaou1, Sotirios Stathakis1, Neil Kirby1, Panayiotis Mavroidis2.   

Abstract

BACKGROUND: The increased use of deformable registration algorithms in clinical practice has also increased the need for their validation. AIMS AND
OBJECTIVES: The purpose of the study was to investigate the quality, accuracy, and plausibility of three commercial image registration algorithms for 4-dimensional computed tomography (4DCT) datasets using various similarity measures.
MATERIALS AND METHODS: 4DCT datasets were acquired for 10 lung cancer patients. 23 similarity measures were used to evaluate image registration quality. To ensure selected method's invertibility and assess resultant mechanical stress, the determinant of the Jacobian for the displacement field and 3-D Eulerian strain tensor were calculated. All the measures and calculations were applied on to extended deformable multi pass (EXDMP) and deformable multi pass (DMP) methods.
RESULTS: The results indicate the same trend for several of the studied measures. The Jacobian determinant values were always positive for the DMP method. The Eulerian strain tensor had smaller values for the DMP method than EXDMP in all of the studied cases. The negative values of the Jacobian determinant point to non-physical behavior of the EXDMP method. The Eulerian strain tensor values indicate less tissue strain for the DMP method. Large differences were also observed in the results between complete and cropped datasets (coefficient of determination: 0.55 vs. 0.93).
CONCLUSION: A number of error and distance measures showed the best performance among the tested measures. The evaluated measures might detect CT dataset differences with higher precision if the analysis is restricted to a smaller volume. Copyright:
© 2020 Journal of Medical Physics.

Entities:  

Keywords:  4DCT; deformable image registration; image dissimilarity indices; Jacobian determinant; image similarity measures; strain tensor

Year:  2020        PMID: 33487928      PMCID: PMC7810144          DOI: 10.4103/jmp.JMP_47_19

Source DB:  PubMed          Journal:  J Med Phys        ISSN: 0971-6203


INTRODUCTION

Deformable image registration (DIR) is extensively used in radiation therapy applications. Possible DIR uses include auto-segmentation of structures,[1] dose accumulation,[2] and treatment optimization. A multi-institution study[3] suggested caution in globally accepting the results from deformable registration. Studies that evaluated different DIR algorithms, including the most common commercial software, showed reasonable overall accuracy of the registration; however, they also observed large DIR errors in some of the studied cases.[345] Thus, it is necessary to assess accuracy of DIR software before it is used for any of its radiation therapy applications. Several studies suggested different ways of quantifying the registration results.[3678] Many of the studies in this field use patient-specific phantoms to evaluate DIR algorithms by looking at fixed points in the images (control points) and measuring their displacement in the registration resultant image (landmark error). Landmark matching is a process that has been applied on large image datasets; however, the limited number of images that are usually analyzed may not give a complete and accurate picture of DIR performance on the whole volume. Furthermore, identifying landmarks is time consuming process as well as not feasible for every patient. Beyond this, there is need for a fast and reliable metrics that can be automatically applied on large volume datasets and can provide a dependable tool for the DIR assessment. Many publications examined the Dice, Tanimoto, or mutual information (MI) measures, but some of these measures proved to be unreliable when used as the only tool for DIR assessment. For example, Dice can show improved similarity when compared with the original unwrapped data[9] but nonphysical behavior may exist, which may strongly influence the accumulated dose distributions and/or contour segmentation. The term “nonphysical behavior“ refers to the situations where the produced registration features are physically impossible based on the geometrical and mechanical characteristics of the involved tissues. Hence, to assess for possible nonphysical behavior of DIR, it is essential to understand tissue mechanics.[1011] This study utilizes a combination of metrics to assess both image similarity and for nonphysical deformation scenarios to evaluate DIR accuracy. More specifically, a broad range of image similarity measures are utilized here, which include methods previously applied in radiation oncology for this purpose and also several similarity measures[361213141516] that were developed and utilized in other fields. In deformable registration, each voxel occupies its own location in space for both the initial and deformed configuration. However, no material element should be permitted to invert as it leads to nonphysical transformations.[17] The goal of the transformation is to be plausible or at least locally invertible. For any transformation to satisfy this requirement, according to the inverse function theorem, it is sufficient for it to be continuously differentiable and have a positive Jacobian determinant (J). As a measure to estimate the expansion and contraction during the deformation (i.e., volume change), the Jacobian determinant is widely used.[18] To avoid a region of positive finite volume to be deformed into a region of zero, negative (folding), or infinite volume, it is required that 0< J <∞. Tissues are composite materials that are continually changing and their behavior is described by continuum models, which have been developed and used in continuum mechanics and biomedical engineering. To describe the kinematics and mechanics of deformations different strain measures are employed as well as measures of volume and surface changes. The most common strain measures are the Lagrangian or material strain tensor and the Eulerian or spatial strain tensor, which are defined by means of the deformation gradient as the basic measure of local deformation and rotational motion. The Jacobian determinant and the Eulerian strain tensors are used here to evaluate for nonphysical deformation scenarios.

