Literature DB >> 33469855

Development and characterization of an all-in-one gamma probe with auto-peak detection for sentinel lymph node biopsy based on NEMA NU3-2004 standard.

Aram Radnia1, Hamed Abdollahzadeh1, Behnoosh Teimourian1, Mohammad Hossein Farahani1, Mohammad Esmaeil Akbari2, Habib Zaidi3,4,5,6, Mohammad Reza Ay7,8.   

Abstract

BACKGROUND: A gamma probe is a handheld device used for intraoperative interventions following interstitial injection of a radiotracer to locate regional lymph nodes through the external detection of radiation. This work reports on the design and performance evaluation of a novel fully integrated gamma probe (GammaPen), recently developed by our group.
MATERIALS AND METHODS: GammaPen is an all-in-one pocket gamma probe with low weight and adequate dimensions, consisting of a detector, a control unit and output all together. The detector module consists of a cylindrical Thallium-activated Cesium Iodide [CsI (Tl)] crystal optically coupled to a Silicon photomultiplier (SiPM), shielded using Tungsten housing on side and back faces. The electronics of the probe consists of two small boards to handle signal processing and analog peak detection tasks. A number of parameters, including probe sensitivity in air/water, spatial resolution in air/water, angular resolution in air/water, and side and back shielding effectiveness, were measured to evaluate the performance of the probe based on NEMA NU3-2004 standards.
RESULTS: The sensitivity of the probe in air at distances of 10, 30, and 50 mm is 18784, 3500, and 1575 cps/MBq. The sensitivity in scattering medium was also measured at distances of 10, 30, and 50 mm as 17,680, 3050, and 1104 cps/MBq. The spatial and angular resolutions in scattering medium were 47 mm and 87 degree at 30 mm distance from the probe, while they were 40 mm and 77 degree in air. The detector shielding effectiveness and leakage sensitivity are 99.91% and 0.09%, respectively.
CONCLUSION: The performance characterization showed that GammaPen can be used effectively for sentinel lymph node localization. The probe was successfully used in several surgical interventions by an experienced surgeon confirming its suitability in a clinical setting.

Entities:  

Keywords:  All-in-one; Intra-operative gamma probe; NEMA NU3; Performance evaluation; Sentinel lymph node

Year:  2021        PMID: 33469855      PMCID: PMC7981324          DOI: 10.1007/s12149-021-01581-z

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  21 in total

1.  Sentinel lymph node evaluation in squamous cell carcinoma of the head and neck.

Authors:  S J Stoeckli; H Steinert; M Pfaltz; S Schmid
Journal:  Otolaryngol Head Neck Surg       Date:  2001-09       Impact factor: 3.497

2.  Evaluation of surgical gamma probes for radioguided sentinel node localisation.

Authors:  T Tiourina; B Arends; D Huysmans; H Rutten; B Lemaire; S Muller
Journal:  Eur J Nucl Med       Date:  1998-09

3.  Sentinel-node biopsy to avoid axillary dissection in breast cancer with clinically negative lymph-nodes.

Authors:  U Veronesi; G Paganelli; V Galimberti; G Viale; S Zurrida; M Bedoni; A Costa; C de Cicco; J G Geraghty; A Luini; V Sacchini; P Veronesi
Journal:  Lancet       Date:  1997-06-28       Impact factor: 79.321

4.  Determination of depth distribution of internal mammary lymph nodes on lateral lymphoscintigraphy.

Authors:  T C Yip; G N Ege
Journal:  Clin Radiol       Date:  1985-03       Impact factor: 2.350

5.  The role of elective lymph node dissection in the management of patients with thick cutaneous melanoma.

Authors:  N J Crowley; H F Seigler
Journal:  Cancer       Date:  1990-12-15       Impact factor: 6.860

6.  The three-dimensional localization of internal mammary lymph nodes by radionuclide lymphoscintigraphy.

Authors:  W D Kaplan; J W Andersen; R L Siddon; B T Connolly; C A McCormick; S M Laffin; E M Rosenbaum; C A Jennings; A Recht; J R Harris
Journal:  J Nucl Med       Date:  1988-04       Impact factor: 10.057

7.  Sentinel lymph node biopsy in breast cancer: initial experience at Memorial Sloan-Kettering Cancer Center.

Authors:  B J O'Hea; A D Hill; A M El-Shirbiny; S D Yeh; P P Rosen; D G Coit; P I Borgen; H S Cody
Journal:  J Am Coll Surg       Date:  1998-04       Impact factor: 6.113

8.  Detection of sentinel lymph nodes in minimally invasive surgery using indocyanine green and near-infrared fluorescence imaging for uterine and cervical malignancies.

Authors:  Elizabeth L Jewell; Juan Juan Huang; Nadeem R Abu-Rustum; Ginger J Gardner; Carol L Brown; Yukio Sonoda; Richard R Barakat; Douglas A Levine; Mario M Leitao
Journal:  Gynecol Oncol       Date:  2014-02-28       Impact factor: 5.482

9.  Prospective comparison of 3 gamma-probes for sentinel lymph node detection in 200 breast cancer patients.

Authors:  Jean-Marc Classe; Maryse Fiche; Caroline Rousseau; Christine Sagan; François Dravet; Raphaëlle Pioud; Albert Lisbona; Ludovic Ferrer; Loic Campion; Isabelle Resche; Chantal Curtet
Journal:  J Nucl Med       Date:  2005-03       Impact factor: 10.057

10.  Mammary lymphoscintigraphy in breast cancer.

Authors:  R F Uren; R B Howman-Giles; J F Thompson; D Malouf; G Ramsey-Stewart; F W Niesche; S B Renwick
Journal:  J Nucl Med       Date:  1995-10       Impact factor: 10.057

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.