Feasibility of microscope-integrated swept-source optical coherence tomography in canaloplasty.

Huan Xu1,2, Wangyi Fang1,2, Guangxing Liu3, Jinyu Fan3, Jian Yu1,2, Yuan Zong1,2, Chunhui Jiang1,2, Guohua Shi3,4, Xinghuai Sun1,2.   

Abstract

BACKGROUND: Several researchers have used commercial microscope-integrated optical coherence tomography (OCT) systems in glaucoma surgery, including ab interno trabeculectomy and canaloplasty. However, the 840 nm wavelength light source of the OCT systems is not ideal for imaging the anterior chamber angle structures because of its limited penetration. We evaluated the potential value of a microscope-integrated swept-source OCT system with a 1,310 nm center-wavelength light in canaloplasty for glaucoma.
METHODS: Sixteen porcine eyes were used to simulate canaloplasty. The critical surgical steps were monitored using a prototype microscope-integrated OCT system with a 1,310 nm light source and a high axial scan rate of 100 kHz. The images from swept-source OCT and three-dimensional images from the microscope were projected simultaneously onto a liquid crystal display three-dimensional monitor (LMD-4251TD, Sony, Japan). The changes in the collector vessel (aqueous drainage structure in the porcine eye, similar to Schlemm's canal in humans) were measured using Image J software. Histological sections stained with hematoxylin and eosin were used to assess surgical efficacy.
RESULTS: High-resolution real-time images of the anterior segment were acquired during canaloplasty using the microscope-integrated OCT system. With the real-time OCT images, the position of the collector vessel was identified and the scleral flap could be created at the ideal location. The expansion of the collector vessel after viscoelastic injection was also visualized in real time. Compared with baseline, there was a significant increase in the cross-sectional area (from 14,502.98±9,242.55 to 59,499.04±20,506.41 µm2, P<0.001) of the collector vessel.
CONCLUSIONS: Using the microscope-integrated OCT system, real-time images of the anterior segment were successfully acquired during the operation. The microscope-integrated OCT system might be useful in future anti-glaucoma surgery. 2020 Annals of Translational Medicine. All rights reserved.

Entities:  

Keywords:  Anterior segment; MIOCT; canaloplasty; intraoperative real-time imaging

Year:  2020        PMID: 33437776      PMCID: PMC7791249          DOI: 10.21037/atm-20-3469

Source DB:  PubMed          Journal:  Ann Transl Med        ISSN: 2305-5839


Introduction

Glaucoma is a leading cause of irreversible blindness worldwide (1). Despite the development of antiglaucoma medications, surgery remains the mainstay of treatment (2). Microinvasive antiglaucoma surgery has recently gained favor among ophthalmologists, and viscoelastic canaloplasty is the preferred technique. Introduced recently, viscoelastic canaloplasty is a modern form of non-penetrating glaucoma surgery (3-5). Multiple follow-up studies have revealed that this surgery may achieve sustained, long-term reductions in intraocular pressure (IOP) in eyes with primary open-angle glaucoma (4,6-10). Compared with traditional methods, canaloplasty lowers IOP with minimal complications or iatrogenic damage (1,11). However, this procedure is technically demanding, with a long learning curve, especially in terms of locating Schlemm’s canal (12). On the other hand, optical coherence tomography (OCT) can provide high-resolution cross-sectional images of the eye without requiring contact. In 2009, Dayani et al. reported the application of handheld OCT in ocular surgery for the first time (13). Since then, OCT has been used to guide various forms of ocular surgery (14-17). With further optimization of the technique, a breakthrough was achieved in the form of real-time imaging during vitreoretinal surgery through microscope-integrated OCT (MIOCT), first reported by Binder et al. (18). Subsequently, several researchers have used commercial MIOCT systems (iOCT, OptoMedical Technologies, Luebeck, Germany; Rescan 700, Carl Zeiss Meditec, Jena, Germany) in glaucoma surgery, including ab interno trabeculectomy and canaloplasty (19-21). The systems include a built-in heads-up display, which provides real-time intraoperative OCT feedback without interrupting the surgery. However, the 840 nm wavelength light source of the OCT systems is not ideal for imaging the structures of the anterior chamber angle because its penetration is relatively limited. In the present study, we used a prototype MIOCT system with a 1,310 nm center-wavelength swept-source light, which achieves deep penetration and real-time imaging with a high signal-to-noise ratio (22). Using this system, a three-dimensional (3D) surgical image from the microscope and the OCT image of the tissue level could be projected together onto the same screen. Here, we simulated surgery in porcine eyes to investigate the potential value of our MIOCT system in canaloplasty. We present the following article in accordance with the ARRIVE reporting checklist (available at http://dx.doi.org/10.21037/atm-20-3469).

