| Literature DB >> 33461286 |
Yoshinori Komazawa1, Mika Yuki1, Nobuhiko Fukuba2, Yoshiya Kobayashi2, Hitomi Ishitobi2, Sayaka Nakashima1, Makoto Nagaoka2, Yoshiko Takahashi2, Toshihiro Shizuku2.
Abstract
BACKGROUND/AIMS: Unsedated transnasal esophagogastroduodenoscopy (EGD) is affected by a poor scope lens-cleaning function. We have previously reported good, albeit limited, effects of an oolong tea washing solution; here, we evaluated the effectiveness of a 5% lens cleaning solution for cleaning an EGD lens.Entities:
Keywords: Lens cleaner; Small-caliber endoscope; Transnasal esophagogastroduodenoscopy; Visual quality
Year: 2021 PMID: 33461286 PMCID: PMC8039737 DOI: 10.5946/ce.2020.062
Source DB: PubMed Journal: Clin Endosc ISSN: 2234-2400
Fig. 1.Photographic images of two different test charts observed through the lens of a (GIF-XP 260N; Olympus, Tokyo, Japan) endoscope soiled by pork grease and then subjected to cleaning. (A) Level 5 cleanliness. (B) Level 3 cleanliness. (C) Level 1 cleanliness.
Fig. 2.The effect of various cleaning solutions on pork grease-soiled lens. The effect of the 5% Cleash solution (C) was consistently superior to that of the 5% dimethicone solution (D) and distilled water (W). Furthermore, visual quality remained high with (C) even with repeated soiling and cleaning procedures.
Study II Patient Characteristics
| C-group | W-group | |
|---|---|---|
| Mean age, yr | 54.1±13.6 | 53.5±12.0 |
| Male : Female | 222 : 276 | 244 : 254 |
| Duodenum observation rate (%) | 100 | 100 |
| Endoscopic procedures | ||
| Biopsy, | 19 (3.8) | 18 (3.6) |
| Chromoendoscopy, | 57 (11.4) | 60 (12.0) |
Fig. 3.Solution volume used for lens cleaning during esophagogastroduodenoscopy procedures. The one used in the C-group was significantly smaller as compared to the W-group (p<0.05).
Fig. 4.Endoscopist stress level due to lens soiling during esophagogastroduodenoscopy procedures. The stress level from the C-group was lower compared to the W-group.
Fig. 5.The level of lens cleanliness during esophagogastroduodenoscopy. The level found in the C-group was significantly higher compared to the W-group.