MATERIALS AND METHODS

Patient selection

To determine which measure has the best correlation with DIR performance, the metrics were applied to 4DCT lung datasets. For each patient 10 computed tomography (CT) datasets were available (the breathing cycle was sampled at 10 different phases). The work was approved by the appropriate ethical committees related to the institution in which it was performed and that subjects gave informed consent to the work. For evaluation purposes, the phase 50 (end of exhale) was registered to the phase 0 (end of inhale), which was chosen as the primary CT set for DIR. Selected patients had varying tumor volume sizes [Table 1].
Table 1

Tumor volume size per patient

Patient #12345678910
Tumor volume (cm3)1.711.02.5108.914.129.921.796.623.517.8
Tumor volume size per patient

Deformable image registration

Several registration options are available in Velocity AI (Varian Medical Systems, Palo Alto, CA): DICOM, rigid, rigid + scale, deformable, deformable multi pass (DMP), rigid + DMP, extended DMP (EXDMP) and structure-guided deformable. Velocity's primary registration algorithm uses a multi-resolution approach based on Mattes MI, the transform used is a cubic B-Spline, the interpolator used is a bi-linear interpolation and the optimizer is based on the method of steepest gradient descent. As far as, degrees of freedom of this approach Velocity is using a B-Spline of order 3 (cubic) with a uniform knot vector. The number of control points (per-dimension) is configurable with a minimum of 5 control points per-axis (no other constraints are imposed onto this value). The registration methods chosen for evaluation were rigid (RIGID – translation and rotation in the x, y, and z planes), DMP (a three pass coarse to medium to fine resolution deformable that yields finer touch up) and EXDMP (a six pass deformable that goes into finer resolution than the DMP). A workflow chart for the Rigid, DMP, and EXDMP registration algorithms is shown in Figure 1. DMP performs DIR sequentially from low to high resolution, i.e., after registration has been completed in one resolution stage, results are used as initial conditions for the next stage. The resolution used in each stage is determined automatically. The multi-resolution approach increases the number of control points used by the B-Spline transform between successive resolution levels. The software manufacturer suggests the use of the DMP method for CT to CT registration and the EXDMP when DMP fails to provide satisfactory results.
Figure 1

A workflow chart for the rigid, deformable multi pass and extended deformable multi pass registration algorithms

A workflow chart for the rigid, deformable multi pass and extended deformable multi pass registration algorithms

Evaluated similarity/dissimilarity measures

For DIR evaluation, registration between the 0 and 50 phases was used for comparison and DIR accuracy assessment, where the 0 phase dataset represented the reference image set. In this study, 23 measures were evaluated using two groups of 3D datasets: The complete CT dataset and the cropped CT dataset (3D CT dataset cropped to the tumor volume region). These measures were: The cross correlation (CC), normalized CC (NCC), distance correlation (DC), root mean squared error (RMSE), normalized absolute error, mean norm of the difference (MND), structural similarity index (SSIM), feature similarity index (FSIM), dimensionless global error (ERGAS), gradient magnitude similarity deviation (GMSD), quality index (Q), Dice similarity coefficient (DSC), Tanimoto coefficient (TC), bias (B), Bray-Curtis dissimilarity (BCD), Pearson correlation coefficient (PCC), Spearman rank correlation coefficient (SRCC), Euclidean distance (ED), Morisita-Horn dissimilarity (MHD), Sorensen dissimilarity (SD), simple matching dissimilarity (SMD), structural content (SC), and the 2D voxel mapping (MI is used in the Velocity AI algorithm, so it was not used in the assessment of the algorithm performance).[361213141516] A short description of those measures is provided in Appendix 1.