Methods

The study was approved by the Institutional Review Board of the Eye and ENT Hospital of Fudan University, Shanghai, China (No. 2019031-1). Eight male Yorkshire pigs (10 kg, 2 months old, clean grade) were purchased from Jiagan Biotechnology Co., Ltd. and maintained in a 12 h light/dark cycle with free access to food and water. After the Yorkshire pigs were acclimatized for at least 7 days, they were used in accordance with the Association for Research in Vision and Ophthalmology Statement for the Use of Animals in Ophthalmic and Vision Research. Canaloplasty was performed on the sixteen eyes of eight Yorkshire pigs. The parameters before and after the operation were acquired and analyzed as self-control.

MIOCT system

In the MIOCT prototype, swept-source OCT (SS-OCT) was optically designed for use in conjunction with the optical path of a microscope. The microscope and OCT shared the same objective lens, with a focal length of 200 mm. SS-OCT used the 1,310 nm wavelength range at a high axial scan rate of 100 kHz and a high-speed tunable laser (HSL-20, Santec, Japan). Its lateral resolution was 14 µm, and its lateral range was 10.7 mm with 768 A-Scans/B-Scan. The axial resolution was 11 µm, and the maximum depth in air was 6 mm. After detection, the OCT signals were processed with a computer (CPU: E5-2620v4@2.1 GHz; RAM: 32 GB; graphics cards: GTX1080Ti) to output real-time OCT images. A superluminescent diode with a wavelength of 640 nm was combined in the sample OCT arm and used to mark the OCT imaging regions. The binocular microscopic images were separated using a beam splitter. They were recorded with two surgical video cameras (MCC-500MDC, Sony, Japan) and displayed on a high-definition (1080P; 1,920×1,080) 42-inch LCD 3D monitor (LMD-4251TD, Sony). Real-time OCT images of the same region of interest are displayed in the lower right corner of the same screen. The surgeon wore circular polarizing 3D glasses to allow precise depth perception (Figure S1). The final time delays in the acquisition of the OCT image and dual-channel microscopic image were less than 120 and 100 ms, respectively.

Surgical protocol

Surgery was conducted on 16 porcine eyes from eight live male Yorkshire pigs (10 kg, 2 months old). After general anesthesia, the eyelids were opened. The real-time OCT images of () the porcine drainage system was acquired. The porcine drainage system at the anterior chamber angle has a series of collector vessels or the angular aqueous plexus, which is similar to the Schlemm’s canal in primates. Then, the conjunctiva and Tennon’s membrane were removed. After the collector vessel was located with the MIOCT system, a 5×5 mm2 limbal-based one-half thickness scleral flap was created. The scalpel was stopped when the flap exceeded the position of the collector vessel (). After that, an incision perpendicular to the limbus was made across the scleral bed and the collector vessel; care was taken not to penetrate the inner wall of the collector vessel. The opening of the collector vessel was then expanded with a custom-made trabeculotome (Figure S2) with the tip adapted from an optical fiber (Φ=250 µm) (). An optical fiber microcatheter (Φ=200 µm) (iTrack 250, Ellex iScience, Inc., Freemont, CA, USA) connected to a red flickering laser light was then used to identify the location of the fiber tip, which had been threaded into the collector vessel. When the distal end of the microcatheter reached the terminal site, the viscoelastic (DisCoVisc, Alcon Laboratories, Fort Worth, USA) was injected via a screw-driven syringe connected to the microcatheter. This screw-driven syringe was twisted again as the microcatheter was withdrawn in a circle with an angle of 30 degrees (). According to the 3Rs principle, all the animals were euthanized at the end of the experiment by intracardial injection of approximately 20 mL of 15% potassium chloride solution. Their eyes were then rapidly enucleated for histopathological evaluation.
Figure 1

