Literature DB >> 33389052

New fixation method for Pauwels type III femoral neck fracture: a finite element analysis of sliding hip screw, L-shaped, and L-shaped with medial plate.

Anderson Freitas1,2, Ricardo Lourenço Bontempo1, Frank Anderson Ramos Azevedo1, Leonardo Rigobello Battaglion3, Marcos Noberto Giordano4, Fabrício Reichert Barin5, Vincenzo Giordano6.   

Abstract

OBJECTIVE: In this study, we aimed to evaluate the biomechanical behavior of three fixations for Pauwels type III fractures (sliding hip screw (SHS), L-shaped, and L-shaped with medial plate), by finite element analysis (FEM).
METHODS: Three internal fixators were developed to treat Pauwels type III fracture by finite elements: SHS; L-shaped; and L-shaped with medial plate. Under the same conditions, localized and total vertical fracture displacement, maximum and minimum principal and von Mises stresses were evaluated.
RESULTS: The localized and total vertical displacement evaluated for the SHS, L-shaped, and L-shaped with medial plate were 0.15 mm, 0.17 mm, and 0.07 mm (localized), and 4.52 mm, 6.97 mm, and 6.83 mm (total), respectively. The maximum values obtained in the upper region of the femoral neck for the internal fixations were 1.43 MPa, 1.29 MPa, and 1.24 MPa, and the minimum values obtained in the lower region of the femoral neck were  - 0.73 MPa,  - 1.09 MPa, and  - 1.03 MPa, respectively. The maximum Von Mises peak stress values were 6.35 MPa, 10.7 MPa, and 16.2 MPa for the fixation models using the SHS, L-shaped, and L-shaped with medial plate, respectively.
CONCLUSION: The present FEM analysis showed that SHS yields better results in terms of total vertical displacements, maximum distribution, and Von Mises peak stresses reduction. On the other hand, the L-shaped construction plus a medial plate decreases localized vertical displacements and maximum principal distribution when compared to the SHS and L-shaped constructions. These results demonstrate that both constructions, SHS and L-shaped plus a medial plate, are biomechanically efficient for the fixation of Pauwels type III femoral neck fractures.
© 2021. Springer-Verlag France SAS, part of Springer Nature.

Entities:  

Keywords:  Biomechanics; Femoral neck fracture; Finite element analysis; Internal fixation

Year:  2021        PMID: 33389052     DOI: 10.1007/s00590-020-02824-x

Source DB:  PubMed          Journal:  Eur J Orthop Surg Traumatol        ISSN: 1633-8065


  2 in total

1.  Results of internal fixation of Pauwels type-3 vertical femoral neck fractures.

Authors:  Martyn J Parker
Journal:  J Bone Joint Surg Am       Date:  2009-02       Impact factor: 5.284

2.  Finite element analysis of a new plate for Pauwels type III femoral neck fractures.

Authors:  Gang Wang; Yong Tang; Xuhua Wu; Huilin Yang
Journal:  J Int Med Res       Date:  2020-02       Impact factor: 1.671

  2 in total
  1 in total

1.  In Silico Finite Element Modeling of Stress Distribution in Osteosynthesis after Pertrochanteric Fractures.

Authors:  Jacek Lorkowski; Mieczyslaw Pokorski
Journal:  J Clin Med       Date:  2022-03-28       Impact factor: 4.241

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.