Nonphysical behavior evaluation (deformation and strain)

For each of the studied cases and for each of the two examined DIR methods (EXDMP and DMP), the binary deformation fields were exported from the Velocity AI (Varian Medical Systems, Palo Alto, CA) and the data were used in MATLAB for the deformation field assessment. First, the deformation gradient, F, was calculated. The calculated deformation gradient F can be decomposed into the product of a proper orthogonal tensor®, describing the rigid body displacements, and a symmetric tensor (U), describing the stretch deformation:[19] F = RU or F = VR      (1) where U is the right stretch tensor and V is the left stretch tensor. Based on these two stretch tensors, two commonly used deformation tensors are defined, the right Cauchy-Green tensor C (= U2), and the left Cauchy-Green tensor B (= V2). Both deformation tensors can be obtained from the deformation gradient: F      (2)

Using the same approach it can be easily verified that

FF      (3) The tensor that was used in the calculations is the Eulerian strain tensor which is defined as follows: where I is the identity matrix. It can be noted that if there is no deformation B-1 = I and e* = 0. A change in the volume due to deformation can be calculated using the Jacobian determinant, and it is defined as:[20] dV = JdV      (5) For incompressible material dV = dV0 and J = 1. In the above expression the Jacobian determinant of the deformation is defined as the determinant of the deformation gradient:[21] J = detF      (8)

RESULTS

There is a considerable computational time difference between the DMP and the EXDMP methods. For all the studied cases, it took 3–4 times longer per image set to complete the EXDMP registration compared to the DMP registration for the same datasets. Not all the measures were evaluated using the complete CT datasets mainly due to the computational memory limitations and the large number of data being evaluated. For example, voxel mapping was applied only on the 3D datasets which were cropped to the wider volume of the tumor location. The same goes for several other coefficients among which were the Dice and Tanimoto. Since this study does not perform an inter-comparison or evaluation of the different measures, this issue has no impact on the analysis. What is important in this analysis is the use of the same CT image volume (complete or cropped) for all three image registration algorithms per measure. Rigid registration visibly produced the worst registration results, which were validated with the overall outcome of the implemented measures. For the complete CT datasets and taking into the account all the evaluated measures, it was found that the RIGID registration was the worst in 75% of the cases, and for the cropped volume data in 96% of the cases.

Evaluated measures

The measures that consistently outlined the RIGID registration as the least accurate (in both datasets) were the Q and DC (CC, RMSE, and GMSD produced same result in 90% of the cases). From the measures that were taken using the limited volume datasets, the RIGID registration was outlined as the worst registration in 100% of the cases for these measures: CC, RMSE, MND, GMSD, DC, PCC, SRCD, BCD, ED, ERGAS, Q, MHD, and 2D voxel mapping. The breakdown of some of the results is shown in Table 2, where the DC and Q measures are used as an example.
Table 2

Results of the distance correlation and universal quality index metrics

MeasureComplete datasetCropped dataset


Measure outlined best registration methodIn percentage of casesMeasure outlined worst registration methodIn percentage of casesMeasure outlined best registration methodIn percentage of casesMeasure outlined worst registration methodIn percentage of cases
DCEXDMP50EXDMP0EXDMP70EXDMP0
DMP50DMP0DMP30DMP0
RIGID0RIGID100RIGID0RIGID100
QEXDMP80EXDMP0EXDMP80EXDMP0
DMP20DMP0DMP20DMP0
RIGID0RIGID100RIGID0RIGID100

EXDMP: Extended deformable multi pass, DMP: Deformable multi pass, DC: Distance correlation, Q: Quality index

