Fast online-customized (FOCUS) parallel transmission pulses: A combination of universal pulses and individual optimization.

Jürgen Herrler1, Patrick Liebig2, Rene Gumbrecht2, Dieter Ritter2, Sebastian Schmitter3, Andreas Maier4, Manuel Schmidt1, Michael Uder5, Arnd Doerfler1, Armin M Nagel5,6,7.   

Abstract

PURPOSE: To mitigate spatial flip angle (FA) variations under strict specific absorption rate (SAR) constraints for ultra-high field MRI using a combination of universal parallel transmit (pTx) pulses and fast subject-specific optimization.
METHODS: Data sets consisting of B0 , B 1 + maps, and virtual observation point (VOP) data were acquired from 72 subjects (study groups of 48/12 healthy Europeans/Asians and 12 Europeans with pathological or incidental findings) using an 8Tx/32Rx head coil on a 7T whole-body MR system. Combined optimization values (COV) were defined as combination of spiral-nonselective (SPINS) trajectory parameters and an energy regularization weight. A set of COV was optimized universally by simulating the individual RF pulse optimizations of 12 training data sets (healthy Europeans). Subsequently, corresponding universal pulses (UPs) were calculated. Using COV and UPs, individually optimized pulses (IOPs) were calculated during the sequence preparation phase (maximum 15 s). Two different UPs and IOPs were evaluated by calculating their normalized root-mean-square error (NRMSE) of the FA and SAR in simulations of all data sets. Seven additional subjects were examined using an MPRAGE sequence that uses the designed pTx excitation pulses and a conventional adiabatic inversion.
RESULTS: All pTx pulses resulted in decreased mean NRMSE compared to a circularly polarized (CP) pulse (CP = ~28%, UPs = ~17%, and IOPs = ~12%). UPs and IOPs improved homogeneity for all subjects. Differences in NRMSE between study groups were much lower than differences between different pulse types.
CONCLUSION: UPs can be used to generate fast online-customized (FOCUS) pulses gaining lower NRMSE and/or lower SAR values.
© 2021 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  7 T; SPINS; kT points; parallel transmission (pTx); ultra-high-field MRI; universal pulses

Year:  2021        PMID: 33400302     DOI: 10.1002/mrm.28643

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  1 in total

1.  Rigid motion-resolved B1+ prediction using deep learning for real-time parallel-transmission pulse design.

Authors:  Alix Plumley; Luke Watkins; Matthias Treder; Patrick Liebig; Kevin Murphy; Emre Kopanoglu
Journal:  Magn Reson Med       Date:  2021-12-27       Impact factor: 3.737

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.