Field output factors for small fields: A large multicentre study.

S Dufreneix1, J Bellec2, S Josset3, L Vieillevigne4.   

Abstract

PURPOSE: The determination of output factors in small field dosimetry is a crucial point, especially when implementing stereotactic radiotherapy (SRT). Herein, a working group of the French medical physicist society (SFPM) was created to collect small field output factors. The objective was to gather and disseminate information on small field output factors based on different detectors for various clinical SRT equipment and measurement configurations.
METHOD: Participants were surveyed for information about their SRT equipment, including the type of linear particle accelerator (linac), collimator settings, measurement conditions for the output factors and the detectors used. Participants had to report both the ratio of detector readings and the correction factors applied as described in the IAEA TRS-483 code of practice for nominal field sizes smaller or equal to 3 cm. Mean field output factors and their associated standard deviations were calculated when data from at least 3 linacs were available.
RESULTS: 23 centres were enrolled in the project. Standard deviations of the mean field output factors were systematically smaller than 1.5% for field sizes larger or equal to 1 cm and reached 5% for the smallest field size (0.5 cm). Deviations with published data were smaller than 2% except for the 0.5 cm circular fixed aperture collimator of the CyberKnife where it reached 3.5%.
CONCLUSION: These field output factor values obtained via a large multicentre study can be considered as an external cross verification for any radiotherapy centre starting a SRT program and should help minimize systematic errors when determining small field output factors.
Copyright © 2021 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  IAEA TRS-483; Output factors; Small field dosimetry

Mesh:

Year:  2021        PMID: 33465756     DOI: 10.1016/j.ejmp.2021.01.001

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Small-field dosimetry with detector-specific output correction factor for single-isocenter stereotactic radiotherapy of single and multiple brain metastases.

Authors:  Tomohiro Ono; Kohei Kawata; Mitsuhiro Nakamura; Megumi Uto; Takashi Mizowaki
Journal:  Radiol Phys Technol       Date:  2022-10-22

2.  A multi-institutional evaluation of small field output factor determination following the recommendations of IAEA/AAPM TRS-483.

Authors:  Wolfgang Lechner; Rodolfo Alfonso; Mehenna Arib; M Saiful Huq; Anas Ismail; Rajesh Kinhikar; José M Lárraga-Gutiérrez; Karthick Raj Mani; Nkosingiphile Maphumulo; Otto A Sauer; Shaima Shoeir; Sivalee Suriyapee; Karen Christaki
Journal:  Med Phys       Date:  2022-07-08       Impact factor: 4.506