Results of the distance correlation and universal quality index metrics EXDMP: Extended deformable multi pass, DMP: Deformable multi pass, DC: Distance correlation, Q: Quality index According to our results, MI did not perform as well on the cropped dataset as it did on the complete dataset, which to some extent is contradictory to the results of some other measures (such as the MND, FSIM, ERGAS, and B), which would give better results on the cropped dataset. Furthermore, the results acquired based on the cropped dataset indicate the EXDMP registration as the favorable one in more cases compared to the results of the complete datasets for the same measures. Unfortunately, due to the computational memory limitations, some of the measures that performed well in the cropped datasets evaluation were not assessed for the complete datasets. These measures are: The BCD, PCC, SRCD, ED, MHD, SD, SMD, and the 2D voxel mapping. The results of the 2D voxel mapping with the corresponding coefficients of determination could be independently validated using the Velocity AI 2D voxel map response option. The cropped comparison volumes were larger in the calculation that was done in MATLAB, using rectangular regions of interest (more voxels included in comparison), than the volumes used for calculation in Velocity AI. As shown in Table 3, the data indicate that the value of R2 (coefficient of determination) increases with the volume involved in mapping (larger number of voxel points).
Table 3

Two dimensional voxel mapping R2 values comparison per patient (best value in bold)

Patient #MethodCropped datasetVelocity AI tumor volume + 0.5 cmVelocity AI tumor volume + 5.0 cm
1DMP0.960.500.91
EXDMP0.960.740.92
RIGID0.910.100.75
2DMP0.900.450.82
EXDMP0.930.610.88
RIGID0.710.160.60
3DMP0.950.710.91
EXDMP0.950.750.92
RIGID0.890.420.81
4DMP0.960.860.93
EXDMP0.960.860.93
RIGID0.930.680.87
5DMP0.930.640.90
EXDMP0.930.800.91
RIGID0.910.380.85
6DMP0.950.530.79
EXDMP0.950.600.78
RIGID0.880.230.63
7DMP0.940.750.88
EXDMP0.950.810.92
RIGID0.870.150.76
8DMP0.950.780.91
EXDMP0.950.770.91
RIGID0.920.620.83
9DMP0.930.160.84
EXDMP0.940.650.90
RIGID0.900.100.79
10DMP0.980.800.93
EXDMP0.970.760.91
RIGID0.930.290.80

EXDMP: Extended deformable multi pass, DMP: Deformable multi pass, AI: Artificial intelligence

Two dimensional voxel mapping R2 values comparison per patient (best value in bold) EXDMP: Extended deformable multi pass, DMP: Deformable multi pass, AI: Artificial intelligence No correlation was found between the tumor volume size and any of the measures regardless of the size of the 3D volume being evaluated. The differences between the measures' values are shown in Figure 2 for the three most common measures, namely the CC, MI and TC. Patients #2 and #6 showed the largest difference in measure values, when compared to the rest of the patients, regardless of the registration method assessed. These two patients have tumor volumes located in the lower lung and posteriorly.[52223] Figure 2 illustrates the variability of behavior of the similarity metrics per patient.
Figure 2

Comparison of the resulting values of the three most commonly used measures in the deformable image registration accuracy assessment (the data were obtained using the complete three-dimensional dataset)

Comparison of the resulting values of the three most commonly used measures in the deformable image registration accuracy assessment (the data were obtained using the complete three-dimensional dataset) When the results are broken down per patient [Table 4], it is seen that when all the calculated measures were taken into account for cropped volumes, the DMP and EXDMP methods share the occurrences in the best value column equally (50:50). Even when we select only the measures, which indicate the RIGID transformation is the worst (least accurate) method, the ranking stays the same [Table 5]. Overall, across all the measures and evaluated volume datasets, DMP was ranked as best in 61% of the cases, EXDMP in 34% of the cases, and RIGID in 5% of the cases for the complete dataset.
Table 4

List of the method preferences using all the studied measures for the cropped three-dimensional computed tomography dataset

Percentage occurrence, all measures

Patient #DMPEXDMP
14258
20100
3955
4955
56337
68911
70100
83763
92663
10955

The largest values per patient are shown in bold. EXDMP: Extended deformable multi pass, DMP: Deformable multi pass

Table 5

List of the method preferences using only the measures where the RIGID method was found to be the least accurate one, using the cropped three-dimensional computed tomography dataset

