The aim of the study was to describe the Mycobacterium chimaera contamination in heater-cooler devices after the application of a protocol of cleaning and disinfection in a tertiary hospital. It was an observational study at the La Paz-Cantoblanco-Carlos III University Hospital, Madrid, Spain. Seven heater-cooler devices are used in our hospital: five 3T Sorin (LivaNova) and two Maquet. We followed the manufacturers' instructions for cleaning and disinfection of the different heater-cooler devices. Environmental testing was developed monthly from January 2017 to July 2019. Samples were obtained from both cardioplegia and patient circuits and before and after the disinfection process and were cultured in appropriate media for non-tuberculous mycobacteria and heterotrophic bacteria (coliforms and Pseudomonas aeruginosa). A total of 320 samples were taken. Mycobacterium chimaera grew in four water samples (1.25%) from three different heater-cooler devices, with two positive results occurring after disinfection. The heterotrophic bacteria Delftia acidovorans and Stenotrophomonas maltophilia were also found. There has not been a case of M. chimaera infection in patients after cardiac surgery in our hospital. In March 2019, we decided to move the heater-cooler device outside the operating room. Mycobacterium chimaera contamination is not always eradicated by disinfection processes. We believe that placing 3T heater-cooler devices outside the operating room is the best option in preventing M. chimaera infection during cardiac surgery.Entities:
Keywords: Aerosolization; Mycobacterium chimaera; cardiac surgery; disinfection; heater–cooler device
Year: 2020 PMID: 33437492 PMCID: PMC7786110 DOI: 10.1016/j.nmni.2020.100757
Source DB: PubMed Journal: New Microbes New Infect ISSN: 2052-2975
Presence of Mycobacterium chimaera in water samples
| HCD | Sample date | |||
|---|---|---|---|---|
| December 2017 | February 2018 | November 2018 | March 2019 | |
| Sorin 3T no. 1 | Positive for | |||
| Removed January 2018 | ||||
| Sorin 3T no. 2 | Positive for | |||
| Removed March 2018 | ||||
| Sorin 3T no. 3 | Positive for | Positive for | ||
| Removed March 2019 | ||||
| Sorin 3T no. 4 | All samples negative | |||
| Removed October 2018 | ||||
| Sorin 3T no. 5 | All samples negative | |||
| Currently active device since October 2018 | ||||
| Maquet no. 1 | All samples negative | |||
| Broken January 2018 | ||||
| Maquet no. 2 | All samples negative | |||
| Currently active device since May 2019 | ||||
Abbreviation: HCD, heater–cooler device.
Fig. 13T heater–cooler device outside the operating room.
Fig. 2Tubes passing through hole in the wall, inside the operating room.