Practical considerations for establishing dead-time corrections in quantitative SPECT imaging.

Claire Cohalan1, Marc-André Morin, Antoine Leblond.   

Abstract

Quantitative SPECT studies require specific information about the equipment being used. Particularly in the context of therapeutic studies, the effect of dead-time can be significant and must be quantified. We explored different techniques for measuring the dead-time constant and applying dead-time corrections to the data.
METHOD: The dead-time constant was measured on four similar SPECT/CT systems by following the response of the system to a uniform phantom initially containing 17 GBq of Lu-177 over a period of 23 days. It was then calculated using the two-source method with 1 332 MBq of Tc-99 m. The dead-time constant found was used to correct SPECT/CT phantom images either applying the correction by projection or globally on the image.
RESULTS: Both methods of calculating the dead-time constant produced equivalent results. However, the dead-time constant varied by as much as 8% between machines of the same model and manufacturer. Correcting for dead-time by projection rather than globally produced slightly more precise results (0.94% error rather than 2.59% error). The benefit of this correction technique will be dependent on the level of asymmetry in the patient as well as the magnitude of the dead-time correction effect.
CONCLUSION: quantification of the dead-time of a system can be performed quickly using the two-source method and any radioisotope. However, it is important to perform this measurement on every system being used. In vastly asymmetric images with high dead-time correction, correcting for dead-time by projection can be pertinent, increasing the precision of dosimetry calculations by several percent. However this additional gain may be within the error of SUV measurements for many clinical acquisitions.

Entities:  

Year:  2020        PMID: 33438643     DOI: 10.1088/2057-1976/ab7500

Source DB:  PubMed          Journal:  Biomed Phys Eng Express        ISSN: 2057-1976


  1 in total

1.  Impact of the dead-time correction method on quantitative 177Lu-SPECT (QSPECT) and dosimetry during radiopharmaceutical therapy.

Authors:  Alessandro Desy; Guillaume F Bouvet; Nancy Lafrenière; Atefeh Zamanian; Philippe Després; Jean-Mathieu Beauregard
Journal:  EJNMMI Phys       Date:  2022-08-17
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.