Percentage occurrence, all measures

Patient #DMPEXDMP
13565
20100
3946
4946
55941
68812
70100
82971
92971
101000

The largest values per patient are shown in bold. EXDMP: Extended deformable multi pass, DMP: Deformable multi pass

List of the method preferences using all the studied measures for the cropped three-dimensional computed tomography dataset The largest values per patient are shown in bold. EXDMP: Extended deformable multi pass, DMP: Deformable multi pass List of the method preferences using only the measures where the RIGID method was found to be the least accurate one, using the cropped three-dimensional computed tomography dataset The largest values per patient are shown in bold. EXDMP: Extended deformable multi pass, DMP: Deformable multi pass Since the results of the registration method accuracy varied so widely across the studied measures, the sensitivity analysis was performed for the most prominent measures as suggested in the study by Yaegashi et al.[7] The similarity measures of the 4DCT images were evaluated with respect to the 50 phase CT dataset. The image similarity with respect to this phase decreases as the respiratory phase increases. To find which measure is the most sensitive we looked at the rate of change of each measure. The dissimilarity measures used were converted to similarity measures and the error and distance measures were normalized to produce compatible comparisons. Yaegashi et al. looked at the image per image correspondence (calculating the degree of similarity between two images), while in our study the complete 3D volume was used for assessment, which may explain some of the differences found between the two studies. The aforementioned study suggested the MI measure as the most sensitive one, whereas according to our study other measures appear to be more sensitive, such as the SSIM, ED, MND, RMSE, and TC [Figure 3]. Among the similarity measures, RMSE, MND, and GMSD indicated the RIGID method as the least accurate one in 100% of the cases, while the TC and FSIM gave the same result in 90% of the cases.
Figure 3

Comparison of the measures for each respiratory phase. The measures were applied on the cropped dataset

Comparison of the measures for each respiratory phase. The measures were applied on the cropped dataset Figure 3 also illustrates the measures that have comparable image similarity sensitivities, namely DC, CC, and MHD with their values almost constant at about 1.0. The 2D voxel mapping method accuracy results, which were obtained through the corresponding coefficient of determination, matched exactly the CC, Pearson correlation dissimilarity, and the Morisita-Horn dissimilarity results from the cropped CT dataset. The Tanimoto coefficient method accuracy results matched the results obtained by simple matching dissimilarity, Sorensen dissimilarity, and gradient magnitude similarity deviation. Patient #1 is a patient with the smallest tumor volume and is also the one that consistently showed rigid transformation as the best one for the number of used measures when the complete dataset was evaluated.

Quantifying deformation and strain

For all the cases, and the two evaluated methods, Eulerian strain tensors were calculated and their maximum values were compared with published data.[2425] Since the volume data that were used for calculation consisted of various tissues with different mechanical properties and biochemical data, the results covered a wide range of values [Table 6].
Table 6

Mechanical properties of different tissues as assessed from the deformation data

Patient #Eulerian strain tensor

EXDMPDMP
10.700.39
20.880.31
30.530.11
40.550.10
50.500.05
61.330.30
71.050.29
80.870.17
90.820.16
100.780.12

EXDMP: Extended deformable multi pass, DMP: Deformable multi pass

Mechanical properties of different tissues as assessed from the deformation data EXDMP: Extended deformable multi pass, DMP: Deformable multi pass The strain tensor comparison shows consistently larger values for the EXDMP method implying larger mechanical deformations as indicated by the Eulerian strain tensor, which is also confirmed by the minimum Jacobian determinant values [Table 7]. The only exception in the above pattern is patient #4, who is the only patient that exhibits plausible physical behavior when EXDMP method is used (J > 0).
Table 7

Jacobian determinant scalar values used for the evaluation of the nonphysical deformable image registration behavior