Surgical procedure of canaloplasty and its corresponding real-time images from the microscope-integrated swept-source optical coherence tomography (MIOCT) system. (A) MIOCT image before surgery; (B) cross-sectional OCT image of the red scanning line (white arrow) clearly shows the anterior segmental structures (AC: anterior chamber; Co: cornea; L: lens; I: iris; S: sclera) and collector vessel (yellow arrow); (C) MIOCT image after incision of a superficial scleral flap; (D) OCT image shows the location of the collector vessel and scleral flap (SF); (E) MIOCT image after insertion of a custom-made trabeculotome with an optical fiber tip (white arrowhead); (F) expansion of the collector vessel was observed on the OCT image; (G) MIOCT image after insertion of a flexible microcatheter (white asterisk) and viscoelastic was injected; (H) OCT image confirming expansion of the collector vessel.

Surgical procedure of canaloplasty and its corresponding real-time images from the microscope-integrated swept-source optical coherence tomography (MIOCT) system. (A) MIOCT image before surgery; (B) cross-sectional OCT image of the red scanning line (white arrow) clearly shows the anterior segmental structures (AC: anterior chamber; Co: cornea; L: lens; I: iris; S: sclera) and collector vessel (yellow arrow); (C) MIOCT image after incision of a superficial scleral flap; (D) OCT image shows the location of the collector vessel and scleral flap (SF); (E) MIOCT image after insertion of a custom-made trabeculotome with an optical fiber tip (white arrowhead); (F) expansion of the collector vessel was observed on the OCT image; (G) MIOCT image after insertion of a flexible microcatheter (white asterisk) and viscoelastic was injected; (H) OCT image confirming expansion of the collector vessel.

MIOCT imaging and measurement

Intraoperative MIOCT images were acquired at the beginning of surgery and after viscoelastic injection for further analysis. The scanning area could be controlled by the surgeon by moving the microscope. In the cross-sectional OCT images, the collector vessel appeared as a dark slit and oval structure. After the surgery, the collector vessels in each of the acquired image were outlined manually by two surgeons (XH and FW) and their areas were automatically quantified using Image J software (http://imagej.nih.gov/ij/; National Institutes of Health, Bethesda, MD, USA). Two surgeons, who were blinded to all sample identities, performed the measurements separately. The average values were used for subsequent analyses.

Statistical analyses

All statistical analyses were conducted using SPSS version 20.0 (SPSS, Inc., Chicago, IL, USA). All variables were checked for conformation to a normal distribution with the one-sample Kolmogorov-Smirnov test. The cross-sectional area of the collector vessel before and after surgery was compared using paired-sample t-tests.

Results

Surgery was conducted on 16 porcine eyes from eight healthy Yorkshire pigs. High-resolution real-time images of the anterior segment of the eyes were acquired intraoperatively with the MIOCT system. The collector vessels were located according to the OCT images. The collector vessel lumina had a low signal, whereas the walls had a relatively highly reflective signal. With the guidance of the images, the scleral flap was created until its tip exceeded the collector vessel. With the collector vessels visible, the incision was made across the scleral bed. In one eye, the scalpel unexpectedly penetrated the anterior chamber while making the incision and this eye was excluded. A flexible microcatheter was then threaded into the collector vessels. After viscoelastic injection, the collector vessel lumen was enlarged in OCT images of all 15 porcine eyes (). The results demonstrated that there was significant increase, compared with baseline, in the area (from 14,502.98±9,242.55 to 59,499.04±20,506.41 µm2, P<0.001) of the collector vessels (). For the measurement of collector vessel area, the intraclass correlation coefficient (ICC) values indicated good consistency between the two readers (Table S1). Histopathological assessment confirmed the expansion of the collector vessels ().
Figure 2

Changes in the lumen of the collector vessel before (A) and after (B) viscoelastic injection during canaloplasty. White arrowheads show the location of collector vessel.

