T Olsen1.
Abstract
A method to measure the power of an intraocular lens (IOL) in situ using an extended range keratometer and a depth measuring apparatus is described. With this method, the IOL power was measured in ten patients with an anterior chamber IOL and in ten patients with a posterior chamber IOL and compared with the labeled power. The total mean error (measured power minus labeled power) was found to be 0.01 diopters (D) with a standard deviation of 0.22 D. The largest error was 0.40 D.Entities:
Mesh:
Year: 1988 PMID: 3339550 DOI: 10.1016/s0886-3350(88)80066-x
Source DB: PubMed Journal: J Cataract Refract Surg ISSN: 0886-3350 Impact factor: 3.351