Patient #Minimum Jacobian determinant

EXDMPDMP
1−0.410.21
2−0.560.56
3−0.180.61
40.000.77
5−0.250.74
6−0.370.42
7−1.190.54
8−0.320.36
9−0.810.57
10−0.630.63

EXDMP: Extended deformable multi pass, DMP: Deformable multi pass

Jacobian determinant scalar values used for the evaluation of the nonphysical deformable image registration behavior EXDMP: Extended deformable multi pass, DMP: Deformable multi pass Local tissue expansion corresponds to a Jacobian determinant >1 and local tissue contraction corresponds to a Jacobian <1. The results for the DMP method based on the J minimum values in Table 7 indicate that in all the studied cases a certain amount of tissue contraction is observed. The difference in behavior between the two studied methods is even more visible in the Jacobian determinant color map for one of the central transverse slices of the cropped CT dataset for the two different patient cases [Figure 4].
Figure 4

Jacobian determinant map emphasizing the transformation difference between the extended deformable multi pass and deformable multi pass method

Jacobian determinant map emphasizing the transformation difference between the extended deformable multi pass and deformable multi pass method

DISCUSSION

The results of DMP showing the best performance in many cases were unexpected due to the fact that the EXDMP method has a longer computational time. By further analyzing the obtained results from the cropped datasets, it could be seen that in the cases where for evaluated measures EXDMP is predominantly best (and RIGID constantly worst) DMP was always better for the patients 3, 4, or 6. Interestingly, only these three patients from all the evaluated cases had tumors located in the upper lung and posteriorly, and this was independent of the tumor volume size as these three cases have widely different tumor sizes [Table 1]. The measures that always scored RIGID registration as the worst one were CC, RMSE, MND, GMSD, 2D voxel map, DC, PCC, SRCC, BCD, MHD, ERGAS, and Q for the cropped CT dataset and DC and Q for the complete CT dataset. The sensitivity study showed that RMSE, MND, ED, GMSD, TC, and FSIM measures have the highest image similarity sensitivity and at the same time found RIGID registration to be the least accurate one in more than 90% of the studied cases. This suggests that these measures can be used for DIR accuracy evaluation. Based on both the Jacobian and strain tensor calculations, it can be noted that while one DIR method may be helpful for the task of contour propagation it can be at the same time problematic when used for the task of dose accumulation and/or for the task of measuring a local volume change. The strain tensor values for the DMP method are well associated with published data, which report that the ultimate tensile strain for different tissues, such as tendon, ligament, skin, and aorta varies from 0.1 to 1.2 in one of the studies[24] and from 0.14 to 0.18 for ligament and tendons in another one.[25] The Jacobian determinant also indicated nonphysical behavior from EXDMP. Together, the Jacobian determinant and strain measures can give valuable information of the DIR's physical behavior. The development of all the inclusive measures for the evaluation of a DIR algorithm, which will take into account all the aforementioned issues (accuracy, quality, similarity, sensitivity, and plausibility) will be addressed in future work. At present, based on the results of the nonphysical behavior analysis and the results of similarity measure analysis, only the GMSD, SD, SMD, and TC indicated the DMP method as the best one in 80% of studied cases for the cropped dataset, and only one measure-B, indicated the same for the complete dataset, without picking RIGID as one of the best methods. Finally, the DMP method was shown to be better than the EXDMP when it comes to the physicality of the deformable registration and to the correct assessment of the volume change and mechanical stress in the deformation process. Although the present study presents some interesting findings, it is also subject to a number of limitations. Task Group 132 (TG-132)[26] presents techniques and workflows for image registration as well as a few common evaluation measures. In the present study, the large majority of registration evaluation measures reported in the literature have been on the same clinical dataset to evaluate the performance of three different registration algorithms given the fact that there has not been yet any measure established as reference or 'golden' standard. However, although the analysis provides a quantitative mean of evaluating the quality of registration and the measures used have been validated by other studies, none of the registrations were assessed by a radiation oncologist or a radiologist. Most studies that evaluate image registration algorithms employ only few evaluation measures and their conclusions are subject to their results. However, as it is shown here, there is a considerable variability in the results of the different evaluation measures even when evaluating exactly the same dataset and registration algorithms. On the other hand, intra- and inter-observer variability (in manual image registration or contour delineation) has been shown to be larger than that of the evaluation measures. Hence, the assessment of the performed registrations by a single physician could not be adequate for our purpose. Finally, 10 patients is a small cohort. Hence, the conclusions derived by the presented findings should be considered with caution in the light of the reduced statistical power of the analysis.