Figure 3

Quantitative evaluation of the cross-sectional area before and after canaloplasty. **, P<0.01, tested with paired-sample t-tests.

Figure 4

Hematoxylin and eosin staining of histological sections obtained before (A) and after (B) canaloplasty (20×).

Changes in the lumen of the collector vessel before (A) and after (B) viscoelastic injection during canaloplasty. White arrowheads show the location of collector vessel. Quantitative evaluation of the cross-sectional area before and after canaloplasty. **, P<0.01, tested with paired-sample t-tests. Hematoxylin and eosin staining of histological sections obtained before (A) and after (B) canaloplasty (20×).

Discussion

This study demonstrates the feasibility of using MIOCT with a 1,310 nm wavelength swept-source light source in canaloplasty. Real-time high-resolution images of the anterior segmental structures were acquired during the procedure. These images provided real-time information for surgeons to make decisions during the procedure and also confirmed the expansion of the collector vessel. Precise and minimally invasive approaches have become popular in modern surgery, especially in ophthalmology. High-resolution, intraoperative real-time imaging is one way to ensure that these surgeries are successful (2,23). In recent years, roentgenograms and B-mode ultrasound have been used and yielded precise and predictable results in prior studies (24-27). In ophthalmology, OCT has been used for several years, and has greatly increased our understanding of many ocular structures. Swept-source OCT was recently introduced and is capable of acquiring a B-scan in 10 ms, which allows intraoperative real-time imaging. With the guidance of MIOCT system intraoperatively, the present study showed that high-resolution real-time images of the anterior segment of the eye during canaloplasty could be acquired. Previous studies have reported that the success rate of cannulation was about 74–89.9% (3,4,28,29). Furthermore, in glaucomatous eyes, anatomical anomalies in Schlemm’s canal, arising from pathological changes or previous surgeries, can make cannulation even more challenging (6,8,10), especially for beginners. In our study, to test the feasibility and potential value of the MIOCT system in guiding canaloplasty, all steps in the surgical procedure were simulated. With OCT guidance, canaloplasty was performed successfully by nonglaucoma specialists. During surgery, the collector vessel was located with real-time OCT imaging, achieved with the MIOCT system. Moreover, a scleral flap was precisely created above the collector vessel. With these collector vessels visible, a vertical incision was made at an appropriate depth. Unlike the routine approach, a deep scleral flap was not required (29). This makes the procedure much easier to perform and reduces potential complications. Previously, canal expansion was only verified postoperatively by ultrasound biomicroscopy (4,9). In this study, the placement of the microcatheter and the expansion of the collector vessel were confirmed in real time with MIOCT, which adds to the confidence of the surgeon while injecting the viscoelastic. Formerly, Siebelmann et al. and Pasricha used intraoperative OCT for the visualization or monitoring of anti-glaucoma surgery, and both found it to be helpful and valuable. However, in their studies and other reports, it was noticed that the manner in which OCT images were projected needed improvement (20,30) and it was suggested that the SS-OCT with a deeper penetration and high imaging speed might have great potential (20,30,31). Based on these findings, our system used SS-OCT with a 1,310 nm light source, and the high-resolution images acquired by the OCT and the microscope were projected on the screen side by side. The results indicated that the MIOCT system was helpful in monitoring and guiding canaloplasty. In addition, the problem of the scleral tissue shadow noticed in the former report using an 840 nm OCT device (20) did not affect the surgeons in our study. Although the collector vessels in the porcine eye differ from Schlemm’s canal in the human eye (32), all of the steps involved in canaloplasty were simulated in this study. Therefore, our results provided evidence that the novel MIOCT system might have a potential value for canaloplasty. The possible correlation between the intraoperative findings (the degree of canal expansion) and the postoperative prognosis (the attenuation of IOP) will be investigated in the future. With this system, combined 3D microscope and OCT images were viewed on the same screen. Although this permitted high-resolution real-time OCT image guidance, the 3D system was complex and beyond the typical imaging routine of most surgeons. Therefore, surgeons will require additional training with their use. Furthermore, although the beacon light clearly indicates the OCT imaging site, when the eye moves during surgical manipulation, its position must be readjusted to the region of interest, as previously noted by Hahn et al. (6,33). Therefore, an automatic tracking system that focuses the OCT scan beam on the preferred target would be extremely valuable and should be developed in the future. Using an animal model of canaloplasty, our study demonstrated that the MIOCT system could acquire real-time, intraoperative high-resolution images of the anterior segment, and the acquired images could be helpful to surgeons. Future clinical trials will continue to demonstrate how the MIOCT system can be used in real practice. The article’s supplementary files as
  33 in total

