Literature DB >> 33431965

Steerable needles for radio-frequency ablation in cirrhotic livers.

Nick J van de Berg1,2, Frédérique C Meeuwsen3, Michail Doukas3, Gernot Kronreif4, Adriaan Moelker5, John J van den Dobbelsteen6.   

Abstract

Accurate needle placement in deep-seated liver tumours can be difficult. In this work, we disclose two new manually controlled steerable needles for 17G radio-frequency ablation probe placement. The needles contain stylets with embedded compliant joints for active tip articulations, and concentric tubes for (curved-path) guidance. Needle steering was evaluated sequentially by intended users and in intended-use tissue types. Six interventional radiologists evaluated the needle in repeated ultrasound-guided steering tasks in liver-mimicking phantoms. Targets were located at a 100 mm depth and 20 mm lateral offset from the initial insertion line. The resulting mean absolute tip placement error was 1.0 ± 1.0 mm. Subsequently, steering-induced tissue damage was evaluated in fresh cirrhotic human liver explants. The surface area of puncture holes was estimated in scanned histology slides, using a connected-components analysis. The mean surface area was 0.26 ± 0.16 mm2 after steering with a median radius of curvature of 0.7 × 103 mm, versus 0.35 ± 0.15 mm2 after straight-path insertions with the steerable needle and 0.15 ± 0.09 mm2 after straight-path RFA probe insertions. The steering mechanisms proposed enable clinically relevant path corrections for 17G needles. Radiologists were quickly adept in curved-path RFA probe placement and the evaluation of histological tissue damage demonstrated a potentially safe use during liver interventions.

Entities:  

Mesh:

Year:  2021        PMID: 33431965      PMCID: PMC7801671          DOI: 10.1038/s41598-020-77869-3

Source DB:  PubMed          Journal:  Sci Rep        ISSN: 2045-2322            Impact factor:   4.379


  20 in total

Review 1.  A systematic review of the quality of liver biopsy specimens.

Authors:  Evangelos Cholongitas; Marco Senzolo; Richard Standish; Laura Marelli; Alberto Quaglia; David Patch; Amar P Dhillon; Andrew Kenneth Burroughs
Journal:  Am J Clin Pathol       Date:  2006-05       Impact factor: 2.493

2.  Histological evaluation of tissue damage caused by rotational needle insertion.

Authors:  Ryosuke Tsumura; Yusuke Takishita; Yuta Fukushima; Hiroyasu Iwata
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2016-08

3.  Highly Articulated Robotic Needle Achieves Distributed Ablation of Liver Tissue.

Authors:  Giada Gerboni; Joseph D Greer; Paul F Laeseke; Gloria L Hwang; Allison M Okamura
Journal:  IEEE Robot Autom Lett       Date:  2017-02-14

4.  A Methodical Quantification of Needle Visibility and Echogenicity in Ultrasound Images.

Authors:  Nick J van de Berg; Juan A Sánchez-Margallo; Arjan P van Dijke; Thomas Langø; John J van den Dobbelsteen
Journal:  Ultrasound Med Biol       Date:  2019-01-14       Impact factor: 2.998

5.  Ovipositor-inspired steerable needle: design and preliminary experimental evaluation.

Authors:  M Scali; T P Pusch; P Breedveld; D Dodou
Journal:  Bioinspir Biomim       Date:  2017-12-04       Impact factor: 2.956

Review 6.  Liver cancer: Approaching a personalized care.

Authors:  Jordi Bruix; Kwang-Hyub Han; Gregory Gores; Josep Maria Llovet; Vincenzo Mazzaferro
Journal:  J Hepatol       Date:  2015-04       Impact factor: 25.083

Review 7.  Radiofrequency ablation of liver tumors.

Authors:  Shaunagh McDermott; Debra A Gervais
Journal:  Semin Intervent Radiol       Date:  2013-03       Impact factor: 1.513

8.  Cancer statistics, 2016.

Authors:  Rebecca L Siegel; Kimberly D Miller; Ahmedin Jemal
Journal:  CA Cancer J Clin       Date:  2016-01-07       Impact factor: 508.702

9.  Needle placement errors: do we need steerable needles in interventional radiology?

Authors:  Tonke L de Jong; Nick J van de Berg; Lisette Tas; Adriaan Moelker; Jenny Dankelman; John J van den Dobbelsteen
Journal:  Med Devices (Auckl)       Date:  2018-08-03

10.  Hybrid control algorithm for flexible needle steering: Demonstration in phantom and human cadaver.

Authors:  Navid Shahriari; Janniko R Georgiadis; Matthijs Oudkerk; Sarthak Misra
Journal:  PLoS One       Date:  2018-12-31       Impact factor: 3.240

View more
  2 in total

1.  Axially rigid steerable needle with compliant active tip control.

Authors:  M de Vries; J Sikorski; S Misra; J J van den Dobbelsteen
Journal:  PLoS One       Date:  2021-12-16       Impact factor: 3.240

2.  Block-matching-based registration to evaluate ultrasound visibility of percutaneous needles in liver-mimicking phantoms.

Authors:  Juan A Sánchez-Margallo; Lisette Tas; Adriaan Moelker; John J van den Dobbelsteen; Francisco M Sánchez-Margallo; Thomas Langø; Theo van Walsum; Nick J van de Berg
Journal:  Med Phys       Date:  2021-10-31       Impact factor: 4.506

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.