CONCLUSION

In this study, we have demonstrated the performance of various measures that may be used for the evaluation of rigid and deformable registration accuracy of a 4DCT dataset. The EXDMP method showed an overwhelming nonphysical and unrealistic behavior as well as poor image similarity in a number of studied cases, making DMP the method of choice. However, care must be taken when deciding which method should be used, because this also depends on the task for which it is applied, i.e., dose accumulation, contour propagation, or measuring local volume (or surface area) change. For example, better voxel mapping (EXDMP) may lead to more accurate contour propagation, etc. It was also shown that the evaluated measures might detect CT dataset differences with higher precision if the analysis is restricted to a smaller volume (i.e., differences were observed in the results of the measures depending on the size of the CT dataset being evaluated).

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  14 in total

1.  Image quality assessment: from error visibility to structural similarity.

Authors:  Zhou Wang; Alan Conrad Bovik; Hamid Rahim Sheikh; Eero P Simoncelli
Journal:  IEEE Trans Image Process       Date:  2004-04       Impact factor: 10.856

2.  Tracking the dose distribution in radiation therapy by accounting for variable anatomy.

Authors:  B Schaly; J A Kempe; G S Bauman; J J Battista; J Van Dyk
Journal:  Phys Med Biol       Date:  2004-03-07       Impact factor: 3.609

3.  Non-local shape descriptor: a new similarity metric for deformable multi-modal registration.

Authors:  Mattias P Heinrich; Mark Jenkinson; Manav Bhushan; Tahreema Matin; Fergus V Gleeson; J Michael Brady; Julia A Schnabel
Journal:  Med Image Comput Comput Assist Interv       Date:  2011

4.  Gradient Magnitude Similarity Deviation: A Highly Efficient Perceptual Image Quality Index.

Authors:  Wufeng Xue; Lei Zhang; Xuanqin Mou; Alan C Bovik
Journal:  IEEE Trans Image Process       Date:  2014-02       Impact factor: 10.856

5.  Objective assessment of deformable image registration in radiotherapy: a multi-institution study.

Authors:  Rojano Kashani; Martina Hub; James M Balter; Marc L Kessler; Lei Dong; Lifei Zhang; Lei Xing; Yaoqin Xie; David Hawkes; Julia A Schnabel; Jamie McClelland; Sarang Joshi; Quan Chen; Weiguo Lu
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

6.  FSIM: a feature similarity index for image quality assessment.

Authors:  Lin Zhang; Lei Zhang; Xuanqin Mou; David Zhang
Journal:  IEEE Trans Image Process       Date:  2011-01-31       Impact factor: 10.856

7.  Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No. 132.

Authors:  Kristy K Brock; Sasa Mutic; Todd R McNutt; Hua Li; Marc L Kessler
Journal:  Med Phys       Date:  2017-05-23       Impact factor: 4.071

8.  Investigation of four-dimensional (4D) Monte Carlo dose calculation in real-time tumor tracking stereotatic body radiotherapy for lung cancers.

Authors:  Mark K H Chan; Dora L W Kwong; Sherry C Y Ng; Eric K W Tam; Anthony S M Tong
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

9.  A Voxel-by-Voxel Comparison of Deformable Vector Fields Obtained by Three Deformable Image Registration Algorithms Applied to 4DCT Lung Studies.

Authors:  Mirek Fatyga; Nesrin Dogan; Elizabeth Weiss; William C Sleeman; Baoshe Zhang; William J Lehman; Jeffrey F Williamson; Krishni Wijesooriya; Gary E Christensen
Journal:  Front Oncol       Date:  2015-02-04       Impact factor: 6.244

10.  Evaluation of various deformable image registration algorithms for thoracic images.

Authors:  Noriyuki Kadoya; Yukio Fujita; Yoshiyuki Katsuta; Suguru Dobashi; Ken Takeda; Kazuma Kishi; Masaki Kubozono; Rei Umezawa; Toshiyuki Sugawara; Haruo Matsushita; Keiichi Jingu
Journal:  J Radiat Res       Date:  2013-07-17       Impact factor: 2.724

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.