Huan Liu1, Hongpeng Wang2, Zeshu Mu1, Lin Ye1, Yingjiu Jiang1.
Abstract
BACKGROUND: This study aims to investigate the efficacy and safety of heat and moisture exchanger on airway resistance in a cardiothoracic surgery intensive care unit.Entities:
Keywords: Airway resistance; intensive care unit; mechanical ventilation; thoracic surgery
Year: 2020 PMID: 33403131 PMCID: PMC7759046 DOI: 10.5606/tgkdc.dergisi.2020.20088
Source DB: PubMed Journal: Turk Gogus Kalp Damar Cerrahisi Derg ISSN: 1301-5680 Impact factor: 0.332
Figure 1The experimental model of this study.
Airway resistance in vitro lung model
| Without HME | With HME adding 0.9% saline | ||||
| 0 mL | 5 mL | 10 mL | 15 mL | ||
| Airway resistance (cmH2O/L/sec) | 17.3±1.0 | 18.8±0.3 | 19.4±0.2 | 20.3±1.0 | 30.2±1.7 |
| P value | 0.066* | 0.060† | 0.065† | <0.001† | |
| <0.001‡ | |||||
| HME: Heat and moisture exchanger; * Represents compared with control; † Represents compared with 0 mL group; ‡ Represents compared with 10 mL group. | |||||
Demographic and clinical characteristics of patients
| Features | 1-Day group | 3-Day group | 7-Day group | |
| n | n | n | p | |
| Sex | ||||
| Male | 5 | 6 | 7 | 0.132 |
| Female | 5 | 5 | 3 | |
| Age (year) | ||||
| ≤50 | 4 | 7 | 7 | 0.279 |
| >50 | 6 | 4 | 3 | |
| Smoking state | ||||
| Smoking | 4 | 6 | 5 | 0.643 |
| Non-smoking | 6 | 5 | 5 | |
| White blood cell count | ||||
| Normal | 9 | 6 | 5 | 0.170 |
| Abnormal | 1 | 5 | 5 | |
| APACHE II score | ||||
| <13 | 4 | 4 | 6 | 0.417 |
| ≥13 | 6 | 7 | 4 | |
| PaO2/FiO2 | ||||
| <220 | 4 | 5 | 5 | 0.101 |
| ≥220 | 6 | 6 | 5 | |
| APACHE: Acute physiology and chronic health evaluation. | ||||
Clinical assessment of humidifying efficacy
| 1-Day group | 3-Day group | 7-Day group | |||
| Mean±SD | Mean±SD | Mean±SD | |||
| Tidal volume (mL) | 480±11 | 482±35 | 469±62 | 0.877* | 0.586† |
| Respiratory rate (breaths per minute) | 13.5±0.7 | 13.6±0.8 | 14.0±1.1 | 0.687* | 0.229† |
| Mean peak airway pressure (cmH2O) | 17.9±0.7 | 17.9±0.7 | 18.7±1.2 | 0.977* | 0.082† |
| No of tracheal suctionings (per day) | 9.0±1.2 | 9.1±0.9 | 9.8±1.2 | 0.845* | 0.151† |
| No of tracheal instillations (per day) | 0.1±0.3 | 0.1±0.3 | 0.3±0.5 | 0.947* | 0.288† |
| SD: Standard deviation; * Represents 3-day group compared with 1-day group; † Represents 7-day group compared with 1-day group. | |||||
Humidity parameters
| 1-Day group | 3-Day group | 7-Day group | |||
| Mean±SD | Mean±SD | Mean±SD | |||
| Absolute humidity (cmH2O/L) | 30.8±0.5 | 30.7±0.4 | 30.7±0.6 | 0.768* | 0.758† |
| Relative humidity (%) | 99.1±0.5 | 99.1±0.5 | 99.2±0.3 | 0.881* | 0.682† |
| SD: Standard deviation; * Represents 3-day group compared to 1-day group; † Represents 7-day group compared to 1-day group. | |||||
Clinical assessment of airway resistance
| 1-Day group | 3-Day group | 7-Day group | ||||
| Mean±SD | Mean±SD | Mean±SD | ||||
| HME resistance (cmH2O/L/sec) | 3.2±0.1 | 3.2±0.1 | 3.5±0.2 | 0.163* | 0.002† | 0.015‡ |
| Ventilation duration (day) | 9.5±2.0 | 10.0±2.5 | 13.1±3.1 | 0.621* | 0.006† | 0.021‡ |
| Hospital stay (day) | 28.6±2.2 | 28.±2.5 | 31.1±2.5 | 0.902* | 0.030† | 0.041‡ |
| SD: Standard deviation; HME: Heat and moisture exchanger; * Represents 3-day group compared with 1-day group; † Represents 7-day group compared with 1-day group; ‡ Represents 7-day group compared with 3-day group. | ||||||