The feasibility of anterior 180° 99m Tc-sestamibi parathyroid SPECT/CT.

Ju-Ling Cheng1, Chien-Chin Hsu1, Pei-Wen Wang1,2, Kuo-Chiang Sung3,4, Nan-Tsing Chiu5, Lan-Yi Chiu1, Yung-Cheng Huang6.   

Abstract

OBJECTIVE: For 99mTc-sestamibi parathyroid single photon emission computed tomography/computed tomography (SPECT/CT), the traditional full-arc 360° acquisition technique has been suggested but not the theoretically low tissue-attenuated anterior 180° method. We aimed to compare the diagnostic performances and target-to-background ratios of anterior 180° and 360° SPECT/CT imaging.
METHODS: Ninety-nine patients who underwent 99mTc-sestamibi scintigraphy and received a surgical-pathological diagnosis of parathyroid adenoma or hyperplasia were enrolled. The SPECT/CT reconstructed images with anterior 180° and full-arc 360° data were interpreted by two physicians using a scoring scale, and the lesions were semi-quantified using target-to-background ratios for both image sets.
RESULTS: In total, 113 abnormal parathyroid lesions were identified on the SPECT/CT images. The agreements of interpretation between the two image sets and readers were very good (κ value: 0.83-1.00). The accuracies of summative interpretation for the anterior 180° and full-arc 360° SPECT/CT were 83.04% and 82.46%, respectively. The target-to-background ratios were significantly higher for the anterior 180° than the full-arc 360° images (P < 0.01).
CONCLUSIONS: The anterior 180° SPECT/CT technique, a time-saving method, can provide satisfactory diagnostic performance and superior lesion contrast compared with the full-arc 360° SPECT/CT technique in 99mTc-sestamibi parathyroid imaging.

Entities:  

Keywords:  99mTc-sestamibi; Hyperparathyroidism; Parathyroid; SPECT/CT

Year:  2021        PMID: 33389667     DOI: 10.1007/s12149-020-01553-9

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  4 in total

1.  Posterior 180 degrees 99mTc-dimercaptosuccinic acid renal SPECT.

Authors:  N J Peng; C G Kwok; Y H Chiou; G H Jao; D G Tsay; R S Liu; H W Strauss
Journal:  J Nucl Med       Date:  1999-01       Impact factor: 10.057

2.  Accuracy of parathyroid imaging: a comparison of planar scintigraphy, SPECT, SPECT-CT, and C-11 methionine PET for the detection of parathyroid adenomas and glandular hyperplasia.

Authors:  Mehmet Ozer Oksüz; Helmut Dittmann; Corinna Wicke; Karsten Müssig; Roland Bares; Christina Pfannenberg; Susanne Martina Eschmann
Journal:  Diagn Interv Radiol       Date:  2011-02-08       Impact factor: 2.630

3.  Detection and localization of parathyroid adenomas in patients with hyperparathyroidism using a single radionuclide imaging procedure with technetium-99m-sestamibi (double-phase study)

Authors:  R Taillefer; Y Boucher; C Potvin; R Lambert
Journal:  J Nucl Med       Date:  1992-10       Impact factor: 10.057

4.  Identifying abnormal parathyroid glands in the thyroid uptake area using technetium-99m-sestamibi and factor analysis of dynamic structures.

Authors:  C Billotey; A Aurengo; Y Najean; E Sarfati; J L Moretti; M E Toubert; J D Rain
Journal:  J Nucl Med       Date:  1994-10       Impact factor: 10.057

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.