| Literature DB >> 33408984 |
Chen Wang1,2,3, Ying Zhou4,3, Wei Li4, Xiaohu Liu1,2, Ling Xi4, Pengcheng Li1,2,5, Juncheng Wei4,6, Jinling Lu1,2,7.
Abstract
To examine the cervical lesions by using the tumor-targeted near-infrared (NIR) fluorescent dyes TMTP1-PEG4-ICG, a dual modal colposcope with visible reflectance imaging and fluorescence imaging was developed. NIR fluorescence imaging and visible light reflectance imaging were integrated together on a colposcope by designing the specific optics and adopting a dual sensor charge coupled device (CCD) camera. Patients with squamous cell carcinoma, adenocarcinoma, cervical intraepithelial neoplasia (CIN) and cervicitis were examined using this dual modal colposcope to validate its potential of cervical cancer detection. Fluorescent dye TMTP1-PEG4-ICG was applied to the cervix 30 minutes before inspection. The fluorescence images were collected after wiping the unbound fluorescent dye using normal saline. Signal to back ratios (SBR) of the fluorescence images were analyzed and compared with the histological analysis. The results suggest that the fluorescent colposcope combined with tumor-specific near-infrared fluorescent dyes TMTP1-PEG4-ICG could help to evaluate cervical lesions in real time.Entities:
Year: 2020 PMID: 33408984 PMCID: PMC7747887 DOI: 10.1364/BOE.410394
Source DB: PubMed Journal: Biomed Opt Express ISSN: 2156-7085 Impact factor: 3.732