1.  Feasibility of intrasurgical spectral-domain optical coherence tomography.

Authors:  Susanne Binder; Christiane I Falkner-Radler; Christoph Hauger; Holger Matz; Carl Glittenberg
Journal:  Retina       Date:  2011 Jul-Aug       Impact factor: 4.256

2.  Doppler ultrasound-guided percutaneous nephrolithotomy: a prospective randomized study.

Authors:  Bing-Chang Tzeng; Chung-Jing Wang; Shi-Wei Huang; Chien-Hsing Chang
Journal:  Urology       Date:  2011-02-12       Impact factor: 2.649

3.  Canaloplasty in Open-angle Glaucoma: Mid-term Results From a Multicenter Study.

Authors:  Paolo Brusini; Guido Caramello; Stefano Benedetti; Claudia Tosoni
Journal:  J Glaucoma       Date:  2016-05       Impact factor: 2.503

4.  Review of intraoperative optical coherence tomography: technology and applications [Invited].

Authors:  Oscar M Carrasco-Zevallos; Christian Viehland; Brenton Keller; Mark Draelos; Anthony N Kuo; Cynthia A Toth; Joseph A Izatt
Journal:  Biomed Opt Express       Date:  2017-02-21       Impact factor: 3.732

5.  Three-year canaloplasty outcomes for the treatment of open-angle glaucoma: European study results.

Authors:  Holger Bull; Kurt von Wolff; Norbert Körber; Manfred Tetz
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-07-06       Impact factor: 3.117

Review 6.  Mechanism of the reconstruction of aqueous outflow drainage.

Authors:  Chen Xin; Ning Tian; Meng Li; Huaizhou Wang; Ningli Wang
Journal:  Sci China Life Sci       Date:  2017-12-22       Impact factor: 6.038

7.  Four-dimensional microscope- integrated optical coherence tomography to enhance visualization in glaucoma surgeries.

Authors:  Neel Dave Pasricha; Paramjit Kaur Bhullar; Christine Shieh; Christian Viehland; Oscar Mijail Carrasco-Zevallos; Brenton Keller; Joseph Adam Izatt; Cynthia Ann Toth; Pratap Challa; Anthony Nanlin Kuo
Journal:  Indian J Ophthalmol       Date:  2017-01       Impact factor: 1.848

8.  Nonoperative dilatation of coronary-artery stenosis: percutaneous transluminal coronary angioplasty.

Authors:  A R Grüntzig; A Senning; W E Siegenthaler
Journal:  N Engl J Med       Date:  1979-07-12       Impact factor: 91.245

Review 9.  Intra-operative optical coherence tomography in glaucoma surgery-a systematic review.

Authors:  Bryan C H Ang; Sheng Yang Lim; Syril Dorairaj
Journal:  Eye (Lond)       Date:  2019-11-26       Impact factor: 3.775

10.  Intrasurgical Human Retinal Imaging With Manual Instrument Tracking Using a Microscope-Integrated Spectral-Domain Optical Coherence Tomography Device.

Authors:  Paul Hahn; Oscar Carrasco-Zevallos; David Cunefare; Justin Migacz; Sina Farsiu; Joseph A Izatt; Cynthia A Toth
Journal:  Transl Vis Sci Technol       Date:  2015-07-01       Impact factor: 3.283

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.