Performance evaluation of the new hematology analyzer UniCel DxH 900.

Alicia Martinez-Iribarren1, Xavier Tejedor1, Àngels Sala Sanjaume1, Alba Leis1, Maria Doladé Botias1, Cristian Morales-Indiano1.   

Abstract

INTRODUCTION: UniCel DxH900 (Beckman Coulter, Miami, Florida, USA) is a quantitative, multi-parameter, automated hematology analyzer for in vitro diagnostic use in clinical laboratories. The aim of this study was to evaluate the analytical performance of the new DxH900 analyzer to verify its diagnostic and clinical utility in the hematology laboratory of a tertiary care hospital in Spain. The most important and novel feature offered by DxH900 analyzer is providing MDW (monocyte distribution width), a new hematologic parameter which is being clinically validated as an early sepsis indicator with promising results.
METHODS: We evaluated imprecision (including MDW), linearity, and carryover of DxH900. Method comparison for cell blood count (CBC) was performed in relation to DxH800 with 100 samples. We compared leukocyte differential (DIFF) from DxH900 with manual 400-cell differential. 390 samples were assessed for flag performance.
RESULTS: Results obtained for between days and within-run imprecision were good. DxH900 showed excellent linearity (R = 1.00) over analytical range for white blood cells, red blood cells, hemoglobin, platelets, and reticulocyte count (RET) (R = 0.96) and no significant carryover effect. CBC and RET on the DxH900 correlated well with DxH800 (R ≥ 0.99). Comparison with manual differential showed excellent correlation (R ≥ 0.88), except for basophils. Flagging performance exhibited sensitivity over 90% for majority of alarm messages and very high negative predictive value (over 95%).
CONCLUSION: UniCel DxH900 Coulter analyzer provides reliable results and fully comparable to DxH800. DxH900 is an accurate, highly precise analyzer with good analytical performances to be used effectively in high-volume laboratories.
© 2021 John Wiley & Sons Ltd.

Entities:  

Keywords:  DxH900; cell blood count; flag; hematology analyzer; leukocyte differential; performance evaluation

Year:  2021        PMID: 33389827     DOI: 10.1111/ijlh.13448

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  2 in total

1.  Establishment of a new protein C detection system based on chromogenic substrate assay and its clinical diagnostic value for deep vein thrombosis.

Authors:  Wenfei Lu; Anshan Ge; Jinxia Zhao; Hui Rong; Chao Zhu; Limin Lun
Journal:  J Clin Lab Anal       Date:  2021-11-13       Impact factor: 2.352

Review 2.  Diagnostic and Prognostic Value of Monocyte Distribution Width in Sepsis.

Authors:  Juehui Wu; Laisheng Li; Jinmei Luo
Journal:  J Inflamm Res       Date:  2022-07-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.