Abstract

PURPOSE: This study analyzes the efficiency of different vitrectomy systems and compares single with double-bladed cutters.
METHODS: The systems EVA™ (DORC), Constellation® Vision System (ALCON), megaTRON S4HPS (Geuder) and Stellaris® PC (Bausch and Lomb) were used. We chose 20G and 23G probes, since not all systems had switched to a smaller G at the time the study was conducted in 2016. Cut rates were varied in increments of 1000 cuts/min from 500 cpm to the system's maximum and vacuum pressures were varied in increments of 100 mmHg, from 100 to 600 mmHg up to the individual system's maximum. In this study water, egg white, Pluronic®-F127 gel and isolated porcine vitreous were used as models of human vitreous. The vitrectomy efficiency was calculated from the aspirated mass (g) within 30 s. The aperture of the different vitrectomy probes was filmed with a high-speed camera.
RESULTS: The area under the curve analysis showed differences in efficiency between vitrectomy systems. For water, a reverse relationship between the aspirated mass and cut rate was shown. By contrast, for most systems aspirated egg white and porcine vitreous showed a non-linear increase or decrease for 4000 cpm and above. For all vitreous surrogates, EVA™'s double-bladed probe aspirated significantly (p < 0.001) more vitreous than its mono-bladed probe. Video recordings showed less vitreous traction for double- in contrast to single-bladed probes.
CONCLUSION: We can demonstrate differences in the efficiency of vitrectomy depending on the vitrectomy system used. Double-bladed probes were more efficient and probably safer than single-bladed probes.

Entities:  

Keywords:  20 gauge; 23 gauge; Vitrectomy; Vitreous; Vitreous model

Year:  2021        PMID: 33468260     DOI: 10.1186/s40942-020-00277-2

Source DB:  PubMed          Journal:  Int J Retina Vitreous        ISSN: 2056-9920


  3 in total

1.  Experiences with partial vitrectomy in patients with proliferative diabetic retinopathy.

Authors:  W A Van Heuven
Journal:  Mod Probl Ophthalmol       Date:  1972

2.  Vitrectomy: a pars plana approach.

Authors:  R Machemer; H Buettner; E W Norton; J M Parel
Journal:  Trans Am Acad Ophthalmol Otolaryngol       Date:  1971 Jul-Aug

Review 3.  Review of Small Gauge Vitrectomy: Progress and Innovations.

Authors:  Shaheeda Mohamed; Carl Claes; Chi Wai Tsang
Journal:  J Ophthalmol       Date:  2017-05-10       Impact factor: 1.909

  3 in total
  1 in total

1.  Human Vitreous Collagen Fragments Dimension As a Function of Vitrectomy Cut Rate.

Authors:  Tommaso Rossi; Andrea Speciale; Paola Menichini; Alberto Izzotti; Isabella D'Agostino; Carlandrea Trillo; Serena Telani; Giorgio Querzoli; Guido Ripandelli
Journal:  Transl Vis Sci Technol       Date:  2022-03-02       Impact factor: 3.283

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.