NeoBeat offers rapid newborn heart rate assessment.

Jessica Brittany Bush1, Victoria Cooley2, Jeffrey Perlman2, Catherine Chang2.   

Abstract

BACKGROUND: Heart rate (HR) is used to guide interventions during delivery room (DR) neonatal resuscitation. Dry electrode ECG (NeoBeat) may detect HR more rapidly than pulse oximetry (PO) and portable ECG, but real-time comparisons of these devices are lacking. DESIGN/
METHODS: PO, ECG and NeoBeat were placed sequentially on newborns in the DR. Time for device placement and time to accurate HR acquisition were noted.
RESULTS: DR resuscitations of 28 preterm/term infants were observed. The NeoBeat was placed faster (ie, 3 s) than PO (20 s, p=<0.0001) and ECG (16 s, p=<0.0001). Total time from initiation of device placement to HR acquisition was fastest with NeoBeat (13 s) versus ECG (42 s, p<0.0001) and PO (105 s, p<0.0001) (duration values=median).
CONCLUSIONS: These observations in a small cohort of relatively well neonates demonstrate that the NeoBeat is significantly faster to place and consistently acquires HR faster than PO and ECG. © Author(s) (or their employer(s)) 2021. No commercial re-use. See rights and permissions. Published by BMJ.

Entities:  

Keywords:  neonatology; resuscitation

Year:  2021        PMID: 33478958     DOI: 10.1136/archdischild-2020-320901

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  5 in total

Review 1.  Toward greater nuance in delayed cord clamping.

Authors:  Laura Marrs; Susan Niermeyer
Journal:  Curr Opin Pediatr       Date:  2022-04-01       Impact factor: 2.856

2.  The Assisted Breathing before Cord Clamping (ABC) Study Protocol.

Authors:  Michael P Meyer; Elizabeth Nevill
Journal:  Children (Basel)       Date:  2021-04-26

3.  Comparing pulse rate measurement in newborns using conventional and dry-electrode ECG monitors.

Authors:  Eris van Twist; Hylke H Salverda; Arjan B Te Pas
Journal:  Acta Paediatr       Date:  2022-01-31       Impact factor: 4.056

4.  Perceptions and experiences of Congolese midwives implementing a low-cost battery-operated heart rate meter during newborn resuscitation.

Authors:  Madeline Thornton; Daniel Ishoso; Adrien Lokangaka; Sara Berkelhamer; Melissa Bauserman; Joar Eilevstjønn; Pooja Iyer; Beena D Kamath-Rayne; Eric Mafuta; Helge Myklebust; Janna Patterson; Antoinette Tshefu; Carl Bose; Jackie K Patterson
Journal:  Front Pediatr       Date:  2022-09-29       Impact factor: 3.569

5.  Comparison of Heart Rate Feedback from Dry-Electrode ECG, 3-Lead ECG, and Pulse Oximetry during Newborn Resuscitation.

Authors:  Siren Rettedal; Joar Eilevstjønn; Amalie Kibsgaard; Jan Terje Kvaløy; Hege Ersdal
Journal:  Children (Basel)       Date:  2021-11-26
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.