Literature DB >> 33398551

Reducing fire and burn risk in the operating room-testing of a novel device.

Soham Roy1, Katherine M Yu2, Mark I Knackstedt2, Nathaniel H Webb3, Lee P Smith3,4.   

Abstract

BACKGROUND: Burn injury and operating room fires are significant risks for both surgical patients and staff. The purpose of this study was to examine the fire and burn risks associated with two types of fiberoptic light cables and evaluate the efficacy of a novel device in reducing the risk of these fire and burn injuries.
METHODS: A 300-W light source was connected sequentially to two standard fiberoptic cables (Storz and Olympus). The distal ends were buried in, or rested on, standard operating room materials including a cotton green towel and a blue propylene drape to assess the risk of fire formation or burn injury. The Gloshield device was then attached to the ends of the cables and the experiment was repeated. Trials simulating improper use of the device were then conducted with the Gloshield device placed either too deep or too shallow on the end of the light cable. All trials were conducted for a maximum of 10 min or until a positive result (burn or fire) was achieved. Trials were conducted in room air or with supplemental 100% oxygen and repeated for accuracy.
RESULTS: Both the Storz and Olympus fiber optic cables were capable of producing burns in standard operating room towels and drapes in control trials. The Gloshield device prevented thermal injury when properly attached in all conditions. Improper use trials demonstrated that the device may be ineffective when not applied properly.
CONCLUSIONS: The Gloshield device is effective in reducing the risk of thermal burn injury by protecting the distal ends of endoscopic light cables from operating room materials. However, the device needs to be attached appropriately in order to provide protective benefits.
© 2021. The Author(s), under exclusive licence to Springer Science+Business Media, LLC part of Springer Nature.

Entities:  

Keywords:  Burn injury; Endoscopic surgery; Fiberoptic cable; Fire hazard; Fire safety; Operating room fire

Mesh:

Year:  2021        PMID: 33398551     DOI: 10.1007/s00464-020-08209-9

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  3 in total

1.  Operating room fire safety.

Authors:  Stuart R Hart; Amit Yajnik; Jeffrey Ashford; Randy Springer; Sherry Harvey
Journal:  Ochsner J       Date:  2011

2.  Operating room fires in periocular surgery.

Authors:  Michael A Connor; Anne M Menke; Ivan Vrcek; John W Shore
Journal:  Int Ophthalmol       Date:  2017-05-20       Impact factor: 2.031

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.